← Product Code [OVD](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVD) · K093629

# SOLITAIRE AND SOLITAIRE PEEK ANTERIOR SPINAL SYSTEM (K093629)

_Biomet Spine · OVD · Mar 9, 2010 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVD/K093629

## Device Facts

- **Applicant:** Biomet Spine
- **Product Code:** [OVD](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVD.md)
- **Decision Date:** Mar 9, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Solitaire™ and Solitaire™ PEEK Anterior Spinal System is designed for use with autograft and is indicated for stand-alone intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Additionally, the Solitaire™ Anterior Spinal System is indicated for use in the thoracolumbar spine (i.e., T10 to L5) to replace a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Solitaire System is also indicated for treating fractures of the thoracic and lumbar spine. The Solitaire System is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period.

## Device Story

System comprises spinal spacers and Osseotite® screws; used for intervertebral body fusion, vertebral body replacement, and fracture stabilization. Screws fabricated from titanium alloy with acid-etched, roughened surface to enhance osseointegration and fixation strength. System implanted by surgeons in clinical/OR settings to restore spinal biomechanical integrity. Output is physical stabilization of spinal column. Benefits include improved fixation and structural support for diseased or fractured vertebrae.

## Clinical Evidence

No clinical data. Bench testing demonstrated mechanical equivalence to existing spacers. Animal testing (sheep model) showed roughened screw surface increases osseointegration and fixation strength.

## Technological Characteristics

Titanium alloy screws with acid-etched, roughened surface; PEEK spacers. Designed for stand-alone fusion or vertebral body replacement. Mechanical fixation via screw-spacer interface.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Solitaire™ PEEK ([K022143](/device/K022143.md))
- Solitaire™ PEEK ([K062810](/device/K062810.md))
- Solitaire™ PEEK ([K081501](/device/K081501.md))
- Solitaire™ PEEK ([K081395](/device/K081395.md))
- Biomet Trauma's Acid Etched Lag Screws ([K070955](/device/K070955.md))
- BioDrive® Cannulated Screw System ([K082874](/device/K082874.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for Biomet Spine. The word "BIOMET" is in a bold, sans-serif font, with each letter connected to the next. Below the word "BIOMET" is the word "SPINE" in a smaller, sans-serif font.

# 510(k) Summary

Preparation Date: November 20, 2009 Applicant/Sponsor: Biomet Spine 100 Interpace Parkway Parsippany, NJ 07054 Contact Person: Vivian Kelly, MS, RAC Phone: 973-299-9300 Fax: 973-257-0232 Trade name: Solitaire™ and Solitaire™ PEEK Anterior Spinal System with Solitaire™ Osteotite® Screws Common Name: Non-cervical spinal spacer Classification Name: Intervertebral fusion device, 21 CFR §888.3080 Spinal Intervertebral Body Fixation Orthosis, 21 CFR § 888.3060 Device Panel /Product Code: Orthopedic MAX & MQP

## Device Description:

The Solitaire™ Ossectite® Screws are used with the spacers in the Solitaire™ and Solitaire™ PEEK Anterior Spinal System. The screws are fabricated from Titanium alloy and are acid closed to create a roughened surface.

## Indications for Use:

The Solitaire™ and Solitaire™ PEEK Anterior Spinal System is designed for use with autograft and is indicated for stand-alone intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

Additionally, the Solitaire™ Anterior Spinal System is indicated for use in the thoracolumbar spine (i.e., T10 to L5) to replace a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Solitaire System is also indicated for treating fractures of the thoracic and lumbar spine. The Solitaire System is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period.

#### Summary of Technologies:

The technological characteristics such as material, design and sizing of the Solitaire™ Osseotite® Screws in the Solitaire™ and Solitaire™ PEEK Anterior Spinal System are the same as, or similar to, the predicate devices.

#### Performance Testing:

Mechanical testing demonstrates that the Solitaire™ Osseotite® Screws when used with the spacers in the Solitaire™ and Solitaire™ PEEK Anterior Spinal System are substantially equivalent to other spacers currently on the market and is adequate for its intended use. Although animal data is not necessarily indicative of human clinical outcomes, animal testing has demonstrated that the roughened surface area of the screw increases osseointegration and enhances screw fixation strength in the spine in a healthy sheep model.

Image /page/0/Picture/13 description: The image shows a simple line drawing of an abstract shape. The shape appears to be composed of several rectangular forms connected at different angles. Some of the lines are solid, while others are dashed, giving the impression of a three-dimensional object or a blueprint-like representation.

MAR - 9 2010

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for Biomet Spine. The word "BIOMET" is in a bold, sans-serif font, with the letters connected to each other. Below the word "BIOMET" is the word "SPINE" in a smaller, sans-serif font. The logo is black and white.

MAR - 9 2010

## 510(k) Summary

| Preparation Date:           | November 20, 2009                                                                                                 |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------|
| Applicant/Sponsor:          | Biomet Spine<br>100 Interpace Parkway<br>Parsippany, NJ 07054                                                     |
| Contact Person:             | Vivian Kelly, MS, RAC<br>Phone: 973-299-9300<br>Fax: 973-257-0232                                                 |
| Trade name:                 | Solitaire™ and Solitaire™ PEEK Anterior Spinal System with<br>Solitaire™ Osteotite® Screws                        |
| Common Name:                | Non-cervical spinal spacer                                                                                        |
| Classification Name:        | Intervertebral fusion device, 21 CFR §888.3080<br>Spinal Intervertebral Body Fixation Orthosis, 21 CFR § 888.3060 |
| Device Panel /Product Code: | Orthopedic MAX & MOP                                                                                              |

#### Device Description:

The Solitaire™ Osseotite Screws are used with the spacers in the Solitaire™ PEEK Anterior Spinal System. The screws are fabricated from Titanium alloy and are acid etched to create a roughened surface.

#### Indications for Use:

The Solitaire™ and Solitaire™ PEEK Anterior Spinal System is designed for use with autograft and is indicated for stand-alone intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

Additionally, the Solitaire™ Anterior Spinal System is indicated for use in the thoracolumbar spine (i.e. T10 to L5) to replace a diseased vertebral body resected or excised for the treatment of turnors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Solitaire System is also indicated for treating fractures of the thoracic and lumbar spine. The Solitaire System is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period.

#### Summary of Technologies:

The technological characteristics such as material, design and sizing of the Solitaire™ Osseotite® Screws in the Solitaire™ and Solitaire™ PEEK Anterior Spinal System are the same as, or similar to, the predicate devices.

#### Performance Testing:

Mechanical testing demonstrates that the Solitaire™ Osseotite® Screws when used with the spacers in the Solitaire™ and Solitaire™ PEEK Anterior Spinal System are substantially equivalent to other spacers currently on the market and is adequate for its intended use. Although animal data is not necessarily indicative of human clinical outcomes, animal testing has demonstrated that the roughened surface area of the screw increases osseointerration and enhances screw fixation strength in the spine in a healthy sheep model.

{2}------------------------------------------------

## Substantial Equivalence:

The Solitaire™ Osseotite® Screws when used with the spacers in the Solitaire™ PEEK Anterior Spinal System are substantially equivalent to its predicate devices with respect to intended use and indications, technological characteristics, and principles of operation and do not present any new issues of safety or effectiveness. Examples of predicate intervertebral body fusion devices and vertebral replacement devices include the Solitaire™ PEEK (K022143, K062810, K081501 & K081395) from Biomet Spine as well as Biomet Trauma's Acid Etched Lag Screws (K070955) and the BioDrive® Cannulated Screw System (K082874) as predicates for the proprietary Osseotite process.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, with three figures intertwined. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the caduceus. The overall design is simple and professional, conveying the organization's focus on health and human services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Biomet Spine % Ms. Vivian Kelly, MS, RAC Regulatory Affairs Project Manager 100 Interpace Parkway Parsippany, NJ 07054

SEP 12 2011

Re: K093629

> Trade/Device Name: Solitaire™ and Solitaire™ PEEK Anterior Spinal System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD, MOP Dated: February 24, 2010 Received: February 25, 2010

Dear Ms. Kelly:

This letter corrects our substantially equivalent letter of March 9, 2010.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other

{4}------------------------------------------------

Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

**Sincerely yours,**

Mark A. Milliken

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known):

Device Name: Solitaire™ and Solitaire™ PEEK Anterior Spinal System

Indications for Use:

The Solitaire™ and Solitaire™ PEEK Anterior Spinal System is designed for use with autograft and is indicated for stand-alone intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

Additionally, the Solitaire™ Anterior Spinal System is indicated for use in the thoracolumbar spine (i.e., T10 to L5) to replace a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Solitaire System is also indicated for treating fractures of the thoracic and lumbar spine. The Solitaire System is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

K093629 5 !! (k) Number_

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVD/K093629](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OVD/K093629)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).