PathLoc SI Joint Fusion System

K231841 · L&K BIOMED Co., Ltd. · OUR · Jul 14, 2023 · Orthopedic

Device Facts

Record IDK231841
Device NamePathLoc SI Joint Fusion System
ApplicantL&K BIOMED Co., Ltd.
Product CodeOUR · Orthopedic
Decision DateJul 14, 2023
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The PathLoc SI Joint Fusion System is intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis.

Device Story

PathLoc SI Joint Fusion System consists of metallic bone screws of varying diameters, lengths, and thread configurations. Designed for sacroiliac joint fusion; implanted into patient bone to stabilize the joint. Features include self-tapping flutes for centering and insertion; arch screws accommodate autograft insertion; locking screws may be used with or without washers. Used by surgeons in clinical settings. Device provides mechanical fixation to facilitate fusion, potentially alleviating pain and instability associated with sacroiliac joint conditions.

Clinical Evidence

No clinical data. Substantial equivalence is supported by risk assessment and finite element (FE) analysis, with mechanical performance data leveraged from predicate devices (K153656/K181600).

Technological Characteristics

Constructed from Ti-6Al-4V ELI titanium alloy (ASTM F136). System comprises bone screws with self-tapping flutes, washers, and locking mechanisms. Mechanical design and surgical approach are identical to predicate devices.

Indications for Use

Indicated for patients requiring sacroiliac joint fusion due to sacroiliac joint disruptions or degenerative sacroiliitis.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ July 14, 2023 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of the Department of Health and Human Services seal on the left, followed by the FDA acronym in a blue square. To the right of the blue square, the words "U.S. FOOD & DRUG" are stacked on top of the word "ADMINISTRATION", also in blue. The logo is clean and professional, reflecting the FDA's role in regulating and overseeing food and drug safety in the United States. L & K Biomed Co., Ltd. Katherine Kim RA #101, 201, 202 16-25, Dongbaekjungang-ro 16 beon-gil, Giheung-gu Yongin-si, Gyeonggi-do 17015 Korea, South ### Re: K231841 Trade/Device Name: PathLoc SI Joint Fusion System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: OUR Dated: June 20, 2023 Received: June 22, 2023 ### Dear Katherine Kim: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/1/Picture/6 description: The image contains the text "Colin O'neill -S" in a simple, sans-serif font. The name "Colin" is positioned above "O'neill -S", creating a stacked arrangement. In the background, there is a faint, watermark-like image of the letters "FDA", which are significantly larger than the text in the foreground. Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K231841 Device Name PathLoc SI Joint Fusion System Indications for Use (Describe) The PathLoc SI Joint Fusion System is intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <span> <span style="font-size: 20px;"> ☑ </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff [PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov) *"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."* {3}------------------------------------------------ # 510(k) SUMMARY The following 510(k) summary is being submitted as required by 21 CFR 807.92(a): ### 1. MANUFACTURER | Submitter's Name: | L&K Biomed Co., Ltd. | |------------------------|-----------------------------------------------------------------------------------------------------------| | Submitter's Address: | #101, 201, 202 16-25, Dongbaekjungang-ro 16 beon-gil<br>Giheung-gu, Yongin-si, Gyeonggi-do, 17015, Korea. | | Submitter's Telephone: | 82-2-6717-1983 | | Contact Person: | Katherine Kim<br>khkim@lnkbiomed.com | | Prepared Date: | June 20, 2023 | #### 2. DEVICE IDENTIFICATION | Device Trade Name | PathLoc SI Joint Fusion System | |--------------------------|----------------------------------------------------| | Common/Usual Name | Sacroiliac Joint Fixation, Bone Screw | | Regulation Class /Number | Class II / 21 CFR 888.3040 | | Regulation Name | Smooth or threaded metallic bone fixation fastener | | Product Code | OUR | | Classification Panel | Spinal Devices (DHT6B) | #### PREDICATE OR LEGALLY MARKETED DEVICES WHICH ARE SUBSTANTIALLY EQUIVALENT. 3. The subject devices are identical to the predicate devices in all characteristics. | Subject Device Name | 510K NO. | Trade or Proprietary or Model Name | Predicate Type | |-----------------------------------|----------|------------------------------------|----------------| | PathLoc SI Joint<br>Fusion System | K153656 | PathLoc SI Joint Fusion System | Primary | | | K181600 | PathLoc SI Joint Fusion System | Additional | | | K210035 | Zespin SI Joint Fusion System | Additional | | | K223521 | Zespin SI Joint Fusion System | Additional | The design feature, indications for use, material and manufacturing process for the subject devices are substantially equivalent to the predicate devices. #### 4. MATERIALS | PATHLOC SI Joint Fusion System | Ti-6Al-4V ELI titanium alloy (ASTM F136) | |--------------------------------|------------------------------------------| |--------------------------------|------------------------------------------| And the additional components material is the same material used in the predicate devices (K 153656/K181600). #### 5. DESCRIPTION OF THE DEVICE The PATHLOC SI Joint Fusion System consists of different diameter bone screws in various lengths and thread configurations to accommodate variations in patient anatomy. The devices are manufactured from titanium alloy per ASTM F136. - Arch Screw will be implanted in patient's bone then autograft will be inserted. • - . Locking Screw can be used with washer or can be used on its own - . Self-tapping flute centers screw for easy insertion #### 6. INDICATION FOR USE The PATHLOC SI Joint Fusion System is intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis. {4}------------------------------------------------ #### 7. PERFORMANCE TESTING ## PATHLOC SI Joint Fusion System A risk assessment, including FE analysis, was conducted to confirm that the additional components do not introduce new issues of safety or effectiveness. And the results of the risk assessment were confirmed, and that there were no problems with safety and effectiveness. None of the additional components is the worst case of the PathLoc SI Joint Fusion System and the additional components to be added through this submission do not require additional mechanical testing. Therefore, we substitute mechanical test data of PathLoc SI Joint Fusion System for additional components with the predicate device data (K153656/K181600). #### 8. SUMMARY OF TECHNOLOGY CHARACTERISTICS Subject devices are identical to the predicate devices in all (Material, Indication for use, Design, Manufacturing process, Surgical approach) characteristics. | | 510K no. | Indication for<br>use | Design | Surgical appro<br>ach | Material | |-----------------------------------|---------------------|-----------------------|--------|-----------------------|------------------------------| | Subject Device | | Same | Same | Same | Ti-6Al-4V ELI<br>(ASTM F136) | | PathLoc SI Joint<br>Fusion System | K153656/<br>K181600 | Same | Same | Same | Same | | Zespin SI Joint<br>Fusion System | K210035<br>K223521 | Same | Same | Same | Same | #### 9. SUBSTANTIAL EQUIVALENCE Subject devices are shown to be substantially equivalent to the predicate devices in indications for use, design, function and materials used. # 10. CONCLUSION The overall technology characteristics lead to the conclusion that the PathLoc SI Joint Fusion System is substantially equivalent to the predicate devices(K153656/K181600).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...