4CIS® Marlin PEEK ACIF Cage is indicated for use in cervical intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at the levels from the C2-C3 disc to the C7-T1 disc. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should have six weeks of non-operative therapy in advance. The 4CIS® Marlin PEEK ACIF Cage is to be used with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, and is to be implanted via an open, anterior approach. It is intended to be used with supplemental spinal fixation systems that have been cleared for use in the cervical spine, such as Anterior Cervical Plate system.
Device Story
The 4CIS Marlin PEEK ACIF Cage is a single-component intervertebral body fusion device designed to restore disc space height and maintain physiological lordosis in the cervical spine. The device is implanted via an open, anterior approach by a surgeon. It features a hollow center to accommodate autogenous or allogenic bone graft to facilitate fusion. The cage surface includes vertical square teeth to prevent subsidence and migration. Three tantalum markers are embedded in the device to allow for radiographic visualization. The device is intended for use in conjunction with supplemental spinal fixation systems, such as an anterior cervical plate. It provides mechanical support to the spinal column during the fusion process.
Clinical Evidence
Bench testing only. Mechanical performance was evaluated via static axial compression, static compression-shear, static torsion, dynamic axial compression, dynamic compression-shear, dynamic torsion, and static subsidence testing in accordance with ASTM F2077-17 and ASTM F2267-04. No clinical data was provided.
Technological Characteristics
Material: Poly-ether-ether-ketone (PEEK) per ASTM F2026. Features: Hollow center for bone graft, vertical square teeth for anchoring, three tantalum markers for radiopacity. Form factor: Various shapes and sizes for anatomical fit. Sterilization: Provided non-sterile for single use.
Indications for Use
Indicated for cervical intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) at levels C2-C3 through C7-T1. Requires six weeks of prior non-operative therapy. Contraindications include patients without skeletal maturity or those not requiring fusion.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K242195 — Gemini Cervical Fusion Cage System · ZheJiang Decans Medical Devices Co., Ltd. · Nov 4, 2024
K212358 — Romero Cervical Cage · Spineup, Inc. · Jan 10, 2022
K200592 — GS Medical AnyPlus PEEK Cage System · GS Medical Co., Ltd. · Aug 10, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 19, 2019
Solco Biomedical Co., Ltd. Hwi-Geun Yu RA & Exam Certification Team Leader 154 Seotan-ro, Seotan-myeon Pyeongtaek-Si, Gyeonggi-do, 17704 Republic of Korea
Re: K192044
Trade/Device Name: 4CIS Marlin PEEK ACIF Cage Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP Dated: November 29, 2019 Received: December 4, 2019
Dear Hwi-Geun Yu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Brent L. Showalter, PhD DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
### K192044
Device Name
4CIS® Marlin PEEK ACIF Cage
#### Indications for Use (Describe)
4CIS® Marlin PEEK ACIF Cage is indicated for use in cervical intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at the levels from the C2-C3 disc to the C7-T1 disc. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should have six weeks of non-operative therapy in advance. The 4CIS® Marlin PEEK ACIF Cage is to be used with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, and is to be implanted via an open, anterior approach. It is intended to be used with supplemental spinal fixation systems that have been cleared for use in the cervical spine, such as Anterior Cervical Plate system.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------|
| | <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Submitter | Solco Biomedical Co., Ltd.<br>154 Seotan-ro, Seotan-myeon, Pyeongtaek, Gyeonggi-do,<br>17704 Republic of Korea<br>Phone. +82-31-664-4101<br>Fax. +82-31-663-8983 |
|-------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Hwi-geun Yu<br>Solco Biomedical Co., Ltd.<br>154 Seotan-ro, Seotan-myeon, Pyeongtaek, Gyeonggi-do,<br>17704 Republic of Korea<br>Phone: +82-31-610-4091<br>Fax: +82-31-663-8983 |
| Preparation Date | July 22, 2019 |
| Trade / Proprietary name | 4CIS® Marlin PEEK ACIF Cage |
| Common / Usual Name | Cervical Intervertebral Body Fusion Device |
| Classification Name<br>Classification Code<br>Regulatory Class<br>Regulation Number | Intervertebral Fusion Device With Bone Graft, Cervical<br>ODP<br>Class II<br>888.3080 |
| Predicate Device | 4CIS® Marlin ACIF Cage System (K162402) [Solco<br>Biomedical Co., Ltd.] - Primary Predicate<br>PATRIOT SPACERS: COLONIAL ACDF (K072991,<br>Additional)<br>TRYPTIK Ca (K091873, Additional)<br>MATISSE Anterior Cervical Interbody Fusion Cage<br>System (K162682, Additional) |
| Description of Device | 4CIS® Marlin PEEK ACIF Cage is single component<br>devices used to restore height of disc space by anterior<br>approach and to facilitate cervical interbody fusion with<br>maintaining physiological lordotic angulation of cervical<br>spine. To allow maximum preservation and ensure ample<br>contact surfaces with bony endplate, a variety of shapes and<br>sizes are available and each device has three tantalum<br>markers for ease of visualization on radiographs. The<br>vertical square teeth on the top and the bottom surface<br>prevent subsidence of the cage into the vertebral body while<br>they increase the anchoring and prevent slipping or<br>expulsion. To make solid fusion of intervertebral body,<br>hollow space in the implant allows bone graft material to be<br>filled. The implant has safety proven structure and material<br>(Poly-ether-ether-ketone, ASTM F2026) to promote<br>biological synostosis and assures mechanical safety against<br>load. |
| Indication for Use | 4CIS® Marlin PEEK ACIF Cage is indicated for use in<br>cervical intervertebral body fusion procedures in skeletally<br>mature patients with degenerative disc disease (DDD) at the<br>levels from the C2-C3 disc to the C7-T1 disc. DDD is defined<br>as neck pain of discogenic origin with degeneration of the disc<br>confirmed by patient history and radiographic studies. These<br>patients should have six weeks of non-operative therapy in<br>advance. The 4CIS® Marlin PEEK ACIF Cage is to be used<br>with autogenous bone graft and/or allogenic bone graft<br>comprised of cancellous and/or corticocancellous bone graft,<br>and is to be implanted via an open, anterior approach. It is<br>intended to be used with supplemental spinal fixation systems<br>that have been cleared for use in the cervical spine, such as<br>Anterior Cervical Plate system. |
| Comparison of<br>Technological<br>Characteristics with<br>the Predicate Devices | The subject device and all the predicates have the same or<br>similar indications for use statements. The subject device is<br>composed of the same material as the predicate devices<br>conforming to recognized industry standards for permanent<br>implants and surgical orthopedic instruments. All they have<br>similar basic design features and functions as well as those<br>dimensions. The subject device and cited predicate devices are<br>provided non-sterile for single use only. The subject device<br>demonstrated equivalent mechanical performance to the cited<br>predicate device under the same test conditions. |
| Performance Data | Mechanical testing (static axial compression, static<br>compression-shear, static torsion, dynamic axial compression,<br>dynamic compression-shear, dynamic torsion and static<br>subsidence) was conducted in accordance with ASTM F2077-<br>17 and F2267-04.<br><br>Above non-clinical performance data in the form of a<br>comprehensive literature review was provided in support of<br>substantial equivalence of the subject device. |
| Conclusion | The overall technology characteristics and mechanical<br>performance data lead to the conclusion that the subject device<br>is substantially equivalent to legally marketed predicate<br>devices. |
{4}------------------------------------------------
{5}------------------------------------------------
K192044
Page 3 of 3
______________________________________________________________________________________________________________________________________________________________________________
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.