Endoskeleton(R) TAS Interbody Fusion Device / Endoskeleton(R) TAS Hyperlordotic Interbody Fusion Device
Applicant
Titan Spine, LLC
Product Code
OVD · Orthopedic
Decision Date
Apr 13, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Endoskeleton® TAS Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolistiss at the involved level(s). Patients should have received 6 months of non-operative treatment with the devices. The device is a standalone system that is intended to be used with the bone screws provided and requires no additional supplementary fixation. The device is indicated to be used with autograft bone. Hyperlordotic Devices ≥16° The Endoskeleton® TAS Hyperlordotic Interbody Fusion Device(≥16°) is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients should have received 6 months of non to treatment with the devices. The device is indicated to be used with autograft bone. The Endosketon® TAS Hyperlordotic Interbody Fusion Device must be used with a posterior supplemental internal spinal fixation that has been cleared by the FDA for use in the lumbar spine.
Device Story
Interbody fusion device for ALIF procedures; titanium alloy (Ti6Al4V-ELI) implant with hollow center for autograft bone; features Chemtex or nanoLOCK surface treatments for bone fixation; includes integrated bone screws. Standard TAS device functions as standalone system; Hyperlordotic versions (≥16°) require posterior supplemental internal fixation. Implant placed in interbody space by surgeon to facilitate fusion; windows allow radiographic visualization of bone graft and new bone formation. Benefits include structural support and fusion at disc space.
Clinical Evidence
Bench testing only. Mechanical testing performed per ASTM F2077 and ASTM 2267 (static/dynamic compression, shear, subsidence, expulsion). Additional evidence includes cadaver study and radiographic/CAD analysis.
Technological Characteristics
Medical grade titanium alloy (Ti6Al4V-ELI). Surface treatments: Chemtex or nanoLOCK (microscopic roughened surface with nano-scale features). Mechanical testing per ASTM F2077 and ASTM 2267. Integrated fixation via bone screws. Form factor: ALIF interbody spacer with hollow center and windows.
Indications for Use
Indicated for skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1, with or without Grade I spondylolisthesis or retrolisthesis, who have failed 6 months of non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
NuVasive Inc. Brigade Hyperlordotic System (K123045)
K232097 — IdentiTi ALIF Interbody Systems · Alphatec Spine, Inc. · Sep 25, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 13, 2017
Titan Spine, LLC % Christine Scifert Managing Partner Memphis Regulatory Consulting (MRC-X), LLC 3416 Roxee Run Cove Bartlett, Tennessee 38133
Re: K163269
Trade/Device Name: Endoskeleton® TAS Interbody Fusion Device / Endoskeleton® TAS Hyperlordotic Interbody Fusion Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVD Dated: March 15, 2017 Received: March 20, 2017
Dear Ms. Scifert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# VincentJ. Devlin -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
## K163269
#### Device Name
Endoskeleton® TAS Interbody Fusion Device / Endoskeleton® TAS Hyperlordotic Interbody Fusion Device
## Indications for Use (Describe)
The Endoskeleton® TAS Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolistiss at the involved level(s). Patients should have received 6 months of non-operative treatment with the devices. The device is a standalone system that is intended to be used with the bone screws provided and requires no additional supplementary fixation. The device is indicated to be used with autograft bone.
Hyperlordotic Devices ≥16°
The Endoskeleton® TAS Hyperlordotic Interbody Fusion Device(≥16°) is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients should have received 6 months of non to treatment with the devices. The device is indicated to be used with autograft bone. The Endosketon® TAS Hyperlordotic Interbody Fusion Device must be used with a posterior supplemental internal spinal fixation that has been cleared by the FDA for use in the lumbar spine.
Type of Use (Select one or both, as applicable)|×| Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
> K163269 Page 1 of 1
{3}------------------------------------------------
# 510(k) Summary
Traditional 510(k) Endoskeleton® TAS System March 16, 2017
| Company: | Titan Spine, LLC<br>6140 West Executive Drive, Suite A<br>Mequon, WI 53092, USA |
|--------------------------------|--------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration: | 3006340236 |
| Primary Contact: | Christine Scifert<br>Phone: 901-831-8053 |
| Company Contact: | Jane Rodd<br>Phone: 866-822-7800<br>Fax: 262-242-7802 |
| Trade Name: | Endoskeleton® TAS Interbody Fusion Device / Endoskeleton® TAS Hyperlordotic<br>Interbody Fusion Device |
| Common Name: | Intervertebral Body Fusion Device with Bone Graft, Lumbar |
| Classification: | Class II |
| Regulation Number: | 21 CFR 888.3080 (Intervertebral body fusion device) |
| Panel: | 87- Orthopedic |
| Product Code: | OVD |
| Predicate Devices: | Primary Predicate<br>NuVasive Inc. Brigade Hyperlordotic System (K123045 S.E. June 16, 2013) |
| | Additional Predicates |
| | Titan Spine Endoskeleton® TAS (K111626 S.E. September 9, 2011) |
| | Titan Spine Endoskeleton® TA (K080615 S.E. June 17, 2008) |
| | Titan Spine nanoLOCK® Surface Technology (K141953 S.E. October 27, 2014) |
| | Centinel Spine/Surgicraft STALIF TT (K073109 S.E. June 4, 2008) |
## Device Description:
The Endoskeleton® TAS Interbody Fusion Device and Endoskeleton® TAS Hyperlordotic Interbody Fusion Device implants are available in a variety of Anterior Lumbar Interbody Fusion (ALIF) sizes with a variety of lordotic angles, to accommodate patient anatomy; Hyperlordotic implants are those defined by a lordotic angle ≥ 16'. Implants are designed with a large hollow region in the center to house autograft bone material. The new bone formation through the implant is intended to provide long-term structural support and fusion at the implanted disc space. The design incorporates "windows" through the implant
{4}------------------------------------------------
to permit visualization of the graft material and over time formation of new bone. The superior and inferior surfaces include either the previously cleared Chemtex® surface treatment or previously cleared nanoLOCK® surface treatment to improve fixation to the adjacent bone. The nanoLOCK® surface technology provides a microscopic roughened surface with nano-scale features. The implant system includes integrated fixation (screws) for stabilizing the implants when placed in the interbody space.
An implant holding feature has been incorporated into the trailing surface of the implant to mate with the implant holder, and to facilitate placement of the interbody space. Screws include internal hex drive features matched to instrumentation. All implantable components are manufactured from medical grade titanium alloy (Ti6A14V-ELI).
# Indication for Use:
The Endoskeleton® TAS Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients should have received 6 months of non-operative treatment prior to treatment with the device is a standalone system that is intended to be used with the bone screws provided and requires no additional supplementary fixation. The device is indicated to be used with autograft bone.
# Hyperlordotic Devices ≥16°:
The Endoskeleton® TAS Hyperlordotic Interbody Fusion Device(≥ 16°) is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients should have received 6 months of non-operative treatment prior to treatment with the device is indicated to be used with autograft bone. The Endoskeleton® TAS Hyperlordotic Interbody Fusion Device must be used with a posterior supplemental internal spinal fixation that has been cleared by the FDA for use in the lumbar spine.
# Substantial Equivalence:
The Endoskeleton®TAS Hyperlordotic IBD is substantially equivalent to the NuVasive Inc. Brigade Hyperlordotic System (K123045 S.E. June 16, 2013), Titan Spine's previously cleared Endoskeleton®TAS (K111626 S.E. September 9, 2011& K141953 S.E. October 27, 2014), Titan Spine Endoskeleton® TA (K080615 S.E. June 17, 2008) and Centinel Spine/Surgicraft STALIF TT (K073109 S.E. June 4, 2008).
## Performance Testing:
Mechanical testing, including static compression shear, dynamic compression, dynamic compression-shear, subsidence and expulsion have been performed per ASTM F2077 and ASTM 2267 on the subject Endoskeleton® TAS Hyperlordotic IBD and the predicate devices. In addition, a cadaver study was performed and radiographic and CAD analysis was performed.
The resulting data from testing demonstrates that the Endoskeleton® TAS and nanoLOCK® TAS Hyperlordotic Interbody Fusion Devices are substantially equivalent to the predicate interbody devices identified in the substantial equivalence section.
{5}------------------------------------------------
# Conclusion:
The subject submission introduces a Hyperlordotic IBD to the Endoskeleton® TAS Interbody Fusion System. The Indications for Use, materials and geometry of the subject device is similar to the predicate devices. Mechanical testing demonstrates that the Endoskeleton® TAS Hyperlordotic IBD is substantially equivalently compared to predicate devices. Thus, it can be concluded that the subject device does not raise new questions about safety and effectiveness.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.