The SImmetry™ Sacroiliac Joint Fusion System is intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis.
Device Story
SImmetry Sacroiliac Joint Fusion System consists of cannulated titanium alloy screws (6.5mm-12.5mm diameter; 30mm-70mm length) and titanium washers. Device used by surgeons for sacroiliac joint fusion to address joint disruptions and degenerative sacroiliitis. Provides mechanical fixation and rigidity to the sacroiliac joint to facilitate fusion. Operates via standard orthopedic surgical fixation techniques.
Clinical Evidence
Bench testing only. Mechanical performance data including axial pull-out strength, driving torque, static bend strength, and bending fatigue demonstrate necessary rigidity to support intended use.
Technological Characteristics
Cannulated bone fixation screws and washers. Material: Titanium Alloy (Ti-6Al-4V ELI). Dimensions: 6.5mm-12.5mm diameter, 30mm-70mm length. Principle: Mechanical fixation/fastener. Class II device per 21 CFR 888.3040.
Indications for Use
Indicated for patients requiring sacroiliac joint fusion due to sacroiliac joint disruptions or degenerative sacroiliitis.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
SImmetry™ Sacroiliac Joint Fusion System (K110512)
Related Devices
K102907 — SIMMETRY SACROILIAC JOINT FUSION SYSTEM · Zyga Technology, Inc. · Dec 14, 2010
K110512 — SIMMETRY SACROILIAC JOINT FUSION SYSTEM · Zyga Technology, Inc. · Mar 23, 2011
K151818 — SImmetry Sacroiliac Joint Fusion System · Zyga Technology, Inc. · Aug 5, 2015
K130092 — SIMMETRY SACROILIAC JOINT FUSION SYSTEM · Zyga Technology, Inc. · Feb 27, 2013
{0}------------------------------------------------
K111801
JUL 2 1 2011
### 510(k) Summary
## [As required by 21 CFR 807.92(c)]
# 1. Submitter / Contact Person / Date of Preparation
| Submitter | Zyga Technology, Inc.<br>700 10th Ave South, Suite 20<br>Minneapolis, MN 55415-1745 |
|---------------------|----------------------------------------------------------------------------------------------------------------|
| Contact Person | Diane Brinza<br>Director of Regulatory and Clinical Affairs<br>Ph. 612.455.1061, ext. 104<br>Fax. 612.455.1064 |
| Date of Preparation | June 22, 2011 |
# 2. General Information
| Trade Name | SImmetry™ Sacroiliac Joint Fusion System |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common / Usual<br>Name | Sacroiliac Joint Fixation / Sacroiliac Joint Fusion |
| Classification Name | Smooth or threaded metallic bone fixation fastener |
| Classification | Class II (per 21 CFR § 888.3040) |
| Manufacturer | Zyga Technology, Inc.<br>700 10th Ave South<br>Minneapolis, MN 55415-1745 |
| Identification of<br>Predicate Devices | K110512<br>Zyga Technology, Inc.<br>SImmetry™ Sacroiliac Joint Fusion System |
| Device Description | The SImmetry™ Sacroiliac Joint Fusion System<br>consists of cannulated screws available in titanium<br>having diameters ranging from 6.5mm-12.5mm; and<br>lengths of 30mm-70mm; titanium washers are<br>available for use with the 6.5mm diameter screws. |
| Intended Use /<br>Indications for Use | The SImmetry™ Sacroiliac Joint Fusion System is<br>intended for sacroiliac joint fusion for conditions<br>including sacroiliac joint disruptions and degenerative<br>sacroiliitis. |
| Technological<br>Characteristic | The principle of operation and fundamental scientific<br>technology of the subject devices is identical to that of<br>the identified predicate. |
| Materials | The subject devices are manufactured from Titanium<br>Alloy (Ti-6Al-4V ELI). |
| Technological<br>Comparison | The principle of operation and fundamental scientific<br>technology of the subject devices is identical to that of<br>the identified predicate device. |
Page of 2
{1}------------------------------------------------
# 长川时时
.
| Summary of<br>Non-clinical<br>Performance Data | Mechanical testing for axial pull out strength, driving<br>torque, static bend strength and bending fatigue<br>provided in this and previous 510(k)s demonstrate<br>that the SImmetry Sacroiliac Joint Fusion System<br>provides the necessary rigidity to support the intended<br>use. |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion | Equivalence for the SImmetry Sacroiliac Joint Fusion<br>System is based on the same intended use, design<br>features, operational principles, and material<br>composition and mechanical performance when<br>compared to the predicate device cleared under<br>K110512. |
and the comments of the comments of the comments of
:
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Zyga Technology, Incorporated % Ms. Diane Brinza Director of Regulatory Clinical and Quality Assurance 700 10th Avenue South, Suite 400 Minneapolis, Minnesota 55415
JUL 2 1 2011
Re: K111801
Trade/Device Name: SImmetryTM Sacroiliac Joint Fusion System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: OUR Dated: June 22, 2011 Received: June 27, 2011
Dear Ms. Brinza:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
#### Page 2 - Ms. Diane Brinza
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melk Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
Device Name: SImmetry™ Sacroiliac Joint Fusion System
### Indications for Use:
The SImmetry™ Sacroiliac Joint Fusion System is intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroillitis.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for M. Melkerson
(Division/Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K11180/
2. lof/
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.