← Product Code [OWY](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/OWY) · K212306

# Tapestry Biointegrative Implant (K212306)

_Embody, Inc. · OWY · Nov 2, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OWY/K212306

## Device Facts

- **Applicant:** Embody, Inc.
- **Product Code:** [OWY](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/OWY.md)
- **Decision Date:** Nov 2, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

TAPESTRY® Biointegrative Implant is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.

## Device Story

Tapestry Biointegrative Implant is a collagen and poly(D,L-lactide) scaffold; functions as a non-constricting, protective layer between tendon and surrounding tissues. Used in surgical settings by surgeons to manage tendon injuries. Implant is conformable; placed between tendon and surrounding tissue; secured via standard fixation techniques. May be hydrated with sterile isotonic solution. Optional co-packaged polyethylene insertion sleeve facilitates application; discarded after use. Provides a scaffold for dense collagenous fibrous connective tissue ingrowth. Device is sterile, non-pyrogenic, and single-use.

## Clinical Evidence

Bench testing only. Performance testing included hydroxyproline analysis for collagen quantity and Fourier Transform Infrared (FTIR) analysis for collagen quality to validate shelf-life stability. Previous safety, biocompatibility, and bench testing from the predicate device (K201572) were deemed directly applicable.

## Technological Characteristics

Materials: Collagen and poly(D,L-lactide). Form factor: Conformable sheet, various sizes (20mm x 25mm to 70mm x 50mm). Sterilization: Sterile. Connectivity: None. Software: None.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- TAPESTRY® Biointegrative Implant ([K201572](/device/K201572.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.

November 2, 2021

Embody, Inc. Brianna Schehr Director of Regulatory Affairs 4211 Monarch Way, Suite 500 Norfolk, Virginia 23508

Re: K212306

Trade/Device Name: Tapestry Biointegrative Implant Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: OWY Dated: September 29, 2021 Received: September 30, 2021

Dear Ms. Schehr:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Deborah Fellhauer, RN, BSN Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K212306

Device Name TAPESTRY® Biointegrative Implant

Indications for Use (Describe)

TAPESTRY® Biointegrative Implant is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

| Device Trade Name:   | TAPESTRY® Biointegrative Implant                                                                               |
|----------------------|----------------------------------------------------------------------------------------------------------------|
| Device Common Name:  | Tendon Protector                                                                                               |
| Device Class:        | II                                                                                                             |
| Classification Name: | Mesh, Surgical                                                                                                 |
| Regulation No.:      | 878.3300                                                                                                       |
| Product Code:        | OWY                                                                                                            |
| Predicate Device:    | TAPESTRY® Biointegrative Implant, K201572                                                                      |
| Owner/Submitter:     | Embody, Inc.                                                                                                   |
|                      | 4211 Monarch Way, Suite 500<br>Norfolk, VA 23508<br>(757) 777-5674                                             |
| Regulatory Contact:  | Brianna Schehr<br>Director of Regulatory Affairs<br>Email: bschehr@embody-inc.com<br>Telephone: (236) 994-7912 |
| Date Prepared:       | September 29, 2021                                                                                             |

#### DEVICE DESCRIPTION

The TAPESTRY® Biointegrative Implant (TAPESTRY) is composed of collagen and poly(D,Llactide). It is designed to function as a non-constricting, protective layer between the tendon and surrounding tissues. TAPESTRY is conformable and designed for easy placement between the tendon and surrounding tissue and may be secured in place using standard fixation techniques. TAPESTRY is provided sterile, non-pyrogenic, for single-use only, in a variety of sizes, ranging from 20mm x 25mm to 70mm x 50mm. TAPESTRY is available with or without a co-packaged polyethylene Insertion Sleeve, which is used to maintain the implant's orientation and to facilitate easy application onto the tendon. The Insertion Sleeve is discarded after use and not implanted. TAPESTRY is designed for stand-alone use. At the discretion of the surgeon. TAPESTRY may be hydrated with sterile isotonic solution.

Preclinical studies of TAPESTRY showed dense collagenous fibrous connective tissue ingrowth into and around the scaffolding.

## INTENDED USE/INDICATIONS FOR USE

TAPESTRY is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.

{4}------------------------------------------------

## SUBSTANTIAL EQUIVALENCE SUMMARY

TAPESTRY is the same device as its TAPESTRY predicate, with the same indications for use, design, materials, and technological characteristics and, therefore is substantially equivalent. Validation methods for establishing collagen stability throughout TAPESTRY's shelf life has been added to support expansion of the shelf life of the subject device. Two new product size/shape offerings, within the originally cleared size range, have been added based on user feedback.

## SUMMARY OF PERFORMANCE TESTING

TAPESTRY is the same device as its predicate, TAPESTRY (K201572), in regard to all aspects of the device. Only the collagen stability validation method changed between the current device and the predicate device. Past performance testing (safety, biocompatibility, bench) of the predicate device is directly applicable to the subject device. Additional tests to confirm the collagen stability of the device were performed. Hydroxyproline analysis was conducted to confirm the collagen quantity and Fourier Transform Infrared (FTIR) analysis was conducted to confirm collagen quality. Together with the product stability testing of the predicate device, the collagen stability testing will be used to support expansion of the shelf life of the subject device. No new performance testing was required to support the additional product sizes, as the sizes are within the cleared size range and the cleared labeling allows for trimming of the device. The performance testing demonstrates that TAPESTRY is substantially equivalent to the predicate device.

### CONCLUSION

The purpose of this 510(k) application was to notify the Food and Drug Administration of the new validation methods for establishing collagen stability throughout TAPESTRY's shelf life. TAPESTRY is the same device as the TAPESTRY device cleared under K201572 with the same intended use, design, and technological characteristics. The new collagen stability validation methods do not raise any different questions of safety or efficacy. Therefore, the TAPESTRY subject device is substantially equivalent to the TAPESTRY predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OWY/K212306](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OWY/K212306)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
