← Product Code [NQP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NQP) · K061774

# NFIX II PEDICLE SCREW SYSTEM (K061774)

_N Spine, Inc. · NQP · Dec 13, 2006 · Orthopedic · SESU_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NQP/K061774

## Device Facts

- **Applicant:** N Spine, Inc.
- **Product Code:** [NQP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NQP.md)
- **Decision Date:** Dec 13, 2006
- **Decision:** SESU
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Like other pedicle screw systems of its type, when used as a pedicle screw fixation system in skeletally mature patients, the N Fix II System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, kyphosis, and failed previous fusion (pseudoarthrosis). In addition, when used as a pedicle screw fixation system, the N Fix II System is indicated in patients who are receiving fusions with autogenous graft only; who are having the device fixed or attached to the lumbar or sacral spine; and who are having the device removed after the development of a solid fusion mass.

## Device Story

N Fix II is a posterior spinal instrumentation system comprising polyaxial top-loading pedicle screws and connecting rods. Designed for surgical implantation by physicians to provide spinal segment immobilization and stabilization as an adjunct to fusion. Screws and rods are manufactured from titanium 6Al-4V ELI alloy; rods also incorporate synthetic polycarbonate urethane (PCU) polymer. The system is intended for use in the thoracic, lumbar, and sacral spine. It provides mechanical support to stabilize the spine during the fusion process, with the device intended for removal following the development of a solid fusion mass. The device is not indicated for spinal stabilization without fusion.

## Clinical Evidence

No clinical data. Substantial equivalence established via non-clinical bench testing performed in accordance with recognized standards.

## Technological Characteristics

Posterior metal/polymer spinal system. Materials: Titanium 6Al-4V ELI alloy and synthetic polycarbonate urethane (PCU). Design: Polyaxial top-loading pedicle screws (5.5mm, 6.5mm, 7.5mm diameters; 35-60mm lengths) and 6.0mm diameter rods (40-200mm lengths). Mechanical fixation device.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Isobar System ([K990118](/device/K990118.md) to [K020245](/device/K020245.md))
- Moss Miami System ([K950697](/device/K950697.md) to [K030383](/device/K030383.md))
- Synergy System ([K950099](/device/K950099.md) to [K011437](/device/K011437.md))
- Optima System ([K024096](/device/K024096.md))
- N Fix System ([K053623](/device/K053623.md))
- DYNESYS® System ([K031511](/device/K031511.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K061774

#### 5.0 510(k) SUMMARY

In accordance with Title 21 of the Code of Federal Regulations Part 807 (21 CFR §807), and in particular §807.92, the following summary of safety and effectiveness information is provided:

## 5.1 Submitted By

N Spine. Inc. 6244 Ferris Square, Suite B, San Diego, California 92121-3239 Telephone: (858) 452-1266

R. Stephen Reitzler, Authorized Regulatory Agent Contact:

Date Prepared: June 19, 2006

## 5.2 Device Name

Trade or Proprietary Names: N Fix II Pedicle Screw System Common or Usual Name: Pedicle Screw System Classification Name: Pedicle Screw Spinal System Product Code: NOP Classification: 21 CFR, §888.3070

#### 5.3 Predicate Devices

The subject device is substantially equivalent to the following predicate devices:

- . Isobar System (Scient'x USA; K990118 to K020245)
- . Moss Miami System (DePuy Acromed; K950697 to K030383)
- . Synergy System (Interpore Cross; K950099 to K011437)
- . Optima System (U & I Corp .; K024096)
- N Fix System (N Spine; K053623) ●
- DYNESYS® System (Centerpulse Spine-Tech; K031511) .

#### 5.4 Device Description

The subject N Fix II device is a posterior instrumentation system consisting of both pedicle screws and connecting rods. Screws are of polyaxial top-loading design, are composed of titanium 6Al-4V ELI alloy, and are available in 5.5mm, 6.5mm, and 7.5mm diameters, in lengths ranging from 35mm to 60mm. Rods are composed of titanium 6Al-4V ELI alloy and synthetic polycarbonate urethane (PCU) polymer, are 6.0mm in diameter, and are available in lengths from 40mm to 200 mm.

{1}------------------------------------------------

## 5.5 Intended Use

Like other pedicle screw systems of its type, when used as a pedicle screw fixation system in skeletally mature patients, the N Fix II System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, kyphosis, and failed previous fusion (pseudoarthrosis).

In addition, when used as a pedicle screw fixation system, the N Fix II System is indicated in patients who are receiving fusions with autogenous graft only; who are having the device fixed or attached to the lumbar or sacral spine; and who are having the device removed after the development of a solid fusion mass.

# 5.6 Comparison to Predicate Devices

Testing and other comparisons have established that the subject N Fix II pedicle screw system is substantially equivalent in design, materials, indications and intended use, packaging, labeling, and performance to other predicate devices of the type currently marketed in the U.S.

### 5.7 Summary of Non-Clinical Tests

Nonclinical tests, including those conducted in accordance with recognized standards, have demonstrated the substantial equivalence of the subject device to commercially-available predicates in terms of performance.

### 5.8 Summary of Clinical Tests

(Not applicable)

# 5.9 Conclusions of Non-Clinical and Clinical Tests

The results of all testing demonstrated the substantial equivalence of the subject devices to the predicate devices.

.2/2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a bird-like figure, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NSpine % R. Stephen Reitzler, RAC Regulatory Agent 13221 Maricotte Place San Diego, California 92130

DEC 1 3 2006

Re: K061774

Trade Name: NFIX II Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Posterior metal/polymer spinal system Regulatory Class: Class II Product Code: NOP Dated: October 30, 2006 Received: October 31, 2006

Dear Mr. Reitzler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling:

"The safety and effectiveness of this device for use in the treatment of spinal stabilization without fusion have not been established."

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 

{3}------------------------------------------------

Page 2 - R. Stephen Reitzler, RAC

510(k) premarket notification if the limitation statement described above is added to your labeling.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,

If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International, and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

D nna-Bea Tillman, Ph.D., M.P.A. Director Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): K061774

Device Names: N Fix II Pedicle Screw System

Indications for Use:

When used as a pedicle screw fixation system in skeletally mature patients, the N Fix II System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, kyphosis, and failed previous fusion (pseudoarthrosis).

In addition, when used as a pedicle screw fixation system, the N Fix II System is indicated in patients:

- . who are receiving fusions with autogenous graft only:
- . who are having the device fixed or attached to the lumbar or sacral spine;
- who are having the device removed after the development of a solid . fusion mass.

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) (Division of General, Restorative, and Neurological Devices

510(k) Number K061774

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NQP/K061774](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NQP/K061774)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
