← Product Code [OUR](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OUR) · K203780

# Harvester SI: Sacroiliac Fixation System (K203780)

_Spinus International, Inc. · OUR · Mar 23, 2021 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OUR/K203780

## Device Facts

- **Applicant:** Spinus International, Inc.
- **Product Code:** [OUR](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OUR.md)
- **Decision Date:** Mar 23, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Harvester SI Fusion System is intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroilitis. This includes conditions whose symptoms began during pregnancy or in the peripartum period and have persisted postpartum for more than 6 months.

## Device Story

Harvester SI Sacroiliac Fixation System consists of threaded fusion implants designed to immobilize and fuse the ilium to the sacrum. Implants are inserted across the sacroiliac joint using specialized instruments; procedure may be unilateral or bilateral. System utilizes two or more threaded implants to achieve arthrodesis. Used in clinical settings by surgeons to treat joint disruptions and degenerative sacroiliitis. Benefits include stabilization of the sacroiliac joint to alleviate symptoms. No software or AI components involved.

## Clinical Evidence

Bench testing only. Performance data includes axial pullout (ASTM F543), torsional properties (ASTM F543), and static/dynamic cantilever bending (ASTM F2193). Results demonstrate mechanical strength sufficient for intended use.

## Technological Characteristics

Threaded fusion implants manufactured from titanium alloy (Ti-6Al-4V ELI) per ASTM F136. Mechanical fixation via threaded design. Various diameters and lengths available. No software, electronics, or energy sources.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Entasis™ Dual-Lead Sacroiliac Implant ([K152237](/device/K152237.md))
- SAMBA™ Screw System ([K121148](/device/K121148.md))
- SambaScrew 3D SI Fusion System ([K183342](/device/K183342.md))
- Life Spine SIMPACT Sacroiliac Joint Fixation System ([K201538](/device/K201538.md))
- Rialto™ SI Fusion System ([K161210](/device/K161210.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

March 23, 2021

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Spinus International, Inc. % Nathan Wright Engineer & Regulatory Specialist Empirical Testing Corp. 4628 Northpark Drive Colorado Springs. Colorado 80918

Re: K203780

Trade/Device Name: Harvester SI: Sacroiliac Fixation System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: OUR Dated: January 26, 2021 Received: January 27, 2021

Dear Nathan Wright:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and probibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br>Indications for Use | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement on last page. |
|------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| 510(k) Number (if known)                                                                       | K203780                                                                                            |
| Device Name                                                                                    | Harvester SI: Sacroiliac Fixation System                                                           |
| Indications for Use (Describe)                                                                 |                                                                                                    |

The Harvester SI Fusion System is intended for sacroiliac joint fusion for conditions including sacroillac joint disruptions and degenerative sacroillitis. This includes conditions whose symptoms began during pregnancy or in the peripartum period and have persisted postpartum for more than 6 months.

| Type of Use (Select one or both, as applicable)                                                                                                                                                                                                                  |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <span style="display:inline-block; width:15px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)             </span> <span> <span style="display:inline-block; width:15px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)             </span> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including
the time to review instructions, search existing data sources, gather and maintain the data needed and
complete and review the collection of information. Send comments regarding this burden estimate or any
other aspect of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services
   

 Food and Drug Administration
   

 Office of Chief Information Officer
   

 Paperwork Reduction Act (PRA) Staff
   

 PRAStaff@fda.hhs.gov

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information
unless it displays a currently valid OMB number."*

FORM FDA 3881 (7/17)

{3}------------------------------------------------

# 510(K) SUMMARY

| Submitter's Name:          | Spinus International, Inc.                                                               |
|----------------------------|------------------------------------------------------------------------------------------|
| Submitter's Address:       | 7200 North Interstate Highway 35<br>Building 1, Suite 101<br>Georgetown, TX 78626        |
| Submitter's Telephone:     | 254-760-4173                                                                             |
| Contact Person:            | Nathan Wright MS<br>Empirical Testing Corp.<br>719-351-0248<br>nwright@empiricaltech.com |
| Date Summary was Prepared: | December 23, 2020                                                                        |
| Trade or Proprietary Name: | Harvester SI: Sacroiliac Fixation System                                                 |
| Common or Usual Name:      | Sacroiliac Joint Fixation                                                                |
| Classification:            | Class II per 21 CFR §888.3040                                                            |
| Product Code:              | OUR                                                                                      |
| Classification Panel:      | Orthopedic                                                                               |

# DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:

The Harvester SI: Sacroiliac Fixation System (also called Harvest SI Sacroiliac Joint Fusion System and Harvester SI Fusion System) is comprised of threaded fusion implants for coupling and fusing the ilium to the sacroiliac fusion). The goal of the fusion arthrodesis is to immobilize the sacroiliac joint with two (2) or more threaded implants. The procedure may or may not be bilateral. The Harvester SI System Fusion implants are manufactured from titanium alloy (Ti-6Al-4V ELI) per ASTM F136. The Harvester SI Fusion System is offered in various diameters and lengths to accommodate patient anatomy. Instruments are provided that facilitate the insertion of the implants.

## INDICATIONS FOR USE

The Harvester SI Fusion System is intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroilitis. This includes conditions whose symptoms began during pregnancy or in the peripartum period and have persisted postpartum for more than 6 months.

# TECHNOLOGICAL CHARACTERISTICS

The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are identical between the subject and predicates:

- Indications for Use
- Materials of manufacture
- Structural support mechanism ●

{4}------------------------------------------------

- Sizes ●
- Surgical approach ●

| 510k Number | Trade or Proprietary or Model Name                  | Manufacturer            | Predicate Type |
|-------------|-----------------------------------------------------|-------------------------|----------------|
| K152237     | Entasis™ Dual-Lead Sacroiliac Implant               | CoreLink                | Primary        |
| K121148     | SAMBA™ Screw System                                 | Medical Designs LLC     | Additional     |
| K183342     | SambaScrew 3D SI Fusion System                      | Orthofix Inc.           | Additional     |
| K201538     | Life Spine SIMPACT Sacroiliac Joint Fixation System | Life Spine Inc.         | Additional     |
| K161210     | Rialto™ SI Fusion System                            | Medtronic Sofamor Danek | Additional     |

#### Table 5-1 Predicate Devices

#### PERFORMANCE DATA

The Harvester SI: Sacroiliac Fixation System has been tested in the following test modes:

- Axial pullout per ASTM F543 ●
- Torsional properties per ASTM F543
- Static cantilever bending per ASTM F2193
- o Dynamic cantilever bending per ASTM F2193

The results of this non-clinical testing show that the strength of the Harvester SI: Sacroiliac Fixation System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.

## CONCLUSION

The overall technology characteristics and mechanical performance data lead to the conclusion that the Harvester SI: Sacroiliac Fixation System is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OUR/K203780](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OUR/K203780)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
