← Product Code [OBT](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OBT) · K093442

# GUIDED GROWTH SYSTEM (K093442)

_Orthofix, Inc. · OBT · Jun 10, 2010 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OBT/K093442

## Device Facts

- **Applicant:** Orthofix, Inc.
- **Product Code:** [OBT](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OBT.md)
- **Decision Date:** Jun 10, 2010
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Guided Growth System plates are designed for the express and sole purpose of redirecting the angle of growth of long bone(s). This is useful for gradually correcting angular deformities in growing children. Specific conditions/diseases for which the device will be indicated include: valgus, varus or flexion, extension deformities of the knee (femur and/or tibia), valgus, varus or plantar flexion deformities of the ankle, valgus or varus deformities of the elbow (humerus), radial or ulnar deviation, flexion or extension deformities of the wrist (radius).

## Device Story

Guided Growth System Quad-Plate is a metallic bone fixation device for pediatric orthopedic use. It consists of a contoured, low-profile titanium alloy plate with a center hole for guide pin placement; attached to the bone surface over the physis using four screws. Screws are not locked, allowing them to swivel and diverge as bone grows; the implant functions as a flexible hinge to redirect growth and gradually straighten limbs. Used in clinical settings by surgeons for congenital or acquired deformities. Post-implantation, patients are permitted immediate mobility and weight-bearing. The device provides a mechanical means to correct angular deformities without fusion of the growth plate.

## Clinical Evidence

Bench testing only. Mechanical and functional testing of the Quad-Plate was conducted in accordance with ASTM F564-02 (Standard Specification and Test Method for Metallic Bone Staples) to demonstrate mechanical properties and performance equivalence to the predicate eight-Plate.

## Technological Characteristics

Materials: Titanium alloy (Ti6Al4V ELI) per ASTM F136. Design: Contoured, low-profile plate with four fixation points. Principle: Flexible hinge for guided growth. Energy: None (mechanical). Connectivity: None. Sterilization: Non-sterile, requires sterilization prior to use.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Orthofix Guided Growth System eight-Plate ([K031493](/device/K031493.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 1/2 K093442

## SUMMARY OF SAFETY AND EFFECTIVENESS

•

| Summary Preparation<br>Date: | June 10, 2010                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |             |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| 510(k) Submitter:            | Mary E. Biggers, RAC<br>Orthofix Inc. 1720 Bray Central Drive<br>McKinney, TX 75609<br>469-742-2500                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | JUN 10 2010 |
| Primary Contact:             | Darla Chew<br>Directory, Regulatory Affairs<br>Orthofix Inc.<br>Phone: 469-742-8824<br>Fax: 469-742-2556<br>darlachew@orthofix.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |             |
| Device Trade Name:           | Guided Growth System™                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |             |
| Device Common<br>Name:       | bone plate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |             |
| Regulation Name:             | Single/multiple component metallic bone fixation appliances and accessories.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |             |
| Classification Name:         | plate, bone, growth control, pediatric, epiphysiodesis<br>(21 CFR Parts 888.3030)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |             |
| Product Code:                | OBT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |             |
| Indications for Use:         | The Guided Growth System plates are designed for the express and sole<br>purpose of redirecting the angle of growth of long bone(s). This is useful for<br>gradually correcting angular deformities in growing children. Specific<br>conditions/diseases for which the device will be indicated include: valgus, varus<br>or flexion, extension deformities of the knee (femur and/or tibia), valgus, varus<br>or plantar flexion deformities of the ankle, valgus or varus deformities of the<br>elbow (humerus), radial or ulnar deviation, flexion or extension deformities of<br>the wrist (radius). |             |
| Predicate Device:            | The Orthofix Inc. Guided Growth System Quad-Plate is substantially equivalent<br>in design, function, and intended use to the Orthofix Guided Growth System<br>eight-Plate®. The Orthofix Guided Growth System eight-Plate, manufactured by<br>Orthofix Inc. of McKinney, Texas, was originally cleared by FDA under K031493<br>on November 20, 2003.                                                                                                                                                                                                                                                    |             |

1

{1}------------------------------------------------

The Guided Growth System is designed for the gradual correction of pediatric Device Description: deformities in both the upper and lower extremities. The device can be used for correction of congenital deformities as well as correction of acquired deformities, provided that the physis (growth plates) are not fused. The plates feature a contoured waist and low profile for pediatric usage. There is a center hole in the plate for a temporary guide pin to be implanted to ensure accurate application of the plate. The plates are attached to the external surface of the bone over the growth plate by two or four screws. These screws are not locked to the plate, but rather are allowed to swivel and diverge in their position as bone growth occurs. The implant acts like a flexible hinge, permitting growth at the growth plate to gradually straighten the limb. Immediately after implantation, the patient is allowed mobility and weight bearing.

| Biomechanical<br>Testing:   | In order to demonstrate that the Guided Growth System Quad-Plate has the<br>mechanical properties necessary to perform its intended use, and that the Quad<br>Plate performs as well as or better than the predicate device, Orthofix has<br>conducted mechanical and functional testing of the Quad-Plate in accordance<br>with ASTM F564-02 Standard Specification and Test Method for Metallic Bone<br>Staples. The testing was successfully completed demonstrating the Quad Plate<br>performs as well as the Guided Growth System eight-Plate. |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Material:                   | The Guided Growth System eight-Plate, Quad Plate and screws are made from<br>titanium alloy, Ti6AL-4V ELI conforming to ASTM F136.                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Sterilization:              | The Guided Growth eight-Plate, Quad-Plate and Bone Screws are supplied<br>NON-STERILE and require sterilization prior to use.                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Substantial<br>Equivalence: | The Guided Growth System Quad Plate is substantially equivalent in design                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |

and function to the Guided Growth System eight-Plate. The Guided Growth System eight-Plate received 510(k) clearance under K031493 on 11-20-03.

| Features                     | eight-Plate®                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Quad-Plate     |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| Indications for<br>Use       | "The Guided Growth Plate is designed for the express and sole<br>purpose of redirecting the angle of growth of long bone(s). This is<br>useful for gradually correcting angular deformities in growing<br>children. Specific conditions/diseases for which the device will be<br>indicated include: valgus, varus or flexion, extension deformities of<br>the knee (femur and/or tibia), valgus, varus or plantar flexion<br>deformities of the ankle, valgus or varus deformities of the elbow<br>(humerus), radial or ulnar deviation, flexion or extension deformities<br>of the wrist (radius)" | Identical      |
| Material                     | Ti6A14V ELI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Identical      |
| Features                     | eight-Plate®                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Quad-Plate     |
| Fixation Method              | Bone Screws                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Identical      |
| Number of<br>Fixation Points | Two                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Four           |
| Design Features              | a contoured waist and low profile for pediatric usage                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Identical      |
| Size Ranges<br>(lengths)     | 12mm and 16 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 16 mm and 22mm |

{2}------------------------------------------------

K09 3442 3/3

- Based upon the results of biomechanical testing the Guided Growth System Quad-Conclusion: Plate has the mechanical properties to perform its indications for use and is considered to be substantially equivalent to the predicate device in design, material and intended use.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, with three stylized lines forming the body and wings.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Orthofix, Inc. % Ms. Mary Biggers 1720 Bray Central Drive McKinney, Texas 75069

JUN 1 0 2010

Re: K093442

Trade/Device Name: Guided Growth System Quad-Plate Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: OBT Dated: May 10, 2010 Received: May 12, 2010

Dear Ms. Biggers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendmonts, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Dr w and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbrandi, nound adulteration. Please note: CDRH does not evaluate information related to contractly, and warranties. We remind you, however, that device labeling must be truthful and not mucleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements not the Act or any Federal statutes and regulations administered by other Federal agencies. Your must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{4}------------------------------------------------

Page 2 - Ms. Mary Biggers

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Dallard Melker
Mark N. Melkerson

Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## INDICATION FOR USE STATEMENT

| Page |  |
|------|--|
| of   |  |

K093442

510(k) Number (if known):

Device Name:

"Guided Growth System Quad-Plate" (bone plate)

Indications for Use:

The Guided Growth Plates are designed for the express and sole purpose of redirecting the angle of growth of long bone(s). This is useful for gradually correcting annyllar deformities in growing children. Specific conditions/diseases for which the device will be indicated include: valgus, varus or flexion, extension deformities of the knee (fiem wr and/or tibia), valgus, varus or plantar flexion deformities of the ankle, valgus or varus deformities of the elbow (humerus), radial or ulnar deviation, flexion or extension deformitus of the wrist (radius)

Prescription Use: X (Per 21 CFR 801.109)

Or

Over-The-Counter (Optional Format 1-2-96)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Santur for mxn
(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K093442

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OBT/K093442](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/OBT/K093442)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
