Salto Talaris Ankle PSI System

K241148 · 3D Systems, Inc. · OYK · Aug 30, 2024 · Orthopedic

Device Facts

Record IDK241148
Device NameSalto Talaris Ankle PSI System
Applicant3D Systems, Inc.
Product CodeOYK · Orthopedic
Decision DateAug 30, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3110
Device ClassClass 2

Indications for Use

The Salto Talaris Ankle PSI System is intended to be used as patient specific surgical planning and instrumentation to assist in the positioning of total ankle replacement components intraoperatively, and in guiding bone cutting. The Salto Talaris Ankle PSI System is intended for use with Smith + Nephew's Salto Talaris Total Ankle System and its cleared indications for use.

Device Story

System provides patient-specific surgical guides, anatomical models, and case reports for total ankle replacement. Input data consists of patient CT scans. Guides are designed to match patient-specific distal tibial and proximal talar anatomy. Used intraoperatively by orthopedic surgeons to guide bone resection and pin placement for the Smith + Nephew Salto Talaris Total Ankle System. Guides facilitate accurate positioning of ankle prosthesis components. Benefits include improved surgical planning and alignment accuracy compared to standard instrumentation.

Clinical Evidence

No clinical data. Non-clinical performance testing included mechanical integrity, debris generation, and inter-designer variability analysis. Cadaveric comparison testing evaluated implant alignment accuracy and guide usability against standard instrumentation for the Salto Talaris Total Ankle System, demonstrating similar accuracy and functionality.

Technological Characteristics

Patient-specific surgical guides manufactured via Selective Laser Sintering (SLS) using DuraForm® ProX PA (polyamide) and surgical grade stainless steel. System utilizes CT-based anatomical modeling. Standalone instrumentation system; non-implantable.

Indications for Use

Indicated for patients undergoing total ankle replacement surgery requiring patient-specific surgical planning and instrumentation to assist in component positioning and bone cutting, specifically for use with the Smith + Nephew Salto Talaris Total Ankle System.

Regulatory Classification

Identification

An ankle joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace an ankle joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces and has no linkage across-the-joint. This generic type of device includes prostheses that have a talar resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a tibial resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. August 30, 2024 3D Systems, Inc. Ashley Dawson Director, Regulatory Affairs 5381 South Alkire Circle Littleton, Colorado 80127 Re: K241148 Trade/Device Name: Salto Talaris Ankle PSI System Regulation Number: 21 CFR 888.3110 Regulation Name: Ankle Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: OYK Dated: July 31, 2024 Received: July 31, 2024 Dear Ashley Dawson: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {2}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Lixin Liu -S Lixin Liu, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use Submission Number (if known) K241148 Device Name Salto Talaris Ankle PSI System Indications for Use (Describe) The Salto Talaris PSI System is intended to be used as patient specific surgical planning and instrumentation to assist in the positioning of total ankle replacement components intraoperatively, and in guiding bone cutting. The Salto Talaris PSI System is intended for use with Smith + Nephew's Salto Talaris Total Ankle System and its cleared indications for use. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the 3D Systems logo. The logo consists of a cube-like shape with the letters '3D' inside, followed by the words '3D SYSTEMS' in a bold, sans-serif font. The color of the logo is gray. # 510(K) SUMMARY ### 1. INTRODUCTION This document contains the 510(k) summary for the Salto Talaris Ankle PSI System. The content of this summary is based on the requirements of 21 CFR 807.92. ### 2. SUBMITTER | Name: | 3D Systems, Inc. | |----------------------|------------------------------------------------------------------------------------------------------| | Address: | 5381 South Alkire Circle<br>Littleton, CO 80127, USA<br>Phone: (720) 643-1001<br>Fax: (720) 643-1009 | | Official Contact: | Dr. Ashley Dawson<br>Director of Regulatory Affairs | | Date Prepared: | August 30, 2024 | | 3. DEVICE | | | Trade Name: | Salto Talaris Ankle PSI System | | Common Name: | Ankle Arthroplasty Implantation System | | Classification Name: | Ankle joint metal/polymer semi-constrained cemented prosthesis | | Classification: | Class II, 21 CFR 888.3110 | | Product Code: | OYK | | 4. PREDICATE DEVICE | | | Predicate Device: | K193432 Vantage® PSI System, Product Code OYK | #### 5. REFERENCE DEVICE Reference Device: K182878 Salto Talaris Total Ankle System, Produce Code HSN ### 6. DESCRIPTION OF THE DEVICE 3D Systems' Salto Talaris Ankle PSI System consists of patient specific outputs including surgical guides, anatomical models, and case reports. The Salto Talaris Ankle PSI System guides are made from biocompatible nylon and surgical grade stainless steel and are designed to fit the contours of the patient's distal tibial and proximal talar anatomy. The surgical guides in combination with the Smith+Nephew Salto Talaris Total Ankle System instruments, facilitate the positioning of Salto Talaris Total Ankle Prostheses. {5}------------------------------------------------ ### 7. INTENDED USE/INDICATIONS FOR USE The Salto Talaris Ankle PSI System is intended to be used as patient specific surgical planning and instrumentation to assist in the positioning of total ankle replacement components intraoperatively, and in guiding bone cutting. The Salto Talaris Ankle PSI System is intended for use with Smith + Nephew's Salto Talaris Total Ankle System and its cleared indications for use. ### 8. INDICATIONS FOR USE COMPARISON The Salto Talaris Ankle PSI System has the same Indications for Use as the predicate device with the only change being the orthopedic implant that the Patient-Specific Instrumentation is guiding. ### 9. TECHNOLOGICAL COMPARISON The principles of operation and technological characteristics are substantially equivalent between the subject Salto Talaris Ankle PSI System and the predicate Vantage PSI System (K193432). Both the predicate and the subject devices are designed with CT-based methods to produce patient-specific instrumentation. The Salto Talaris Ankle PSI System guides are manufactured using SLS technology with DuraForm® ProX PA (polyamide) material and surgical stainless steel components. Both the subject and predicate devices guide the placement of pins and resections of tibial and talar anatomies for total ankle arthroplasties. The difference between the subject device and the predicate device is that the subject device is designed to guide the accurate placement of the Smith+Nephew Salto Talaris Total Ankle Prosthesis. ### 10. NON-CLINICAL AND/OR CLINICAL TESTS SUMMARY Non-clinical performance testing on the Salto Talaris Ankle PSI System included Mechanical Integrity (post-processing), Debris Generation, and Inter-Designer Variability analysis. The Salto Talaris Ankle PSI System met all acceptance criteria for all performance tests. Non-clinical cadaveric comparison testing was performed to compare Implant Alignment Accuracy and Guide Usability between the subject device and reference device. Accuracy and functionality were shown to be similar to that of the standard instrumentation used for the Salto Talaris Total Ankle System. ### 11. CONCLUSION Based on a comparison of the intended use, indications for use, and technological characteristics, the Salto Talaris Ankle PSI System is substantially equivalent to the identified predicate device. Additionally, the non-clinical testing supports that the system performs in accordance with its intended use and is as safe, as effective, and performs as well as the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...