← Product Code [OWW](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/OWW) · K240817

# VersaWrap (K240817)

_Alafair Biosciences, Inc. · OWW · Jun 14, 2024 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OWW/K240817

## Device Facts

- **Applicant:** Alafair Biosciences, Inc.
- **Product Code:** [OWW](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/OWW.md)
- **Decision Date:** Jun 14, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

VersaWrap is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue. The device may also be used in the management and protection of surrounding tissues such as skeletal muscle and ligaments. In these procedures, VersaWrap may encounter a variety of implanted structures such as anchors, grafts, staples, and sutures.

## Device Story

VersaWrap is an absorbable surgical implant designed to provide a non-constricting, protective encasement for injured tendons, skeletal muscle, and ligaments. The device consists of a clear hydrogel sheet (alginate and hyaluronic acid) and an aqueous citrate wetting solution. During surgery, the clinician applies the solution to the sheet to create a tissue-adherent gel, which is then placed under, around, or over the target tissue. The device is intended to interface with surrounding tissues and implanted structures (anchors, grafts, staples, sutures) to facilitate management and protection. It is supplied sterile for single-use. The device does not alter the surgical procedure but provides a physical barrier to protect the injury site.

## Clinical Evidence

No clinical testing was required. Substantial equivalence is supported by prior non-clinical and animal testing previously conducted for the device.

## Technological Characteristics

Absorbable implant consisting of an alginate and hyaluronic acid hydrogel sheet and an aqueous citrate wetting solution. Supplied sterile, non-pyrogenic, for single use in double peel pouches. Mechanical barrier/encasement function.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- VersaWrap® ([K213163](/device/K213163.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the date June 14, 2024. The month is June, the day is the 14th, and the year is 2024. The text is in a simple, sans-serif font. The date is likely used for a calendar, invitation, or other time-sensitive document.

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized depiction of a human figure. To the right of it is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Alafair Biosciences Inc. Dr. Sarah Mayes Official Correspondent 6101 W Courtvard Dr Ste 1-225 Austin, Texas 78730

Re: K240817

Trade/Device Name: VersaWrap® Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: OWW Dated: April 23, 2024 Received: April 24, 2024

Dear Dr. Mayes:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

 Robert M. Stefani -S

Digitally signed by Robert M Stefani -S Date: 2024.06.14 17:19:00 -04'00'

For: Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K240817

Device Name

VersaWrap®

Indications for Use (Describe)

VersaWrap is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue. The device may also be used in the management and protection of surrounding tissues such as skeletal muscle and ligaments.

In these procedures, VersaWrap may encounter a variety of implanted structures such as anchors, grafts, staples, and sutures.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

| Contact Details                   |                                                                                                                                                                            |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted by:                     | Alafair Biosciences, Inc.<br>6101 W Courtyard Drive Ste. 1-225<br>Austin, TX 78730                                                                                         |
| Contact:                          | Dr. Sarah Mayes<br>Alafair Biosciences Inc.<br>6101 W Courtyard Dr Ste 1-225<br>Austin TX 78730 United States<br>Phone: 512-739-9510<br>Email: info@alafairbiosciences.com |
| Device Name                       |                                                                                                                                                                            |
| Product Name                      | VersaWrap®                                                                                                                                                                 |
| Common Name                       | Surgical Mesh                                                                                                                                                              |
| Classification number             | 21 CFR 878.3300                                                                                                                                                            |
| Classification name               | Mesh, Surgical, Absorbable, Orthopaedics, Reinforcement of Tendon                                                                                                          |
| Product Code                      | OWW                                                                                                                                                                        |
| Legally Marketed Predicate Device |                                                                                                                                                                            |

VersaWrap® (K213163)

#### Device Description Summary

VersaWrap® is an absorbable implant (device), designed to serve as an interface between the target tissues to provide a non-constricting, protective encasement. VersaWrap consists of a clear hydrogel (Sheet) and a wetting Solution. The clear Sheet is a thin membrane of alginate and hyaluronic acid. VersaWrap Sheet is easy to handle, and is designed for placement under, around, or over injured tissues and/or surrounding tissues. VersaWrap may encounter a variety of implanted structures such as anchors, grafts, staples. VersaWrap Sheet is supplied sterile, non-pyrogenic, for single use, in double peel pouches. The VersaWrap Solution is applied to the Sheet a gelatinous, tissue adherent layer (gel). The Solution, comprised of aqueous citrate, is provided sterile, non-pyrogenic, for single use, in a double peel pouch.

Intended Use/Indications for Use

VersaWrap is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue. The device may also be used in the management and protection of surrounding tissues such as skeletal muscle and ligaments. In these procedures, VersaWrap may encounter a variety of implanted structures such as anchors, grafts, staples,

Indications for Use Comparison

The additional description in the labeling is further identifying the surrounding environment and is not intended therapeutic, diagnostic, prosthetic, or surgical use of the difference does not affect the safety and effectiveness of the device when used as labeled, and describe a new disease, condition, or patient population. The device is not being used differently. The risk-based assessment did not identify any new risks or significantly modified existing risks.

Therefore, the subject device can be considered to be substantially equivalent to the predicate device

Technology Comparison

The subject device has the identical technological characteristics as the predicate device.

Non-Clinical and/or Clinical Tests Summary & Conclusions

Based on the risk-based assessment no additional testing is required to support the additional determination of substantial equivalence. The prior non-clinical and animal testing for the device were used to support the extended indications.

Based on the risk-based assessment no clinical testing is required to support the additional determination of substantial equivalence.

## 510(k) Summary

Prepared on: June 10, 2024

{5}------------------------------------------------

The subject device can be considered to be substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OWW/K240817](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OWW/K240817)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
