← Product Code [OLC](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/OLC) · K170364

# Collagen Meniscus Implant XL (K170364)

_Ivy Sports Medicine, LLC · OLC · May 2, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OLC/K170364

## Device Facts

- **Applicant:** Ivy Sports Medicine, LLC
- **Product Code:** [OLC](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/OLC.md)
- **Decision Date:** May 2, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Ivy Sports Medicine Collagen Meniscus Implant (CMI) is intended for use in surgical procedures for the reinforcement and repair of soft tissue injuries of the medial meniscus. In repairing medial meniscal defects, the patient must have an intact meniscal rim and anterior and posterior horns for attachment of the mesh. In addition, the surgically prepared site for the CMI must extend at least into the red/white zone of the meniscus to provide sufficient vascularization. The CMI reinforces soft tissue and provides a resorbable scaffold that is replaced by the patient's own soft tissue. The CMI is not a prosthetic device and is not intended to replace normal body structure.

## Device Story

Resorbable collagen matrix; bovine type I collagen; semi-lunar shape; triangular cross-section. Used by surgeons in orthopedic procedures to reinforce weakened soft tissue in the medial meniscus. Surgeon trims device to fit defect; device acts as scaffold for patient's own tissue ingrowth; eventually resorbed. Not a prosthetic; does not replace normal body structure. Provided sterile for single use.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench performance testing and identical material composition/technology to predicate.

## Technological Characteristics

Resorbable bovine type I collagen matrix; semi-lunar shape with triangular cross-section; sterile, single-use; dimensions modified (12.5% height increase) from predicate.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- ReGen Collagen Scaffold Surgical Mesh ([K082079](/device/K082079.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines suggesting movement or transformation.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Ivy Sports Medicine, LLC John Dichiara Senior Director of Regulatory and Clinical 545 Penobscot Drive Redwood City, California 94063

May 2, 2017

Re: K170364

Trade/Device Name: Ivy Sports Medicine Collagen Meniscus Implant XL (CMI XL) Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: OLC Dated: January 27, 2017 Received: February 6, 2017

Dear Mr. Dichiara:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

## Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K170364

Device Name

Ivy Sports Medicine Collagen Meniscus Implant XL (CMI XL)

Indications for Use (Describe)

The Ivy Sports Medicine Collagen Meniscus Implant (CMI) is intended for use in surgical procedures for the reinforcement and repair of soft tissue injuries of the medial meniscus. In repairing medial meniscal defects, the patient must have an intact meniscal rim and anterior and posterior horns for attachment of the mesh. In addition, the surgically prepared site for the CMI must extend at least into the red/white zone of the meniscus to provide sufficient vascularization.

The CMI reinforces soft tissue and provides a resorbable scaffold that is replaced by the patient's own soft tissue. The CMI is not a prosthetic device and is not intended to replace normal body structure.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Ivy Sports Medicine, LLC. The logo consists of a stylized green ivy leaf design on the left, followed by the company name in bold, black text. The text is aligned to the right of the ivy leaf design.

### SECTION 6

## 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

#### 510(k) Number: K170364

#### Applicant Information:

| Owner Name: | Ivy Sports Medicine, LLC                      |
|-------------|-----------------------------------------------|
| Address:    | 545 Penobscot Drive<br>Redwood City, CA 94063 |
| Main Phone: | 650-562-0800                                  |
| Fax:        | 650-562-0808                                  |

Establishment Registration Number: 2956141

| Contact Person: | John Dichiara       |
|-----------------|---------------------|
|                 | Phone: 917-439-2597 |
|                 | Fax: 650-562-0808   |

Date Prepared: January 27, 2017

#### Device Information:

| Classification:      | Class II                                                                                                   |
|----------------------|------------------------------------------------------------------------------------------------------------|
| Trade Name:          | Collagen Meniscus Implant XL (CMI XL)                                                                      |
| Common name:         | Surgical Mesh                                                                                              |
| Classification name: | Scaffold, Partial Medial Meniscal Defects Extending into the<br>Red/White Zone, Resorbable Bovine Collagen |
| Regulation number:   | 21 CFR §878.3300                                                                                           |
| Product Code:        | OLC                                                                                                        |

#### Predicate Device:

The Collagen Meniscus Implant XL is substantially equivalent in Intended Use, Fundamental Scientific Technology and Performance to the following legally marketed device in commercial distribution: ReGen Collagen Scaffold Surgical Mesh cleared

{4}------------------------------------------------

# バル Ivy Sports Medicine, LLC

under K082079 (Also known as the ReGen CS and the Ivy Sports Medicine Collagen Meniscus Implant (CMI)).

#### Device Description:

The Ivy Sports Medicine Collagen Meniscus Implant XL (CMI XL) is a resorbable collagen matrix comprised primarily of bovine type I collagen. The CMI is provided in a semi-lunar shape with a triangular cross section to be used to reinforce weakened soft tissue and provide a resorbable scaffold that is replaced by the patient's own tissue. The surgeon trims the device to the size necessary for repair of the damaged or weakened soft tissue. The CMI XL is provided sterile for single use only.

#### Intended Use:

The Ivy Sports Medicine Collagen Meniscus Implant (CMI) is intended for use in surgical procedures for the reinforcement and repair of soft tissue injuries of the medial meniscus. In repairing and reinforcing medial meniscal defects, the patient must have an intact meniscal rim and anterior and posterior horns for attachment of the mesh. In addition, the surqically prepared site for the CMI must extend at least into the red/white zone of the meniscus to provide sufficient vascularization. The CMI reinforces soft tissue and provides a resorbable scaffold that is replaced by the patient's own soft tissue. The CMI is not a prosthetic device and is not intended to replace normal body structure.

#### Comparison to Predicate Device(s):

The CMI XL has the same intended use and technological characteristics as the predicate device, ReGen Collagen Scaffold, with the exception of the physical dimensions and mass. Any differences identified do not raise new questions of safety or effectiveness, as supported by appropriate testing.

#### Technological Characteristics/Performance Data:

The Collagen Meniscus Implant XL is substantially equivalent to the predicate device in intended use, fundamental scientific technology, and performance specifications. Design verification testing was performed to verify that the performance of the Collagen Meniscus Implant XL remains substantially equivalent to the predicate device. Testing performed on the Collagen Meniscus Implant XL demonstrated that all of the predetermined acceptance criteria were met with passing results.

#### Clinical Testing:

Clinical evaluation is not required for this device. TThe product has the same material composition, fundamental scientific technology, shape and performance specifications. The only difference between the CMI XL and the predicate is that the overall height dimension is approximately 12.5% greater than the predicate to better match the height of the native meniscus rim in patients where the existing two sizes of the CS device do not meet the surgeon and patient needs.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Ivy Sports Medicine, LLC. The logo consists of a stylized green plant with three leaves on the left, followed by the text "Ivy Sports Medicine, LLC" in black. The text is in a bold, sans-serif font.

#### Substantial equivalence:

The Collagen Meniscus Implant XL has the following similarities to the ReGen Collagen Scaffold predicate device cleared under K082079.

- has the same intended use and indication for use: ●
- . has the same technological characteristics;
- . has the same principles of operation;
- . has equivalent performance characteristics; and
- . has the same sterilization process.

#### Summary

In summary, the Collagen Meniscus Implant XL, the subject of this submission, is as safe and effective as the ReGen Collagen Scaffold. It has the same indication for use, the same principles of operation and the same technological characteristics as the ReGen Collagen Scaffold. The differences between the Collagen Meniscus Implant XL and the ReGen Collagen Scaffold raise no new issues of safety or effectiveness. Bench performance test data demonstrate that the Collagen Meniscus Implant XL is as safe and effective as the ReGen Collagen Scaffold and is therefore substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OLC/K170364](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OLC/K170364)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
