← Product Code [ODP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP) · K173159

# Cervical Spinal Truss System (CSTS) Interbody Fusion Device (K173159)

_4Web, Inc. · ODP · Jan 8, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP/K173159

## Device Facts

- **Applicant:** 4Web, Inc.
- **Product Code:** [ODP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP.md)
- **Decision Date:** Jan 8, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Cervical Spine Truss System (CSTS) Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the cervical spine at one level or two contiguous disc levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. CSTS Interbody Fusion Devices are used as an adjunct to fusion in the cervical spine and are placed via an anterior approach at the C2 to T1 disc levels using autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. Patients should have received 6 weeks of non-operative treatment prior to treatment with the devices. The device must be used with supplement fixation.

## Device Story

Cervical Spine Truss System (CSTS) is an interbody fusion device; utilizes open architecture truss design; manufactured via additive manufacturing process; provides structural support while allowing bone through-growth and fusion. Implanted via anterior approach at C2-T1 levels; used as adjunct to fusion; requires supplemental fixation. Device made of Ti6Al4V alloy; available in various sizes to match patient anatomy. Used by surgeons in clinical settings to treat DDD; provides mechanical stability to facilitate spinal fusion; benefits patients by addressing discogenic pain.

## Clinical Evidence

Bench testing only. Testing performed per ASTM F2077 (static/dynamic axial compression, compression shear, torsion), ASTM F2267-04 (subsidence), ASTM F2119 (MR artifact), ASTM F2052 (MR displacement), ASTM F2213 (MR torque), and ASTM F2182 (MR heating). Expulsion testing also conducted.

## Technological Characteristics

Material: Ti6Al4V alloy. Design: Open architecture truss with hierarchical surface roughness via additive manufacturing. Standards: ASTM F2077, ASTM F2267-04, ASTM F2119, ASTM F2052, ASTM F2213, ASTM F2182. Connectivity: None. Sterilization: Sterile packaging added.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- 4WEB CSTS ([K121741](/device/K121741.md))
- [K142112](/device/K142112.md)
- [K171496](/device/K171496.md)
- K11119

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the agency's acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 8, 2018

4Web, Inc. % Rich Jansen Consultant Silver Pine Consulting 11821 Bramble Cove Drive Ft. Myers, Florida 33905

Re: K173159

Trade/Device Name: Cervical Spinal Truss System (CSTS) Interbody Fusion Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP Dated: December 13, 2017 Received: December 14, 2017

Dear Rich Jansen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Katherine D. Kavlock -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES |  |
|-----------------------------------------|--|
| Food and Drug Administration            |  |

# Indications for Use

510(k) Number (if known) K173159

Device Name

Cervical Spine Truss System (CSTS) Interbody Fusion Device

#### Indications for Use (Describe)

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

| The Cervical Spine Truss System (CSTS) Interbody Fusion Device is indicated for use in skeletally mature  |
|-----------------------------------------------------------------------------------------------------------|
| patients with Degenerative Disc Disease (DDD) of the cervical spine at one level or two contiguous disc   |
| levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient         |
| history and radiographic studies. CSTS Interbody Fusion Devices are used as an adjunct to fusion in the   |
| cervical spine and are placed via an anterior approach at the C2 to T1 disc levels using autograft and/or |
| allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. Patients should         |
| have received 6 weeks of non-operative treatment prior to treatment with the devices. The device          |
| must be used with supplement fixation.                                                                    |

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An aqency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

(301) 443-6740

{3}------------------------------------------------

# 510(k) Summary

| Date Prepared:      | January 5, 2018                                                                                                                         |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Contact:            | Jessee Hunt, President<br>4WEB, Inc.<br>2801 Network Blvd., Suite 620<br>Frisco, TX 75034<br>Phone: (800) 285-7090<br>Fax: 972-488-1816 |
| Regulatory Contact: | Rich Jansen, Pharm. D.<br>Silver Pine Consulting, LLC<br>richj@s-pineconsulting.com                                                     |
| Trade Name:         | Cervical Spine Truss System (CSTS) Interbody Fusion Device                                                                              |
| Product Class:      | Class II                                                                                                                                |
| Classification:     | 21 CFR §888.3080                                                                                                                        |
| Common Name:        | Intervertebral Body Fusion Device                                                                                                       |
| Product Codes:      | ODP                                                                                                                                     |
| Panel Code:         | 87                                                                                                                                      |

#### Purpose:

The purpose of this submission is to update the system with design changes to implants and instruments, add sterile packaging, revise the Indications for Use, and add MR Conditional labeling.

#### Indications for Use:

The Cervical Spine Truss System (CSTS) Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the cervical spine at one level or two contiguous disc levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. CSTS Interbody Fusion Devices are used as an adjunct to fusion in the cervical spine and are placed via an anterior approach at the C2 to T1 disc levels using autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. Patients should have received 6 weeks of non-operative treatment with the devices. The device must be used with supplement fixation.

#### Device Description:

The device is an open architecture truss design mathematically formulated to provide structural support with open space throughout the implant for bone through growth and fusion. The 4WEB additive manufacturing process provides a hierarchical surface roughness. The implant is made from Ti6Al4V alloy. The device is available in a variety of sizes to accommodate the patient's anatomy.

{4}------------------------------------------------

## Predicate Device(s):

The primary predicate device is the 4WEB CSTS (K121741). Additional predicates include K142112, K171496 and K11119.

### Performance Standards:

Performance testing has been completed per the following standards: ASTMF2077 - Static and dynamic axial compression, static and dynamic compression shear, and static and dynamic torsion ASTM F2267-04 - Subsidence Testing ASTM F2119 – MR Image Artifact ASTM F2052 – MR Induced Displacement Force ASTM F2213 – MR Induced Torque ASTM F2182 - MR Induces Heating Expulsion testing per accepted industry standard.

### Technological Characteristics:

4Web, Inc. has compared these changes to the previously cleared predicate devices in regards to indications for use, materials, function, sizes and mechanical test results. These comparisons demonstrate substantial equivalence to the predicate devices.

#### Conclusion:

4WEB, Inc. concludes that the CSTS devices are substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP/K173159](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/ODP/K173159)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
