← Product Code [OUR](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OUR) · K183342

# SambaScrew 3D SI Fusion System (K183342)

_Orthofix, Inc. · OUR · Apr 30, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OUR/K183342

## Device Facts

- **Applicant:** Orthofix, Inc.
- **Product Code:** [OUR](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OUR.md)
- **Decision Date:** Apr 30, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The SambaScrew 3D SI Fusion System is intended for fixation of sacroiliac joint disruptions, and intended for sacrolliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis.

## Device Story

SambaScrew 3D SI Fusion System consists of sterile, cannulated titanium alloy screws (11mm and 12mm) with fenestrations and a 3D-printed porous mid-shaft region; designed for packing with autograft or allograft materials. Used for fixation and fusion of sacroiliac joint disruptions and degenerative sacroiliitis. Implants are placed by surgeons in a clinical setting to stabilize the joint; porous structure facilitates bone ingrowth. System includes non-sterile instruments for implantation. Benefits include joint stabilization and fusion to address pain and instability associated with SI joint conditions.

## Clinical Evidence

Bench testing only. Mechanical performance evaluated via static and dynamic cantilever beam tests (ASTM F2193), static torsion tests (ASTM F543), and static axial pull-off tests (ASTM F543). No clinical data provided.

## Technological Characteristics

Constructed from medical-grade titanium alloy (Ti-6Al-4V ELI). Features 3D-printed porous mid-shaft region with open macroscopic pores and microscopic roughened surface. Cannulated design for screw placement. Mechanical testing performed per ASTM F2193 and ASTM F543. Non-sterile instruments provided with sterile implants.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- SAMBA Screw System ([K121148](/device/K121148.md))
- Simmetry Sacroiliac Joint Fusion System ([K151818](/device/K151818.md))
- iFuse Implant System ([K182983](/device/K182983.md))

## Reference Devices

- CONSTRUX Mini PEEK Ti Spacer System ([K172437](/device/K172437.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 30, 2019

Orthofix Inc. Ms. Jacki Koch Senior Regulatory Affairs Specialist 3451 Plano Parkway Lewisville, Texas 75056

Re: K183342

Trade/Device Name: SambaScrew 3D SI Fusion System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: OUR Dated: April 1, 2019 Received: April 2, 2019

Dear Ms. Koch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K183342

#### Device Name SambaScrew 3D SI Fusion System

Indications for Use (Describe)

The SambaScrew 3D SI Fusion System is intended for fixation of sacroiliac joint disruptions, and intended for sacrolliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis.

Type of Use (Select one or both, as applicable)

| <div> <span> <span style="font-size: 16px;">☑</span>      Prescription Use (Part 21 CFR 801 Subpart D)     </span> </div> |
|---------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">□</span>      Over-The-Counter Use (21 CFR 801 Subpart C)     </span> </div>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape on the left, followed by the word "ORTHOFIX" in black, bold letters. A small registration mark is visible to the upper right of the "X" in "ORTHOFIX".

# 510(k) SUMMARY

## SambaScrew 3D SI Fusion System

| 510(k) Owner Information<br>Name:<br>Address:       | Orthofix Inc.<br>3451 Plano Parkway<br>Lewisville, TX 75056                                                                    |
|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Telephone Number:<br>Fax Number:<br>Email:          | 214.937.2100<br>214-937-3322<br>jackikoch@orthofix.com                                                                         |
| Registration Number:                                | 2183449                                                                                                                        |
| Contact Person:                                     | Jacki Koch, Senior Regulatory Affairs Specialist                                                                               |
| Date Prepared:                                      | April 1, 2019                                                                                                                  |
| Name of Device<br>Trade Name / Proprietary<br>Name: | SambaScrew 3D SI Fusion System                                                                                                 |
| Common Name:                                        | Sacroiliac Joint Fixation Bone                                                                                                 |
| Product Code:                                       | OUR                                                                                                                            |
| Regulatory Classification:                          | Class II – 21 CFR § 888.3040                                                                                                   |
| Review Panel:                                       | Orthopedic Device Panel                                                                                                        |
| Primary Predicate Device:                           | K121148 – SAMBA Screw System – Orthofix Inc.                                                                                   |
| Additional Predicate<br>Devices:                    | K151818 - Simmetry Sacroiliac Joint Fusion System – Zyga<br>Technology, Inc.<br>K182983 - iFuse Implant System – SI-Bone, Inc. |
| Reference Devices:                                  | K172437 – CONSTRUX Mini PEEK Ti Spacer System – Orthofix Inc.                                                                  |

Reason for 510(k) Submission:

Orthofix is submitting this Traditional 510(k) premarket notification for the new SambaScrew 3D SI Fusion System.

### Device Description

The SambaScrew 3D SI Fusion System is a temporary, multiple component system consisting of non-sterile instruments as well as sterile, 11mm and 12mm cannulated screws of various lengths featuring multiple fenestrations along the shaft. The 11mm cannulated screw features a 

{4}------------------------------------------------

tapered proximal end and dual-pitch threads while the 12mm cannulated screw features a single pitch thread on the proximal and distal ends. The SambaScrews are constructed from medicalgrade titanium alloy (Ti-6AI-4V ELI). The 11mm and 12mm SambaScrews are 3D printed with a mid-shaft porous region. The porous titanium region has open macroscopic 3D pores with a microscopic roughened surface. SambaScrew allows for packing of autograft and allograft materials.

## Indications for Use

The SambaScrew 3D SI Fusion System is intended for fixation of sacroillac joint disruptions, and intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis.

### Summary of the Technological Characteristics of the Device Compared to the Selected Predicate Devices

The technological characteristics of the SambaScrew 3D SI Fusion System is similar to the predicate devices in terms of design, dimensions, intended use, materials, and performance characteristics. There are no significant differences between the SambaScrew 3D SI Fusion System and the predicate devices which would adversely affect the use of the product.

#### PERFORMANCE DATA – Summary of Non-Clinical Test Conducted for Determination of Substantial Equivalence

Mechanical testing consisting of Static and Dynamic cantilever beam Test. Static Torsion Test and Static Axial Pull-Off Test were conducted in accordance to ASTM F2193 Standard Specifications and Test Methods for Components Used in the Surgical Fixation of the Spinal Skeletal System and in accordance to ASTM F543 Standard Specification and Test Methods for Metallic Medical Bone Screws.

| Subject Device                      | Mechanical Testing                                                                                                                                                        |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SambaScrew 3D SI Fusion<br>implants | Static cantilever beam test per ASTM F2193-18a Standard<br>Specifications and Test Methods for Components Used in the<br>Surgical Fixation of the Spinal Skeletal System  |
| SambaScrew 3D SI Fusion<br>implants | Dynamic cantilever beam test per ASTM F2193-18a Standard<br>Specifications and Test Methods for Components Used in the<br>Surgical Fixation of the Spinal Skeletal System |
| SambaScrew 3D SI Fusion<br>implants | Static torsion test per ASTM F543-17, Standard Specification<br>and Test Methods for Metallic Medical Bone Screws                                                         |
| SambaScrew 3D SI Fusion<br>implants | Static axial pull-off test per ASTM F543-17 Standard<br>Specification and Test Methods for Metallic Medical Bone<br>Screws                                                |

Table 1: Mechanical Performance Testing

### Basis of Substantial Equivalence

The new SambaScrew 3D SI Fusion System has the same intended use, the same indications for use, the same technological characteristics and design, same materials and the same principles of operation as the predicate SAMBA Screw System (K121148), the Slmmentry Sacroiliac Joint Fusion System (K151818) and the iFuse Implant System (K162733).

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OUR/K183342](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OUR/K183342)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
