← Product Code [PBF](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PBF) · K220706

# MyPAO Guides (K220706)

_Medacta International S.A. · PBF · May 30, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PBF/K220706

## Device Facts

- **Applicant:** Medacta International S.A.
- **Product Code:** [PBF](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PBF.md)
- **Decision Date:** May 30, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic

## Indications for Use

MyPAO patient-specific guides are devices intended to be used as anatomical guides are specifically designed based on the CT images of the patient (i.e., pelvis, proximal femur). The guides are intended to assist the surgeon in the positioning of the acetabular fragment during periacetabular osteotomies to treat patients who require to undergo periacetabular osteotomy. MyPAO guides are intended for single use only.

## Device Story

MyPAO Guides are patient-specific, single-use anatomical guides designed from preoperative CT scans; used to assist surgeons in positioning the acetabular fragment during periacetabular osteotomies. System includes repositioning PSI guides (pre-op drill guide, ilium/acetabulum guides, realignment arc), preoperative/postoperative bone models, and reusable instruments. Surgeon uses MyPAO software to perform 3D preoperative planning, now including 3D femoral head coverage and acetabular anteversion parameters. Guides replicate desired correction before final fixation of acetabular fragment. Device used in clinical setting; aids surgeon in achieving planned realignment; benefits patient through improved surgical accuracy in complex pelvic osteotomies.

## Clinical Evidence

Single-arm study (n=10; 8 female, 2 male; ages 18-53) conducted outside the U.S. compared preoperative planned correction to postoperative results using 2D planar x-rays. Primary endpoints: lateral center edge angle and acetabular index. Acceptance criteria: mean error ≤ 4 degrees. Results met criteria; no adverse effects or complications observed.

## Technological Characteristics

Patient-specific, single-use, non-sterile guides and bone models. Designed via CT-based 3D planning. Software-driven design. Biocompatibility leveraged from K200792. Reusable instruments used in conjunction with guides.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- VSP Orthopedics System ([K190044](/device/K190044.md))

## Reference Devices

- MySpine S2-Alar/Alar-Iliac Pedicle Screw Placement Guides ([K200792](/device/K200792.md))
- MyPAO Planning Report ([K200589](/device/K200589.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 30, 2023

Medacta International S.A. % Chris Lussier Senior Director, Quality, Regulatory and Clinical Research Medacta USA 3973 Delp Street Memphis, Tennessee 38118

Re: K220706

Trade/Device Name: MyPAO Guides Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: PBF Dated: March 9, 2022 Received: March 10, 2022

Dear Chris Lussier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Limin Sun -S

Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K220706

Device Name MyPAO Guides

### Indications for Use (Describe)

MyPAO patient-specific guides are devices intended to be used as anatomical guides are specifically designed based on the CT images of the patient (i.e., pelvis, proximal femur). The guides are intended to assist the surgeon in the positioning of the acetabular fragment during periacetabular osteotomies to treat patients who require to undergo periacetabular osteotomy. MyPAO guides are intended for single use only.

| Type of Use (Select one or both, as applicable)                                      |                                                                                     |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

#### I. Submitter

Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66

Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Applicant Correspondent: Chris Lussier, Sr. Director of Quality, Regulatory, and Clinical, Medacta USA Date Prepared: March 09, 2022 Date Revised: May 30, 2023

#### II. Device

| Device Proprietary Name: | MyPAO Guides                                                                   |
|--------------------------|--------------------------------------------------------------------------------|
| Common or Usual Name:    | Orthopaedic Surgical Planning And Instrument Guides                            |
| Classification Name:     | Single/multiple component metallic bone fixation appliances<br>and accessories |
| Primary Product Code:    | PBF                                                                            |
| Regulation Number:       | 21 CFR 888.3030                                                                |
| Device Classification    | II                                                                             |

#### III. Predicate Device

Substantial equivalence is claimed to the following primary predicate devices:

- Primary predicate device:
	- >> VSP Orthopedics System, K190044, 3D Systems, Inc.

Reference device :

- A MySpine MySpine S2-Alar/Alar-Iliac Pedicle Screw Placement Guides, K200792, Medacta International SA
- A MyPAO Planning Report, K200589, Medacta International SA

#### Device Description IV.

The MyPAO Guides enlarge Medacta's MyPAO realignment system, already including MyPAO Planning report, in order to provide the surgeons a further help in the realignment phase after a Periacetabular Osteotomy.

Specifically, the current submission aims at obtaining the clearance for the changes made to MyPAO Planning report and for the following new devices:

{4}------------------------------------------------

- Reposition PSI guides (Pre-op drill guide, Ilium and Acetabulum Rep Guides, Realignment ● Arc):
- Preoperative and postoperative bone models; ●
- Reusable instruments to be used with the guides. ●

MyPAO Guides are patient specific guides designed on the basis of the patient's preoperative CT scan and subsequent preoperative planning and they aim at replicating the desired correction before the final fixation of the acetabular fragment.

The MyPAO Guides as well as their bone models are single-use and they are provided non-sterile.

The 3D pre-operative planning based on the patient's CT scans is performed through MyPAO software that is the same software already used to generate MyPAO Planning Report. The only change made to the already cleared MyPAO Planning report (K200589) regards the addition of two parameters in the report, i.e. 3D femoral head coverage and acetabular anteversion.

#### V. Indications for Use

MyPAO patient-specific guides are devices intended to be used as anatomical guides. The guides are specifically designed based on the CT images of the patient (i.e., pelvis, proximal femur). The guides are intended to assist the surgeon in the positioning of the acetabular fragment during periacetabular osteotomies to treat patients who require to undergo periacetabular osteotomy. MyPAO guides are intended for single use only.

The indications for use differ because the predicate for use in non-joint replacing osteotomies in the non-sacrum pelvis area, whereas the subject device has specific indications for use in periacetabular osteotomy (a type of non-joint replacement, non-sacrum pelvis indication). This specificity does not change the intended use of the instruments for guided orthopedic corrections of the pelvis.

#### VI. Comparison of Technological Characteristics

The subject and primary predicate device (K190044) indications for use are substantially equivalent: both are intended to be used as a surgical instrument to assist the surgeon during non-joint replacing osteotomies.

Both the subject and the predicate device include single use patient matched guides. The subject device differs from the predicate because it allows orientation of one fragment to the remaining pelvic bone, whereas the predicate allows for orientation of instruments to bone.

{5}------------------------------------------------

## Discussion

The subject and primary predicate device (K190044) indications for use are substantially equivalent: both are intended to be used as a surgical instrument to assist the surgeon during non-joint replacing osteotomies.

The comparison of technological characteristics and performance data provided within this submission, shows that there are no new risks associated with the subject devices design, and supports the substantial equivalence of the MyPAO Guides to the identified predicate devices.

The only change made to the already cleared MyPAO Planning report (K200589) has not any impact on device's safety and effectiveness since it simply regards the addition of two parameters in the report, i.e. 3D femoral head coverage and acetabular anteversion.

#### VII. Performance Data

Based on the risk analysis, testing activities were conducted to written protocols. The following validation and tests are being provided in support of the substantial equivalence determination:

## Non-Clinical Studies

- Software validation ●
- MyPAO Design validation to validate functionality and usability ●
- MyPAO Design validation to demonstrate the ability of MyPAO guides to replicate ● preoperative planning
- MyPAO guides positioning variability Cadaveric test validation ●

Biocompatibility data submitted in support of the reference device (K200792) was leveraged.

## Clinical Studies:

- 10 patients were treated by three surgeons using the MyPAO surgical planning software and . guides. The primary safety and effectiveness data included a surgeon formative evaluation and comparison of the preoperative planned correction as compared to the post-operative correction results. 2D planar x-rays were used to measure the final lateral center edge and acetabular index.
	- This was a single-arm study with a literature control. The study was conducted outside O of the United States.
	- The acceptance criteria was a mean error of 4 degrees or less for the lateral center edge o angle and acetabular index. This criteria was identified based on literature for conventional surgical technique accuracy.
	- The clinical data set met the acceptance criteria. o
- The 10 patients included: ●
	- 0 8 female
	- 2 male o
	- Ages: from 18 to 53 o

{6}------------------------------------------------

- . There were no observed adverse effects or complications.
## VIII. Conclusion

The information provided above supports that the MyPAO Guides are substantially equivalent to the identified predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PBF/K220706](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PBF/K220706)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
