MedCAD AccuPlan Orthopedics System

K223421 · Medcad · PBF · Sep 20, 2023 · Orthopedic

Device Facts

Record IDK223421
Device NameMedCAD AccuPlan Orthopedics System
ApplicantMedcad
Product CodePBF · Orthopedic
Decision DateSep 20, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2

Indications for Use

MedCAD® AccuPlan® Orthopedics System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the system and the result is an output data file. This file may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models of the fibula and ilium. surgical guides for harvesting bone grafts from the fibula or ilium, and surgical planning case reports for use in maxillofacial reconstructive surgeries. MedCAD® AccuPlan® Orthopedics System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options. The MedCAD® AccuPlan® Orthopedics System is indicated for use in adolescents (greater than 12 to 21 years of age) and adults.

Device Story

System processes 3D CT images to create patient-specific anatomical models, surgical guides, and planning reports for maxillofacial reconstructive surgery; involves harvesting bone from fibula or ilium. Operated by trained MedCAD employees using COTS software; physician provides input/feedback during planning. Outputs produced via additive manufacturing. Used in clinical settings to assist surgeons in pre-operative simulation and surgical execution. Benefits include patient-specific surgical planning and guidance, potentially improving surgical accuracy and outcomes.

Clinical Evidence

Bench testing only. Wear debris testing (ASTM F1877-16) on titanium guides confirmed safety. Fit and form validation performed using worst-case CT inputs and anatomical models; all samples met acceptance criteria.

Technological Characteristics

Materials: Biocompatible polymers and Ti-6Al-4V ELI Titanium Alloy (ASTM F136). Additive manufacturing process. Steam sterilization (SAL 10^-6). COTS software used for design. Connectivity: Standalone system for image processing and additive manufacturing.

Indications for Use

Indicated for adolescents (>12 to 21 years) and adults requiring maxillofacial reconstructive surgery involving bone grafts from the fibula or ilium.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ September 20, 2023 Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. MedCAD % Justin Gracyalny Regulatory Affairs Manager Secure BioMed Evaluations 7828 Hickory Flat Highway Suite 120 Woodstock. Georgia 30188 Re: K223421 Trade/Device Name: MedCAD® AccuPlan® Orthopedics System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: PBF, LLZ Dated: August 18, 2023 Received: August 18, 2023 Dear Mr. Gracyalny: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shumaya Ali-S Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K223421 Device Name MedCAD® AccuPlan® Orthopedics System #### Indications for Use (Describe) MedCAD® AccuPlan® Orthopedics System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the system and the result is an output data file. This file may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models of the fibula and ilium. surgical guides for harvesting bone grafts from the fibula or ilium, and surgical planning case reports for use in maxillofacial reconstructive surgeries. MedCAD® AccuPlan® Orthopedics System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options. The MedCAD® AccuPlan® Orthopedics System is indicated for use in adolescents (greater than 12 to 21 years of age) and adults. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### K223421 510(k) Summary MedCAD® AccuPlan® Orthopedics System ### September 18, 2023 #### Sponsor MedCAD 501 S 2nd Ave, Suite A-1000 Dallas, TX 75226 (214) 453-8864 x305 ## Contacts Secure BioMed Evaluations Justin Gracyalny, MSE Linda Braddon, Ph.D. 7828 Hickory Flat Highway, Suite 120 Woodstock, GA 30188 770-837-2681 Regulatory(@SecureBME.com ### Name of Device and Classification Name Device Name: MedCAD® AccuPlan® Orthopedics System Common Device: Orthopaedic Surgical Planning And Instrument Guide Regulation Name and Number: 21 CFR 888.3030 Single/multiple component metallic bone fixation appliances and accessories, 21 CFR 892.2050 Medical image management and processing system Product Code: PBF, LLZ Classification Panel: Orthopedic ### Predicate Device CenterMed Patient Matched Assisted Surgical Planning (ASP) System (K211614) Reference Devices MedCAD® AccuPlan® System (K192282 / K223024) ### Indications for Use: MedCAD® AccuPlan® Orthopedics System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the system and the result is an output data file. This file may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models of the fibula and ilium, surgical guides for harvesting bone grafts from the fibula or ilium, and surgical planning case reports for use in maxillofacial reconstructive surgeries. MedCAD® AccuPlan® Orthopedics System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options. The MedCAD® AccuPlan® Orthopedics System is indicated for use in adolescents (greater than 12 to 21 years of age) and adults. {4}------------------------------------------------ # K223421 ## Device Description The MedCAD® AccuPlan® Orthopedics System is a collection of software and associated additive manufacturing equipment intended to provide a variety of outputs to support harvesting of bone to support maxillofacial reconstructive surgeries. The system uses electronic medical images of the patient's anatomy with input from the physician to manipulate original patient images for planning and executing surgery. The patient specific outputs from the system include anatomical models, surgical guides, and patient-specific case reports. Following the MedCAD® Quality System and specific Work Instructions, trained employees utilize Commercial Off-The-Shelf (COTS) software to manipulate 3-D medical Computed Tomography (CT) images to create patient-specific physical and digital outputs. The process requires clinical input and review from the physician during and prior to delivery of the final outputs. While the process and dataflow vary somewhat based on the requirements of a given patient and physician, the following description outlines the functions of key sub-components of the system, and how they interact to produce the defined system outputs. It should be noted that the system is operated only by trained MedCAD employees, and the physician does not directly input information. The physician provides input for model manipulation and interactive feedback through viewing of digital models of system outputs that are modified by the engineer during the planning session. The MedCAD® AccuPlan® Orthopedics System is made up of two individual pieces of software for the design and various manufacturing equipment integrated to provide a range of anatomical models (physical and digital), surgical guides, and patient-specific planning reports for harvesting of bone from the fibula and ilium for use in maxillofacial reconstructive surgeries. The MedCAD® AccuPlan® Orthopedics System requires an input 3-D image file from medical imaging systems (i.e., CT). This input is then used, with support from the prescribing physician to provide the following potential outputs to support maxillofacial reconstructive surgery. Each system output is designed with physician inout and reviewed by the physician prior to finalization. All outputs are used only with direct physician involvement to reduce the criticality of the outputs. System outputs include: - Anatomical Models ● - Surgical Guides - Patient-Specific Case Reports ● # Sterilization Validation Sterilization validation was conducted in accordance with international standard ISO 17665 and FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling." to a sterility Assurance Level (SAL) of 1x10 . All test method acceptance criteria were met. # Biocompatibility Validation Biocompatibility validation was conducted in accordance with international standard ISO 10993-1 and FDA guidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The results of the testing adequately address biocompatibility for the output devices and their intended use. {5}------------------------------------------------ # K223421 # Performance Testing Performance testing for the MedCAD® AccuPlan® Orthopedics System is summarized in the table below: | Test | Test Method Summary | Results | |----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Wear Debris<br>Testing | Cutting / drilling instruments were used on a<br>worst-case titanium surgical guide. The quantity<br>of wear debris generation was characterized.<br>Wear particles were characterized per ASTM<br>F1877-16. | PASS<br>The quantity and<br>morphology of wear<br>debris generated by the<br>subject device under<br>worst case use<br>conditions aligns with<br>values as reported in the<br>literature to be safe. | | Fit and Form<br>Validation | Subject devices were manufactured from worst-<br>case CT scan input data. The manufactured<br>devices were optically scanned to verify<br>alignment with the 3D model. Evaluation of fit<br>was also validated by fitting the guide over the<br>corresponding defect in a representative<br>anatomical model. | PASS<br>All samples met the<br>predetermined<br>acceptance criteria. | # Substantial Equivalence MedCAD® AccuPlan® Orthopedics System is substantially equivalent to the identified predicate based on indications for use, principles of operation, technological characteristics, inputs, and outputs. | Property | MedCAD® AccuPlan®<br>Orthopedics System | CenterMed Patient Matched<br>Assisted Surgical Planning (ASP)<br>System | |---------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) number | K223421 | K211614 | | Indications for Use | MedCAD® AccuPlan® Orthopedics<br>System is intended for use as a<br>software system and image<br>segmentation system for the transfer<br>of imaging information from a<br>medical scanner such as a CT based<br>system. The input data file is<br>processed by the system and the<br>result is an output data file. This file<br>may then be provided as digital<br>models or used as input to a rapid<br>prototyping portion of the system that<br>produces physical outputs including<br>anatomical models of the fibula and<br>ilium, surgical guides for harvesting | CenterMed Patient Matched Assisted<br>Surgical Planning (ASP) System is<br>intended for use as a software system<br>and image segmentation system for<br>the transfer of imaging information<br>from a medical scanner such as a CT<br>based system. The input data file is<br>processed by the ASP system and the<br>result is an output data file. This file<br>may then be provided as digital<br>models or used as input to a rapid<br>prototyping portion of the system that<br>produces physical outputs including<br>anatomical models of the Fibula and<br>Ilium, surgical guides for harvesting | | Property | MedCAD® AccuPlan®<br>Orthopedics System | CenterMed Patient Matched<br>Assisted Surgical Planning (ASP)<br>System | | | bone grafts from the fibula or ilium, and surgical planning case reports for use in maxillofacial reconstructive surgeries. MedCAD® AccuPlan® Orthopedics System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options. The MedCAD® AccuPlan® Orthopedics System is indicated for use in adolescents (greater than 12 to 21 years of age) and adults. | bone grafts from the Fibula or Ilium, and surgical planning case reports for use in Maxillofacial reconstructive surgeries. CenterMed Patient Matched ASP System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options. | | Preoperative software | Yes | Yes | | Additive<br>manufacturing of<br>surgical guides and<br>models | Yes | Yes | | Data inputs | Images from medical scanners | Images from medical scanners | | Data outputs | Output for patient-specific surgical<br>guides and anatomical models | Output for patient-specific surgical<br>guides and anatomical models | | Physical outputs | Patient-specific surgical guides,<br>anatomical models, and patient-<br>specific case reports | Patient-specific surgical guides,<br>anatomical models, and patient-<br>specific case reports | | Materials | Biocompatible polymers and<br>Ti-6Al-4V ELI Titanium Alloy per<br>ASTM F136 | Polyamide-12 (PA-12) with 316L<br>stainless steel sleeves when used as<br>cutting guides | | Recommended<br>Temporary Screw<br>Diameter | 1.5mm – 2.1mm | 1.5mm – 2.1mm | | Recommended<br>Temporary Screw<br>Length | 5mm – 22mm | 5mm – 22mm | | Recommended<br>Temporary Screw<br>Style | Non-Locking | Drill-Free, Tapping-Free | | Sterilization | Provided non-sterile and is steam<br>sterilized by the end-user (SAL 10⁻⁶) | Provided non-sterile and is steam<br>sterilized by the end-user (SAL 10⁻⁶) | | Manufacturing<br>Method | Additive Manufacturing | Additive Manufacturing | | Patient Contact | Implant – Tissue / Bone; Limited (<<br>24 hours) | Implant – Tissue / Bone; Limited (<<br>24 hours) | | Patient Population | Adults and Pediatrics (12+) | Adults | {6}------------------------------------------------ {7}------------------------------------------------ # K223421 # Comparison of Technological Characteristics with the Predicate Device MedCAD® AccuPlan® Orthopedics System is substantially equivalent to its predicate device (K211614 CenterMed Patient Matched Assisted Surgical Planning (ASP) System). ## Similarities to Predicate The MedCAD® AccuPlan® Orthopedics System has the same intended use and similar technological characteristics as the identified predicate device. The system employs similar fundamental technologies as the identified predicate including software image transfer, manipulation, and surgical planning. The principles of operation and technological characteristics are either identical or substantially equivalent to the predicate. The system has similar technological characteristics including: - System inputs: Images from medical scanner systems (CT) - System outputs: Additively manufactured physical and/ or digital outputs such as patientspecific anatomical models and surgical guides for use in harvesting bone to support maxillofacial reconstructive surgeries - Software: Use of commercial off-the-shelf software to support the design phase of the device. ● - Sterility assurance level of 1x10-6 ● The intended use of the subject device and the predicate both provide tools and accessories (software for image manipulation, anatomical models, and guides) for use in maxillofacial reconstructive surgery. Additionally, both the subject and the predicate device are intended to be used by trained personnel with active support from the surgeon. ## Differences to Predicate The only differences between the subject and predicate device relate to the addition of the 12+ pediatric population for the subject device and minor geometrical differences. A pediativ risk analysis was performed to support the change in patient population. Performance testing demonstrates that the minor material and geometrical differences do not raise new questions for safety and effectiveness. ### Conclusion The MedCAD® AccuPlan® Orthopedics System is substantially equivalent to its predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...