← Product Code [OWI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OWI) · K201240

# Mariner Cap System (K201240)

_SeaSpine Orthopedics Corporation · OWI · Aug 5, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OWI/K201240

## Device Facts

- **Applicant:** SeaSpine Orthopedics Corporation
- **Product Code:** [OWI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OWI.md)
- **Decision Date:** Aug 5, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3010
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Mariner Cap System is intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures. The indications for use include the following applications: - 1. Spinal trauma surgery, used in sublaminar or facet wiring techniques; - 2. Spinal reconstructive surgery, incorporated into constructs for the purpose of correction of spinal deformities such as adolescent idiopathic scoliosis, adult scoliosis, kyphosis and spondylolisthesis; - 3. Spinal degenerative surgery, as an adjunct to spinal fusions. The Mariner Cap System shall be used in conjunction with the SeaSpine Mariner Pedicle Screw System whenever "wiring" may help secure the attachment of other implants.

## Device Story

Mariner Cap System provides temporary bone stabilization and fracture repair aid; functions as bone anchor for spinal surgery. Comprises three components: Mariner Cap SP Connector, Mariner Cap SP Locking Base, and Locking Insert. Attaches to proximal end of SeaSpine Mariner Pedicle Screw System; enables attachment of JAZZ Band/JAZZ Passer Band to skeletal structures. Used in spinal trauma, reconstructive, and degenerative surgeries; operated by surgeons in clinical/OR settings. Provides lower-profile method for braid attachment to fixation constructs. Benefits patient by securing implants during bony fusion development.

## Clinical Evidence

No clinical data. Substantial equivalence is based on non-clinical bench testing, including dynamic and static compression testing performed in accordance with ASTM 1717 standards.

## Technological Characteristics

System consists of three components: Mariner Cap SP Connector, Mariner Cap SP Locking Base, and Locking Insert. Designed for attachment to Mariner Pedicle Screw System. Locking mechanism for braid component is identical to K182771. Mechanical performance validated via dynamic and static compression testing per ASTM 1717.

## Regulatory Identification

A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.

## Predicate Devices

- ISS-JAZZ Screw System and JAZZ CAP SP ([K182771](/device/K182771.md))

## Reference Devices

- Mariner Pedicle Screw System ([K173882](/device/K173882.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

August 5, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it. The words "U.S. FOOD & DRUG" are on the first line, and "ADMINISTRATION" is on the second line.

SeaSpine Orthopedics Corporation Alicia McArthur Specialist, Regulatory Affairs 5770 Armada Drive Carlsbad, California 92008

Re: K201240

Trade/Device Name: Mariner Cap System Regulation Number: 21 CFR 888.3010 Regulation Name: Bone Fixation Cerclage Regulatory Class: Class II Product Code: OWI Dated: May 6, 2020 Received: May 8, 2020

#### Dear Alicia McArthur:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'Neill, M.B.E. Acting Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K201240

Device Name Mariner Cap System

Indications for Use (Describe)

The Mariner Cap System is intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures.

The indications for use include the following applications:

- 1. Spinal trauma surgery, used in sublaminar or facet wiring techniques;
- 2. Spinal reconstructive surgery, incorporated into constructs for the purpose of correction of spinal deformities such as adolescent idiopathic scoliosis, adult scoliosis, kyphosis and spondylolisthesis;
- 3. Spinal degenerative surgery, as an adjunct to spinal fusions.

The Mariner Cap System shall be used in conjunction with the SeaSpine Mariner Pedicle Screw System whenever "wiring" may help secure the attachment of other implants.

| Type of Use (Select <i>one</i> or <i>both</i> , as applicable) |
|----------------------------------------------------------------|
|----------------------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Summary

### Contact Details

| Applicant Name:      | SeaSpine® Orthopedics Corporation              |
|----------------------|------------------------------------------------|
| Address:             | 5770 Armada Drive, Carlsbad, CA 92008          |
| Phone number:        | (760) 216-5176                                 |
| Fax number:          | (760) 683-6874                                 |
| Primary Contact:     | Alicia McArthur, Regulatory Affairs Specialist |
| Secondary Contact:   | Jesse Albright, Regulatory Affairs Specialist  |
| Date Prepared:       | July 29, 2020                                  |
| Device Name          |                                                |
| Trade Name:          | Mariner Cap System                             |
| Common Name:         | Bone Fixation Cerclage, Sublaminar             |
| Classification Name: | Bone Fixation Cerclage (21 CFR 888.3010)       |

OWI Product Code:

Class:

# Legally Marketed Predicate Devices

II

| 510(k) Number            | Product Code | Trade Name                                  | Manufacturer                        |
|--------------------------|--------------|---------------------------------------------|-------------------------------------|
| PRIMARY PREDICATE Device |              |                                             |                                     |
| K182771                  | NKB, OWI     | ISS-JAZZ Screw<br>System and JAZZ CAP<br>SP | Implanet                            |
| REFERENCE Device         |              |                                             |                                     |
| K173882                  | NKB          | Mariner Pedicle Screw<br>System             | SeaSpine Orthopedics<br>Corporation |

# Device Description

The Mariner Cap is composed of three components: the Mariner Cap SP Connector, the Mariner Cap SP Locking Base, and the Locking Insert. The cap is designed to mate with the Mariner Pedicle Screw System screws. The Mariner Cap is added to the proximal end of the Mariner Screw and provides for an alternate method to attach the JAZZ Band and JAZZ Passer Band to skeletal structures as compared to the current method using a JAZZ connector (K171881).

{4}------------------------------------------------

#### Indications for Use

The Mariner Cap System is intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures.

The indications for use include the following applications:

- 1. Spinal trauma surgery, used in sublaminar or facet wiring techniques;
- 2. Spinal reconstructive surgery, incorporated into constructs for the purpose of correction of spinal deformities such as adolescent idiopathic scoliosis, adult scoliosis, kyphosis and spondylolisthesis:
- 3. Spinal degenerative surgery, as an adjunct to spinal fusions.

The Mariner Cap System shall be used in conjunction with the SeaSpine Mariner Pedicle Screw System whenever "wiring" may help secure the attachment of other implants.

# Summary of Technological Characteristics

The Mariner Cap System provides an alternate way to allow for braid attachment to the spine. The Mariner Cap System when used with the Mariner Pedicle Screw System, provides a more of ficient, lower profile method of adding the braid attachment to the fixation construct by attaching the JAZZ Passer Band to the Mariner Cap designed to attach to the proximal end of the screw. The locking mechanism used for the braid component in the Mariner Cap is identical to that already cleared in the JAZZ CAP (K182771).

The Mariner Cap System was shown to be substantially equivalent and have equivalent technological characteristics to the cited predicate and reference device in regard to components, device description, intended use/indications for use, technological characteristics (i.e., operating principle, design, materials, sterility, manufacturing, etc.) and performance (i.e., mechanical testing).

# Non-Clinical Testing

The Mariner Cap System demonstrated similar performance to the predicate system through dynamic compression testing with reference to ASTM 1717 and static compression testing.

# Clinical Testing

Not applicable; determination of substantial equivalence is not based on an assessment of clinical performance data.

# Conclusions

The submitted data demonstrates that the Mariner Cap System has been shown to be substantially equivalent to the legally marketed predicate device for its intended use.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OWI/K201240](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OWI/K201240)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
