The EDEN Peek Cage - PLIF Type is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involvedlevel(s).This device is to be used with autogenous bone graft. The EDEN Peek Cage System is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage. The EDEN Peek Cage - ACIF Type is indicated for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from the C2-C3 disc to the C7-T1 disc. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. Patients should have at least six weeks of non-operative treatment with an intervertebral cage.
Device Story
Intervertebral body fusion device; implanted via anterior (ACIF) or posterior (PLIF) surgical approach; provides structural support to disc space during fusion; used with autogenous bone graft and supplemental fixation; manufactured from PEEK with tantalum markers for radiographic visualization; intended for use by surgeons in clinical settings; benefits patients by stabilizing spinal segments and facilitating fusion in DDD cases.
Clinical Evidence
No clinical data. Bench testing only: mechanical testing (static/dynamic axial compression, torsion per ASTM F2077; subsidence per ASTM F2267) and sterilization validation (ISO 11137) with 36-month shelf life validation.
Technological Characteristics
Materials: Polyether ether ketone (VESTAKEEP i4 R, ASTM F2026) and tantalum markers (ASTM F560). Form factor: Intervertebral cage configurations for ACIF and PLIF approaches. Sterilization: Gamma radiation. Connectivity: None (passive implant).
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) requiring intervertebral body fusion. PLIF Type: L2-S1 (1-2 levels), DDD with/without Grade 1 spondylolisthesis/retrolisthesis, requires 6 months non-operative treatment. ACIF Type: C2-T1 (1 level), DDD with radicular symptoms, requires 6 weeks non-operative treatment. Both require autograft and supplemental fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K183659 — Omnia Medical Rotary PLIF System · Omnia Medical, LLC · Feb 6, 2019
K073502 — VERTEBRON VBR SYSTEM · Vertebron, Inc. · Mar 24, 2008
K100865 — AVA NAVIGATOR PEEK SPACERS, MODEL 48392XXX, 48393XXX · Stryker Spine · Aug 11, 2010
K131077 — LIFE SPINE PLATEAU SPACER SYSTEM · Life Spine · Jul 17, 2013
K160074 — Rampart D Lumbar Interbody Fusion Device · Spineology, Inc. · Oct 18, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" on the bottom.
September 13, 2021
JMT Co., Ltd Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 1150 Roosevelt STE 200 Irvine, California 92620
Re: K201793
Trade/Device Name: EDEN Peek Cage Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP, MAX Dated: July 13, 2021 Received: July 15, 2021
Dear Priscilla Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201793
Device Name EDEN Peek Cage
#### Indications for Use (Describe)
The EDEN Peek Cage - PLIF Type is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involvedlevel(s).This device is to be used with autogenous bone graft. The EDEN Peek Cage System is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.
The EDEN Peek Cage - ACIF Type is indicated for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from the C2-C3 disc to the C7-T1 disc. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. Patients should have at least six weeks of non-operative treatment with an intervertebral cage.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <div> <span> <b> \[X] Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div> | <div> <span> \[] Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
(K201793)
This summary of 510(K) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: 9/3/2021
## 1. Submitter
JMT Co., Ltd (11517) 70-39, Gwonyul-ro 1203 beon-gile, Baekseok-eup, Yangju-si Gyeonggi-do, Republic of Korea Tel. +82-31-868-0951
## 2. U.S Agent/Contact Person
Priscilla Chung LK Consulting Group USA, Inc. 1150 Roosevelt STE 200, Irvine CA 92620 Phone: 714.202.5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com
## 3. Device
- Trade Name: EDEN Peek Cage
- Common Name: Intervertebral Body Fusion System ●
- Classification:
- Product Code: ODP (888.3080, Intervertebral Fusion Device With Bone Graft, Cervical, Class II, 510k)
- Product Code: MAX (888.3080, Intervertebral Fusion Device With Bone Graft, Lumbar, Class II, 510k)
## 4. Predicate Devices:
- LnK Lumbar Interbody Fusion Cage System by L&K BIOMED Co., Ltd. ●
- (K181380) Velofix™ Interbody Fusion System by U & I Corporation (K171749) .
## 5. Device Description:
The EDEN Peek Cage is offered in various device configurations based on based on surgical approach and patient anatomy, and consist of:
{4}------------------------------------------------
Cervical Interbody Fusion Device (ACIF Type) which may be implanted as a single device via an anterior approach.
Lumbar Interbody Fusion Device (PLIF Type) which can be implanted Bi-laterally via a posterior approach.
- · As a single device via an Anterior approach;(ACIF Type)
- · Bi-laterally via a posterior approach:(PLIF Type)
The EDEN Peek Cage is made of polyether ether ketone (VESTAKEEP® i4 R) that conforms to ASTM F2026 and they have tantalum markers (ASTM F560) to assist the surgeon for proper placement of the device. The subject devices are offered gamma sterile.
### 6. Indication for use:
The EDEN Peek Cage - PLIF Type is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).This device is to be used with autogenous bone graft. The EDEN Peek Cage System is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.
The EDEN Peek Cage - ACIF Type is indicated for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from the C2-C3 disc to the C7-T1 disc. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. Patients should have at least six weeks of non-operative treatment prior to treatment with an intervertebral cage.
### 7. Basis for Substantial Equivalence
The subject device is substantially equivalent to LnK Lumbar Interbody Fusion Cage System (K181380) by L&K BIOMED Co., Ltd. and VelofixTM Interbody Fusion System (K171749) by U & I Corporation.
They have the same indications for use and use the same materials with very similar designs. We have conducted the performance tests and the results support that the subject device is substantially equivalent to the predicate devices.
{5}------------------------------------------------
## 8. Non-Clinical Testing
- Sterilization validating testing has been performed in accordance with ISO 11137 .
- . 36-month shelf life validation
- . Mechanical test (static and dynamic axial compression, static and dynamic torsion per ASTM F2077 and subsidence per ASTM F2267)
## 9. Conclusion
The subject device and the predicate devices have the same intended use and have the same technological characteristics. Based on the similarities and the test results, we conclude that the EDEN Peek Cage is substantially equivalent to the predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.