← Product Code [NQW](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NQW) · K231232

# Curiteva Laminoplasty System (K231232)

_Curiteva, Inc. · NQW · Jun 28, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NQW/K231232

## Device Facts

- **Applicant:** Curiteva, Inc.
- **Product Code:** [NQW](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NQW.md)
- **Decision Date:** Jun 28, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3050
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Curiteva Laminoplasty System is intended for use in the lower cervical and upper thoracic spine (C3 to T3) in laminoplasty procedures. The Curiteva Laminoplasty System is used to hold or buttress the allograft or autograft material in place in order to prevent the graft material from expulsion or impinging the spinal cord.

## Device Story

Internal fixation system for spinal surgery; consists of plates and screws in various configurations/sizes to accommodate patient anatomy. Plates feature pre-drilled holes for bone screws; hinge plates provided for additional stabilization. Used to hold/buttress allograft or autograft material during laminoplasty procedures in C3-T3 spine. Operated by surgeons in clinical/OR settings. Provides mechanical stabilization to prevent graft expulsion or spinal cord impingement.

## Clinical Evidence

Bench testing only. Performed static and dynamic four-point bending per ASTM F2193 and axial screw pullout per ASTM F543. Results demonstrate performance equivalent to primary predicate.

## Technological Characteristics

Components manufactured from Titanium alloy (Ti-6Al-4V ELI) per ASTM F136 or PEEK per ASTM F2026. System consists of plates and screws. Mechanical fixation device.

## Regulatory Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

## Predicate Devices

- Synthes Arch Fixation System ([K032534](/device/K032534.md))
- Choice Spine Laminoplasty Fixation System ([K173215](/device/K173215.md))
- Globus Medical Canopy Laminoplasty Fixation System ([K121732](/device/K121732.md))
- Nuvasive Laminoplasty System ([K091623](/device/K091623.md))

## Reference Devices

- Curiteva Cervical Interbody Fusion System ([K181261](/device/K181261.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

June 28, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Curiteva, Inc. Eric Linder Chief Technology Officer 25127 Will McComb Drive Tanner, Alabama 35671

Re: K231232

Trade/Device Name: Curiteva Laminoplasty System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal Interlaminal Fixation Orthosis Regulatory Class: Class II Product Code: NQW Dated: April 28, 2023 Received: April 28, 2023

Dear Eric Linder:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Image /page/1/Picture/5 description: The image contains the text "Colin O'neill -S" in black font. To the right of the text is a faded blue logo that appears to be "FDA". The text is arranged in two lines, with "Colin" on the first line and "O'neill -S" on the second line. The logo is partially obscured by the text.

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K231232

Device Name Curiteva Laminoplasty System

#### Indications for Use (Describe)

The Curiteva Laminoplasty System is intended for use in the lower cervical and upper thoracic spine (C3 to T3) in laminoplasty procedures. The Curiteva Laminoplasty System is used to hold or buttress the allograft material in place in order to prevent the graft material from expulsion or impinging the spinal cord.

Type of Use (Select one or both, as applicable)

| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Curiteva. The logo consists of a geometric shape resembling a stylized flower or gear, with six petals or segments arranged in a circular pattern. Five of the segments are colored in a dark blue, while one segment is a lighter blue. Below the geometric shape, the word "CURITEVA" is written in a bold, sans-serif font, with the trademark symbol (™) next to it.

# 510(k) Summary

### A. Submitter Information

| Submitter:      | Curiteva, Inc.<br>25127 Will McComb Drive<br>Tanner, AL 35671<br>Phone: (256) 213-1057<br>Fax: (256) 213-1058 |
|-----------------|---------------------------------------------------------------------------------------------------------------|
| Contact Person: | Eric Linder<br>regulatory@curiteva.com                                                                        |
| Date Prepared:  | April 28, 2023                                                                                                |

### B. Device Information

| Trade Name:            | Curiteva Laminoplasty System                                                                                                                                                                                                                                                                                           |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:           | Spinal Interlaminal Fixation Orthosis                                                                                                                                                                                                                                                                                  |
| Classification Name:   | Orthosis, Spine, Plate, Laminoplasty, Metal                                                                                                                                                                                                                                                                            |
| Device Classification: | Class II (per 21 CFR 888.3050)                                                                                                                                                                                                                                                                                         |
| Product Code:          | NQW                                                                                                                                                                                                                                                                                                                    |
| Classification Panel:  | Division of Orthopedic Devices                                                                                                                                                                                                                                                                                         |
| Predicate Device(s):   | Primary: Synthes Arch Fixation System (K032534)<br>Additional: Choice Spine Laminoplasty Fixation System (K173215)<br>Additional: Globus Medical Canopy Laminoplasty Fixation System (K121732)<br>Additional: Nuvasive Laminoplasty System (K091623)<br>Reference: Curiteva Cervical Interbody Fusion System (K181261) |

## C. Device Description

The Curiteva Laminoplasty System is an internal fixation device for spinal surgery that consists of various configurations of plates and screws. The implant configurations are available in different types and sizes so that adaptations can be made to take into account pathology and individual patient anatomy. The plates come preformed with holes to receive bone screws. Screws are used to attach the plates to bone. System plate configurations may be used with allograft or autograft material. A hinge plate is provided when additional stabilization is necessary.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Corteva Agriscience. The logo features a stylized geometric shape resembling a flower or a stylized molecule, with several dark blue hexagonal elements arranged in a circular pattern. One of the hexagonal elements is colored in a lighter shade of blue. Below the geometric shape, the word "CURITEVA" is written in bold, sans-serif font, with a trademark symbol to the right of the word.

All system components are manufactured from Titanium alloy (Ti-6Al-4V ELI) that conforms to ASTM F136, or PEEK (Polyetheretherketone) as described by ASTM F2026.

# D. Indications for Use

The Curiteva Laminoplasty System is intended for use in the lower cervical and upper thoracic spine (C3 to T3) in laminoplasty procedures. The Curiteva Laminoplasty System is used to hold or buttress the allograft or autograft material in place in order to prevent the graft material from expulsion or impinging the spinal cord.

# E. Comparison of Technological Characteristics

As was established in this submission, the subject Curiteva Laminoplasty System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and to have similar technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, function, and range of sizes.

# F. Performance Data

Non-clinical testing performed on the Curiteva Laminoplasty System supports substantial equivalence to predicate devices. The following testing was performed:

- Static four-point bending per ASTM F2193 ●
- Dynamic four-point bending per ASTM F2193
- Axial screw pullout per ASTM F543

The results of non-clinical testing demonstrate that the strength and performance of the Curiteva Laminoplasty System is sufficient for its intended use and is substantially equivalent to the legally marketed primary predicate device, K032534.

# G. Conclusion

Based on the indications for use, technological characteristics, non-clinical performance testing, and comparison to predicate devices, the subject Curiteva Laminoplasty System has been shown to be substantially equivalent to legally marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NQW/K231232](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NQW/K231232)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
