KMTI Tesera SA Anterior Lumbar Interbody Fusion (ALIF) System
K193320 · Kyocera Medical Technologies, Inc. · OVD · Feb 4, 2020 · Orthopedic
Device Facts
Record ID
K193320
Device Name
KMTI Tesera SA Anterior Lumbar Interbody Fusion (ALIF) System
Applicant
Kyocera Medical Technologies, Inc.
Product Code
OVD · Orthopedic
Decision Date
Feb 4, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The KMTI Tesera SA Anterior Lumbar Interbody Fusion (ALIF) System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). KMTI Tesera SA ALIF System implants are to be used with autogenous bone graft. Patients should be skeletally mature and have at least six months of non-operative treatment. The KMTI Tesera SA ALIF System is a standalone device and is intended to be used with the cover plate and screws provided and requires no additional supplementary fixation. The anterior cover plate must be utilized whenever the device is implanted using the bone screws provided. Should the physician choose to use fewer than the four screws provided, additional supplemental fixation cleared by the FDA for use in the lumbar spine must be used. Supplemental fixation, cleared by the FDA for use in the lumbosacral spine, must be used with implants ≥20°.
Device Story
System consists of additively manufactured Ti6Al4V intervertebral cages, cover plates, and screws; used for anterior lumbar interbody fusion (ALIF). Cages implanted in lumbar spine (L2-S1) to facilitate fusion with autogenous bone graft. Standalone use permitted with provided cover plate and screws; supplemental fixation required for implants ≥20° or if fewer than four screws used. Operated by surgeons in clinical/OR settings. Provides structural support to intervertebral space; promotes fusion. Benefits patients by stabilizing spinal segments and addressing discogenic pain.
Clinical Evidence
Bench testing only. Performance validated via dynamic shear compression testing per ASTM F2077-18 and tension testing per ASTM E8/E8M-16.
Technological Characteristics
Ti6Al4V cages manufactured via additive manufacturing (powder bed fusion) compliant with ASTM F2924. Stainless steel instruments compliant with ASTM A564. Gamma sterilized. Standalone system with cover plate and screws.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine (L2-S1) at one or two contiguous levels. Includes patients with up to Grade 1 spondylolisthesis or retrolisthesis who have failed at least six months of non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Renovis S128 Anterior Lumbar Interbody Fusion (ALIF) System (K131122)
Renovis S128 Anterior Lumbar Interbody Fusion (ALIF) System (K140106)
Renovis S128 Anterior Lumbar Interbody Fusion (ALIF) System (K180502)
Reference Devices
Kyocera S180 Lateral Lumbar Interbody Fusion (LLIF) System (K181655)
Related Devices
K201671 — A-Link Z · Acuity Surgical Devices, LLC · May 4, 2021
K230872 — Halis Lumbar Cage System, Lydia Anterior Lumbar Fusion System, Dica Direction Changeable Lumbar Cage System, KEYSTONE Cage System · Shanghai Sanyou Medical Co, Ltd. · Dec 12, 2023
K212980 — Tesera-k ALIF System · Kyocera Medical Technologies, Inc. · Jan 7, 2022
K180556 — Corelink Foundation 3D Anterior Lumbar System · Corelink, LLC · Jul 27, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
February 4, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Kyocera Medical Technologies, Inc. % Sharyn Orton Senior Consultant MEDIcept, Inc. 200 Homer Ave Ashland, Massachusetts 01721
## Re: K193320
Trade/Device Name: KMTI Tesera SA Anterior Lumbar Interbody Fusion (ALIF) System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVD Dated: December 17, 2019 Received: December 18, 2019
Dear Dr. Orton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Brent L. Showalter, Ph.D. Assistant Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K193320
#### Device Name
KMTI Tesera SA Anterior Lumbar Interbody Fusion (ALIF) System
#### Indications for Use (Describe)
The KMTI Tesera SA Anterior Lumbar Interbody Fusion (ALIF) System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). KMTI Tesera SA ALIF System implants are to be used with autogenous bone graft. Patients should be skeletally mature and have at least six months of non-operative treatment. The KMTI Tesera SA ALIF System is a standalone device and is intended to be used with the cover plate and screws provided and requires no additional supplementary fixation. The anterior cover plate must be utilized whenever the device is implanted using the bone screws provided. Should the physician choose to use fewer than the four screws provided, additional supplemental fixation cleared by the FDA for use in the lumbar spine must be used. Supplemental fixation, cleared by the FDA for use in the lumbosacral spine, must be used with implants ≥20°.
Type of Use (Select one or both, as applicable)
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# rg Kyocera
# Traditional 510(k) Premarket Notification Summary as required by 21 CFR 807.92(a) K193320
| A ) Submitted by: | Kyocera Medical Technologies, Inc.<br>1200 California St. Suite 210<br>Redlands, CA 92374<br>Phone: 909-557-2360<br>Fax: 909-839-6269 |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | Anthony DeBenedictis<br>Divisional Vice President of Quality Assurance |
| Consultant: | Sharyn Orton, Ph.D.<br>MEDIcept, Inc.<br>200 Homer Ave<br>Ashland, MA 01721 |
| B) Classification Name: | Intervertebral Fusion Device With Bone Graft, Lumbar |
| Common Name: | Intervertebral body fusion device |
| Proprietary Name: | KMTI Tesera SA Anterior Lumbar Interbody Fusion System |
| Device Class: | Class II |
| Regulations<br>and Product Code: | 21 CFR 888.3080<br>OVD |
| Classification panel: | Orthopedic |
| C) Predicates: | Primary: K131122 Renovis S128 Anterior Lumbar Interbody<br>Fusion (ALIF) System<br>Secondary: K140106 Renovis S128 Anterior Lumbar Interbody<br>Fusion (ALIF) System |
Fusion (ALIF) System
Secondary: K180502 Renovis S128 Anterior Lumbar Interbody
{4}------------------------------------------------
January 31, 2020 D) Date Prepared:
## E) Device Description:
The Kyocera Medical Technologies, Inc. Tesera SA Anterior Lumbar Interbody Fusion (ALIF) System is a change to the FDA cleared Kyocera S128 Anterior Lumbar Interbody Fusion (ALIF) System (Ti6Al4V and PEEK cages, Cover Plate, screws and instruments; K131122. K140106. and K180502). Only the Ti6Al4V cages are relevant to this application which describes a change in manufacturing and material standard. There are no other changes in cage dimensions (L, W, H and lordosis), sterilization, packaging or shelf-life. The new system will be marketed as the Tesera SA Anterior Lumbar Interbody Fusion (ALIF) System ("Tesera SA ALIF System") and will include Ti6Al4V cages only.
The Tesera SA ALIF System includes cages of a variety of lengths, widths, and lordosis to suit the individual pathology and anatomical conditions of the patient. The different shape of the footprint allows for different surgical approaches for insertion. Tesera SA ALIF System cages are standalone when implanted with the Cover Plate and screws with the exception of implants ≥20° which require supplemental fixation.
- The Ti6Al4V is compliant with ASTM F2924 Standard Specification for Additive Manufacturing Titanium-6 Aluminum-4 Vanadium with Powder Bed Fusion
The Tesera SA ALIF System cages that are the subject of this application are gamma sterilized.
The system also includes the previously FDA cleared Cover Plate and screws; previously FDA cleared and new instruments to allow for implant determination, trialing and disc preparation. New instruments are manufactured from stainless steels compliant with ASTM A564 Standard Specification for Hot-Rolled and Cold-Finished Age-Hardening Stainless Steel Bars and Shapes.
- F) Intended Use/Indications For Use:
The KMTI Tesera SA Anterior Lumbar Interbody Fusion (ALIF) System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). KMTI Tesera SA ALIF System implants are to be used with autogenous bone graft. Patients should be skeletally mature and have at least six months of non-operative treatment. The KMTI Tesera SA ALIF System is a stand-alone device and is intended to be used with the cover plate and screws provided and requires no additional supplementary fixation. The anterior cover plate must be utilized
{5}------------------------------------------------
whenever the device is implanted using the bone screws provided. Should the physician choose to use fewer than the four screws provided, additional supplemental fixation cleared by the FDA for use in the lumbar spine must be used. Supplemental fixation, cleared by the FDA for use in the lumbosacral spine, must be used with implants ≥20°.
G) Substantial Equivalence Comparison and Discussion
The Tesera SA ALIF System has the same Indications for Use as the predicate devices. The cages are additively manufactured; have the same dimensions and sizes, are packaged and gamma sterilized the same as the predicate cages. The new cages are manufactured to ASTM F2924, which is appropriate for additive manufacturing. The Tesera SA ALIF System cages are additively manufactured using the same technology and material standard as the FDA cleared K181655 Kyocera S180 Lateral Lumbar Interbody Fusion (LLIF) System cages (reference device). The manufacturing change has been validated.
H) Performance – Bench
The new changes were assessed for risk and successful dynamic shear compression testing conducted.
- Compliance with Standards or FDA Guidance I)
The Tesera SA ALIF System complies with the following:
- ASTM F2077-18 Test Methods For Intervertebral Body Fusion Devices .
- . ASTM E8/E8M-16 Standard Test Methods for Tension Testing of Metallic Materials
- . Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Intervertebral Body Fusion Device, June 2007
## Conclusion
The change in manufacturing and material standard do not result in new or different issues of safety or effectiveness associated with the new cages.
The Tesera SA ALIF System is substantially equivalent to the predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.