Curiteva Porous PEEK Cervical Interbody Fusion System
K213030 · Curiteva, Inc. · ODP · Feb 13, 2023 · Orthopedic
Device Facts
Record ID
K213030
Device Name
Curiteva Porous PEEK Cervical Interbody Fusion System
Applicant
Curiteva, Inc.
Product Code
ODP · Orthopedic
Decision Date
Feb 13, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Curiteva Porous PEEK Cervical Interbody Fusion System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The Curiteva Porous PEEK Cervical Interbody Fusion System is intended for use for anterior cervical interbody fusion in patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, myelopathy, and/or pain at multiple contiguous levels from C2 - T1. The System is intended to be used with supplemental fixation. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion.
Device Story
Interbody fusion system consisting of sterile, single-use, box-shaped implants; designed for anterior cervical spine surgery. Implants feature open central corridor for bone graft packing; porous PEEK structure on superior/inferior surfaces and graft corridor; hydroxyapatite (HA) surface coating. Used by surgeons in clinical settings to stabilize spine segments; requires supplemental fixation. Implants accommodate varying patient anatomy; facilitate fusion via autogenous/allogeneic bone graft. Benefits include spinal stabilization and symptom relief for degenerative conditions.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing including axial compression, compression-shear, torsion (ASTM F2077), subsidence (ASTM F2267), wear debris characterization (ASTM F1877), expulsion, and impaction testing.
Technological Characteristics
Implant-grade PEEK (ASTM F2026) with Titanium alloy markers (ASTM F136). Features porous PEEK structure and 20nm thick hydroxyapatite (HA) coating. Sterile, single-use, box-shaped form factor. No software or energy source.
Indications for Use
Indicated for skeletally mature patients with cervical disc degeneration and/or cervical spinal instability resulting in radiculopathy, myelopathy, and/or pain at multiple contiguous levels from C2-T1. Requires use with supplemental fixation and bone graft.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Curiteva Cervical Interbody Fusion System (K181261)
Astura Medical ALTA Anterior Cervical Interbody Spacer (K173324)
Cutting Edge Spine, LLC EVOL SI Joint Fusion System (K190025)
Related Devices
K254061 — Curiteva Porous PEEK Cervical Interbody Fusion System · Curiteva, Inc. · Jan 15, 2026
K251080 — IdentiTi II Cervical Interbody System · Alphatec Spine, Inc. · Jul 3, 2025
K172676 — NuVasive® Modulus-C Interbody System · Nu Vasive, Incorporated · Dec 14, 2017
K173324 — ALTA Anterior Cervical Interbody Spacer · Astura Medical · Feb 27, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 13, 2023
Curiteva, Inc. % Meredith May, MS, RAC Director of Consulting Empirical Testing Corp. 4628 Northpark Drive Colorado Springs, Colorado 80918
Re: K213030
Trade/Device Name: Curiteva Porous PEEK Cervical Interbody Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP Dated: January 11, 2023 Received: January 11, 2023
Dear Ms. May:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent Showalter -S
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213030
#### Device Name
Curiteva Porous PEEK Cervical Interbody Fusion System
Indications for Use (Describe)
The Curiteva Porous PEEK Cervical Interbody Fusion System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The Curiteva Porous PEEK Cervical Interbody Fusion System is intended for use for anterior cervical interbody fusion in patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, myelopathy, and/or pain at multiple contiguous levels from C2 - T1. The System is intended to be used with supplemental fixation. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - K213030
#### A. Submitter Information
| Submitter: | Curiteva, Inc.<br>25127 Will McComb Drive, Suite 100<br>Tanner, AL 35671<br>Phone: (256) 213-1057<br>Fax: (256) 213-1058 |
|---------------------|--------------------------------------------------------------------------------------------------------------------------|
| Submitter Contact: | Eric Linder<br>Chief Technology Officer |
| Regulatory Contact: | Meredith May MS, RAC<br>Empirical Testing Corp.<br>Phone: (719) 337-7579<br>mpvanderbilt@empiricaltech.com |
| Date Prepared: | February 9, 2023 |
#### B. Device Information
| Trade Name: | Curiteva Porous PEEK Cervical Interbody Fusion System |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Intervertebral Body Fusion Device |
| Classification Name: | Intervertebral Fusion Device With Bone Graft, Cervical |
| Device Classification: | Class II (per 21 CFR 888.3080) |
| Product Code: | ODP |
| Classification Panel: | Division of Orthopedic Devices |
| Predicate Device(s): | Primary: Curiteva Cervical Interbody Fusion System – K181261<br>Additional: Astura Medical ALTA Anterior Cervical Interbody<br>Spacer - K173324<br>Additional: Cutting Edge Spine, LLC EVOL SI Joint Fusion<br>System - K190025 |
#### C. Device Description
The Curiteva Porous PEEK Cervical Interbody Fusion System implants are sterile, single-use devices and available in a variety of different footprints, styles and sizes to accommodate the individual pathology and anatomical conditions of the patient. The implants are generally boxshaped with an open central corridor to permit packing with bone graft to facilitate fusion. The implants have a dense central ring with a porous structure lining the vertical graft corridor and on
{4}------------------------------------------------
the superior and inferior surfaces of the construct. Each implant has been surface treated with a hydroxyapatite (HA) coating that is approximately 20nm thick.
The Curiteva Porous PEEK Cervical Interbody Fusion System implants are manufactured from implant-grade PEEK (per ASTM F2026) with Titanium alloy markers (per ASTM F136).
## D. Indications for Use
The Curiteva Porous PEEK Cervical Interbody Fusion System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The Curiteva Porous PEEK Cervical Interbody Fusion System is intended for use for anterior cervical interbody fusion in patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT. MRI), that results in radiculopathy, myelopathy, and/or pain at multiple contiguous levels from C2 - T1. The System is intended to be used with supplemental fixation. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion.
### E. Comparison of Technological Characteristics
As was established in this submission, the subject Curiteva Porous PEEK Cervical Interbody Fusion System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and to have similar technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, function, and range of sizes.
## F. Performance Data
Non-clinical testing performed on the Curiteva Porous PEEK Cervical Interbody Fusion System supports substantial equivalence to predicate devices. The following testing was performed:
- Axial compression per ASTM F2077
- Compression-shear per ASTM F2077 ●
- Torsion per ASTM F2077 ●
- Subsidence per ASTM F2267
- Wear debris characterization per ASTM F1877
Additional testing (e.g. expulsion, impaction) was performed to further support substantial equivalence. The results of non-clinical testing demonstrate that the strength and performance of the Curiteva Porous PEEK Cervical Interbody Fusion System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.
#### G. Conclusion
Based on the indications for use, technological characteristics, non-clinical performance testing, and comparison to predicate devices, the subject Curiteva Porous PEEK Cervical Interbody Fusion System has been shown to be substantially equivalent to legally marketed predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.