The Halis™ Lumbar Cage System is indicated for use in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S 1 whose condition requires use of interbody fusion combined with supplemental fixation. These DDD patients may also have up to Grade 1 Spondylolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have had six months of non-operative treatment prior to surgery. The Halis™ Lumbar Cage System is intended to be used with autogenous bone graft comprised of cancellous and/or corticocancellous bone graft. The devices are intended to be used with supplemental fixation. The devices may be implanted through several surgical approaches: transforaminal, posterior, oblique, and direct lateral. When implanted through the direct lateral approach, the Halis™ Lumbar Cage System is only indicated for use in levels L2-L5. The Caro™ Cervical Cage System is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These patients should have had six weeks of non-operative treatment prior to surgery. Caro™ Cervical Cage System implants are placed via an anterior approach and is to be used with autogenous bone graft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The devices are intended to be used with supplemental fixation. The Lydia™ Anterior Lumbar Spinal System is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as discogence back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implanted via a laparoscopic or an open anterior approach. The devices are intended to be used with supplemental fixation. This device is not intended for cervical spine use. The Dica™ Direction Changeable Lumbar Cage System is indicated for interbody fusion with autogenous bone graft in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have six months of non-operative treatment prior to surgery. The devices are intended to be used with supplemental fixation.
Device Story
Shanghai Sanyou PEEK Cage System comprises four intervertebral body fusion device models (Halis, Caro, Lydia, Dica); used for spinal stabilization and fusion. Implants consist of PEEK cages with serrated surfaces for fixation; hollow geometry allows packing with autogenous or allogenic bone graft; embedded tantalum markers provide radiographic visualization. Devices implanted by surgeons in clinical settings via various approaches (transforaminal, posterior, oblique, direct lateral, laparoscopic, or open anterior) to support vertebral bodies during fusion. Output is structural support for the spine; clinical decision-making relies on radiographic confirmation of DDD and failure of non-operative treatment. Benefits include stabilization of spinal segments and promotion of bone fusion.
Clinical Evidence
Bench testing only. Static and dynamic axial compression, static shear compression, static torsion, and subsidence testing performed per ASTM F2077-14 and ASTM F2267-04. Results demonstrate performance meets or exceeds predicate devices.
Technological Characteristics
Materials: Polyetheretherketone (PEEK) and unalloyed tantalum (markers). Design: Hollow cage with serrated superior/inferior surfaces. Standards: ASTM F2077-14 (mechanical testing), ASTM F2267-04 (subsidence). Connectivity: None. Energy source: None. Sterilization: Not specified.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) requiring interbody fusion with supplemental fixation. Lumbar (L2-S1) indications include DDD with/without Grade 1 Spondylolisthesis/retrolisthesis; requires 6 months non-operative treatment. Cervical (C2-T1) indications include DDD with radicular symptoms; requires 6 weeks non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K133205 — CAPSTONE PTC SPINAL SYSTEM, CLYDESDALE PTC SPINAL SYSTEM · Medtronic Sofamor Danek USA, Inc. · Mar 13, 2014
K143163 — AVS® AL and ALign PEEK Spacers, AVS® PL and UniLIF PEEK Spacers, AVS® TL PEEK Spacer, AVS® Navigator PEEK Spacer, AVS® ARIA PEEK Spacer, AccuLIF TL and PL Cage, AVS® Anchor-L Spacer, Aero-AL Lumbar Cage System · Stryker Corporation · Jan 26, 2015
K150521 — Interbody Fusion (IBF)/Vertebral Body Replacement(VBR) System · Rti Surgical, Inc. Dba Rti Biologics · May 15, 2015
K170389 — TAURUS PEEK Cage System · Medyssey USA, Inc. · Dec 14, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring a symbol that resembles a caduceus or a representation of human figures.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 18, 2017
Shanghai Sanyou Medical Co, Ltd % Christine Scifert Executive Vice President MRC/X, LLC 6075 Poplar Avenue, Suite 500 Memphis, Tennessee 38119
Re: K163422
Trade/Device Name: Shanghai Sanyou PEEK Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP, MAX Dated: August 25, 2017 Received: August 30, 2017
Dear Ms. Scifert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163422
Device Name Shanghai Sanyou PEEK Cage System
#### Indications for Use (Describe)
The Halis™ Lumbar Cage System is indicated for use in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S 1 whose condition requires use of interbody fusion combined with supplemental fixation. These DDD patients may also have up to Grade 1 Spondylolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have had six months of non-operative treatment prior to surgery. The Halis™ Lumbar Cage System is intended to be used with autogenous bone graft comprised of cancellous and/or corticocancellous bone graft. The devices are intended to be used with supplemental fixation. The devices may be implanted through several surgical approaches: transforaminal, posterior, oblique, and direct lateral. When implanted through the direct lateral approach, the Halis™ Lumbar Cage System is only indicated for use in levels L2-L5.
The Caro™ Cervical Cage System is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These patients should have had six weeks of non-operative treatment prior to surgery. Caro™ Cervical Cage System implants are placed via an anterior approach and is to be used with autogenous bone graft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The devices are intended to be used with supplemental fixation.
The Lydia™ Anterior Lumbar Spinal System is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as discogence back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implanted via a laparoscopic or an open anterior approach. The devices are intended to be used with supplemental fixation. This device is not intended for cervical spine use.
The Dica™ Direction Changeable Lumbar Cage System is indicated for interbody fusion with autogenous bone graft in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have six months of non-operative treatment prior to surgery. The devices are intended to be used with supplemental fixation.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
Shanghai Sanyou PEEK Cage System September 8, 2017
| Company: | Manufacturing Facility and Headquarters:<br>Shanghai Sanyou Medical Co, LTD<br>1988 Jiatang Road<br>Jiading District, Shanghai, 201807, China |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Manufacturing Facility:<br>Shanghai Sanyou Medical Co, LTD<br>Rm 101/102/106/107<br>356 Renqing Rd, Building 3-1F,<br>Pudong New District, Shanghai 201201, China |
| Primary Contact: | Christine Scifert<br>Phone: 901-831-8053<br>christine.scifert@mrc-x.com |
| Company Contact: | David Fan, VP, Marketing<br>Phone: +86 21 58389980<br>Fax: +86 21 38682915<br>david.fan@sanyou-medical.com |
| Trade Name: | Shanghai Sanyou PEEK Cage System |
| Common Name: | Intervertebral Body Fusion Device with Bone Graft, Cervical<br>Intervertebral Body Fusion Device with Bone Graft, Lumbar |
| Classification: | Class II |
| Regulation Number: | 21 CFR 888.3080 (Intervertebral Body Fusion Device) |
| Panel: | 87- Orthopedic |
| Product Code: | ODP, MAX |
| Primary<br>Predicate Device: | K132897 – Medtronic Clydesdale® Spinal System |
| Additional Predicate<br>Devices: | K072791 - Depuy Synthes Spine, OPAL Spacer<br>K094025 – Medtronic Sofamor Danek, CRESCENT™ Spinal System<br>K151128 – Medtronic Sofamor Danek, CAPSTONE® Spinal System<br>K153373 – Medtronic Sofamor Danek, CORNERSTONE® PSR Cervical Fusion<br>System |
{5}------------------------------------------------
## Device Description:
The Shanghai Sanyou PEEK Cage System consists of four different models of fusion devices:
- Caro™ Cervical Cage System
- Halis™ Lumbar Cage System
- Lydia™ Anterior Lumbar Fusion System
- . Dica™ Direction Changeable Lumbar Cage System
All components are made of polyetheretherketone (PEEK). All fusion devices are intended for stabilization use and to promote bone fusion during the normal healing process following surgical correction of disorders of the cervical or lumbar spine. These Fusion Devices consist of PEEK cages of various heights, widths and lengths, which can be inserted between two lumbar or cervical vertebral bodies to give support and correction during intervertebral body fusion surgeries. All components have serrations on the superior and inferior surfaces of the implant to aid in fixation. The hollow geometry of the implants allows them to be packed with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.
Radiopaque Markers manufactured from unalloyed tantalum have been embedded within the implant to allow for visualization in radiographic images. Associated instruments are available to facilitate the implantation of device.
Several additional predicates and reference devices are identified and are used to support the substantial equivalence of the individual components of the Shanghai Sanyou PEEK Cage System as described in Table 1 below.
| Predicate Devices | | | |
|---------------------------------------|--------------------|--------------------------------------------------|---------------------------------------------------------------------|
| Predicate Device | Predicate<br>Usage | Subject Device Supported | Rationale for Support |
| K132897, Clydesdale<br>Spinal System | Lumbar | Halis™ Lumbar Cage System | Equivalent size range, anatomic<br>usage, and implantation approach |
| K072791, OPAL Spacer | Lumbar | Halis™ Lumbar Cage System | Equivalent size range, anatomic<br>usage, and implantation approach |
| K094025, Crescent<br>Spinal System | Lumbar | Dica™ Direction Changeable<br>Lumbar Cage System | Equivalent size range, anatomic<br>usage, and implantation approach |
| K151128, Capstone<br>Spinal System | Lumbar | Halis™ Lumbar Cage System | Equivalent size range, anatomic<br>usage, and implantation approach |
| K153373, Cornerstone<br>Spinal System | Cervical | Caro™ Cervical Cage System | Equivalent size range, anatomic<br>usage, and implantation approach |
| Reference Devices | | | |
| Reference Device | Reference<br>Usage | Subject Device Supported | Rationale for Support |
| K150765, ROI-C Fusion<br>System | Cervical | Caro™ Cervical Cage System | Equivalent size range, anatomic<br>usage, and implantation approach |
| K121982, Sovereign<br>Spinal System | Lumbar | Lydia™ Anterior Lumbar<br>Fusion System | Equivalent size range, anatomic<br>usage, and implantation approach |
#### Table 1. Predicate and Reference Devices
{6}------------------------------------------------
#### Indications for Use:
The Halis™ Lumbar Cage System is indicated for use in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1 whose condition requires use of interbody fusion combined with supplemental fixation. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have had six months of non-operative treatment prior to surgery. The Halis™ Lumbar Cage System is intended to be used with autogenous bone graft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The devices are intended to be used with supplemental fixation. The devices may be implanted through several surgical approaches: transforaminal, posterior, oblique, and direct lateral. When implanted through the direct lateral approach, the Halis™ Lumbar Cage System is only indicated for use in levels L2-L5.
The Caro™ Cervical Cage System is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These patients should have had six weeks of non-operative treatment prior to surgery. Caro™ Cervical Cage System implants are placed via an anterior approach and is to be used with autogenous bone graft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The devices are intended to be used with supplemental fixation.
The Lydia™ Anterior Lumbar Spinal System is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via a laparoscopic or an open anterior approach. The devices are intended to be used with supplemental fixation. This device is not intended for cervical spine use.
The Dica™ Direction Changeable Lumbar Cage System is indicated for interbody fusion with autogenous bone graft in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have six months of non-operative treatment prior to surgery. The devices are intended to be used with supplemental fixation.
## Performance Testing:
Static and dynamic axial compression, static shear compression, static torsion and were completed for the Shanghai Sanyou PEEK Cage System according the ASTM F2077-14 and Guidance for Industry and FDA Staff: Spinal System 510(k)s issued May 3, 2004. Additionally, subsidence testing was completed according to ASTM F2267-04. Performance testing demonstrates that the subject device meets or exceeds performance of predicate devices demonstrating that the subject devices are substantially equivalent to the predicate devices.
{7}------------------------------------------------
#### Substantial Equivalence:
The subject systems components are similar in sizes, materials and geometry to the predicate components. The subject components have the same indications as the predicate components. The subject devices are substantially equivalent to the predicate devices.
K132897 – Medtronic Clydesdale® Spinal System
K072791 – Depuy Synthes Spine, OPAL Spacer
K094025 – Medtronic Sofamor Danek, CRESCENT™ Spinal System
K151128 – Medtronic Sofamor Danek, CAPSTONE® Spinal System
K153373 – Medtronic Sofamor Danek, CORNERSTONE® PSR Cervical Fusion System
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.