← Product Code [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB) · K203607

# MOSS VRS Spinal System (K203607)

_Biedermann Motech GmbH & Co. KG · NKB · Aug 20, 2021 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K203607

## Device Facts

- **Applicant:** Biedermann Motech GmbH & Co. KG
- **Product Code:** [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB.md)
- **Decision Date:** Aug 20, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Intended Use

The MOSS VRS Spinal System is intended to provide immobilization of spinal segments in sketally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine. The MOSS VRS Spinal System is intended for posterior, non-cervical pedicle fixation for the following indications: - degenerative disc disease (DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.) - spondylolisthesis - trauma (i. e., fracture or dislocation) - spinal stenosis - curvatures (i. e., scoliosis, kyphosis and/or lordosis) - tumor - pseudoarthrosis - failed previous fusion When used for posterior non-cervical pedicle screw fixation in pediatric patients, the MOSS VRS System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The MOSS VRS System is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

## Device Story

MOSS VRS Spinal System is a thoracolumbosacral posterior pedicle fixation system; provides immobilization/stabilization as adjunct to fusion. Components include screws (polyaxial, cannulated, fenestrated), polyaxial heads, rods, locking caps, connectors, and hooks. Used in OR by surgeons for spinal pathologies including DDD, trauma, and scoliosis. Implants are rigid, configured to patient anatomy, and secured to spine to facilitate fusion. Benefits include mechanical stabilization of spinal segments. Fenestrated screws are intended for use with saline or radiopaque dye only; not for bone cement. System is non-sterile, single-use.

## Clinical Evidence

Bench testing only. Mechanical performance evaluated per ASTM F1717 (static/dynamic compression, static torsion) and ASTM F1798 (static/dynamic flexion-extension, static axial grip, static axial pull-off). Results demonstrate substantial equivalence to predicate devices.

## Technological Characteristics

Materials: ASTM F136 titanium alloy (Ti-6Al-4V ELI) and ASTM F1537 cobalt chromium alloy. System consists of pedicle screws, rods, connectors, and hooks. Non-sterile, single-use. Mechanical fixation via polyaxial screw-rod construct.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- MOSS VRS Spinal System ([K181821](/device/K181821.md))
- EXPEDIUM Spine System ([K131802](/device/K131802.md))
- Nuvasive® GSB Global Spinal Balance System ([K132014](/device/K132014.md))
- Nuvasive® Reline® System ([K160989](/device/K160989.md))
- MOSS 100 Spinal System ([K162232](/device/K162232.md))
- Synthes MATRIX System ([K120838](/device/K120838.md))
- Synthes Universal Spinal System ([K990745](/device/K990745.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 20, 2021

Biedermann Motech GmbH & Co. KG Gerd Federle Director of Quality Management Bertha-von Suttner-Strasse 23 Villingen-Schwenningen, Baden-Wuerttemberg 78054 Germany

Re: K203607

Trade/Device Name: MOSS VRS Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB, KWP Dated: July 23, 2021 Received: July 26, 2021

Dear Gerd Federle:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K203607

Device Name MOSS VRS Spinal System

Indications for Use (Describe)

The MOSS VRS Spinal System is intended to provide immobilization of spinal segments in sketally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.

The MOSS VRS Spinal System is intended for posterior, non-cervical pedicle fixation for the following indications: - degenerative disc disease (DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.)

- spondylolisthesis
- trauma (i. e., fracture or dislocation)
- spinal stenosis
- curvatures (i. e., scoliosis, kyphosis and/or lordosis)
- tumor
- pseudoarthrosis
- failed previous fusion

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the MOSS VRS System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The MOSS VRS System is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Type of Use (*Select one or both, as applicable*)

| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)     </span> </div> | <div> <span> <span style="font-size: 16px;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C)     </span> </div> |
|----------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K203607 510(k) Summary

# Submitter, Correspondent and Manufacturer

| Submitter Name:          | Biedermann Motech GmbH & Co. KG                                                                                           |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Submitter Address:       | Bertha-von-Suttner-Str. 23<br>78054 Villingen-Schwenningen<br>Germany                                                     |
| Contact Person:          | Gerd Federle<br>Director of Quality Management<br>Phone: +49 7720 8510-545<br>Fax: +49 7720 8510-66<br>qfe@biedermann.com |
| Date of Submission:      | 07/23/2021                                                                                                                |
| Manufacturer Name:       | Biedermann Motech GmbH & Co. KG                                                                                           |
| Manufacturer<br>Address: | Bertha-von-Suttner-Str. 23<br>78054 Villingen-Schwenningen<br>Germany                                                     |
| Device Identification    |                                                                                                                           |
| Device Trade Name:       | MOSS VRS Spinal System                                                                                                    |
| Device Common<br>Name:   | Pedicle Screw Spinal System                                                                                               |
| Device Class:            | Class II                                                                                                                  |
| Classification Name:     | Thoracolumbosacral Pedicle Screw<br>System, Spinal interlaminal fixation<br>orthosis                                      |
| Classification Code:     | NKB, KWP                                                                                                                  |
| Classification Panel:    | Orthopedic                                                                                                                |
| Regulation Number:       | 21 CFR 888.3070, 21 CFR 888.3050                                                                                          |
| Regulation Name:         | Thoracolumbosacral Pedicle Screw<br>System, Spinal interlaminal fixation<br>orthosis                                      |

{4}------------------------------------------------

#### Predicate Devices

| Primary Predicate:     | MOSS VRS Spinal System, Biedermann Motech GmbH & Co. KG (K181821)   |
|------------------------|---------------------------------------------------------------------|
| Additional Predicates: | EXPEDIUM Spine System, DePuy Spine, Inc. (K131802)                  |
|                        | Nuvasive® GSB Global Spinal Balance System, Nuvasive Inc. (K132014) |
|                        | Nuvasive® Reline® System, Nuvasive Inc. (K160989)                   |
|                        | MOSS 100 Spinal System, Biedermann Motech GmbH & Co. KG (K162232)   |
|                        | Synthes MATRIX System, Synthes Spine (K120838)                      |
|                        | Synthes Universal Spinal System, Synthes Spine (K990745)            |

# Device Description

The Biedermann Motech MOSS VRS Spinal System is a comprehensive thoracolumbosacral spinal system that offers posterior clinical solutions and is indicated for the treatment of significant mechanical instability or deformity of the spine which requires fusion with instrumentation.

The MOSS VRS System consists of a variety of screws, polyaxial heads, rods, locking caps, connectors, hooks and associated general instruments. The implant components are available in a variety of sizes and can be rigidly locked into a variety of different configurations to suit the individual pathology and anatomical conditions of the patient. The purpose of this submission is to add components like additional polyaxial heads, cannulated screws, fenestrated screws, screws with larger diameters and lengths and hooks and connectors to the MOSS VRS Spinal System (see K181821).

The safety and effectiveness of the fenestrated screws has not been established when used in conjunction with bone cement or for use in patients with poor bone quality (e.g., osteoporosis, osteopenia). This device is intended only to be used with saline or radiopaque dye.

All implants of the MOSS VRS System are sinqle-use only and the system is provided non sterile.

{5}------------------------------------------------

# Material

The implants are manufactured from ASTM F 136 implant grade titanium alloy Ti-6Al-4V ELI and cobalt chromium allov according to ASTM F1537.

# Indication for Use

The MOSS VRS Spinal System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.

The MOSS VRS Spinal System is intended for posterior, non-cervical pedicle fixation for the following indications:

- . degenerative disc disease (DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.)
- . spondylolisthesis
- 트 trauma (i. e., fracture or dislocation)
- 트 spinal stenosis
- curvatures (i. e., scoliosis, kyphosis and/or lordosis)
- 트 tumor
- . pseudoarthrosis
- 트 failed previous fusion

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the MOSS VRS System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The MOSS VRS System is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

# Technological Characteristics and Substantial Equivalence

As was established in this submission, the subject MOSS VRS Spinal System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States.

The subject device was shown to be substantially equivalent and have equivalent technological characteristics to its predicate devices through comparison in areas including design, labeling, intended use, material composition and function.

{6}------------------------------------------------

# Performance Data

The new additional components of the MOSS VRS Spinal System were evaluated per the FDA Guidance Document for Industry and FDA Staff: Spinal System 510(k)s.

The following mechanical testing was carried out to show safety and effectiveness of our system:

# Bench testing - mechanical

Mechanical testing according to ASTM F1717 (Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model) standard and to ASTM F1798 (Standard Test Method for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants) standard was performed in order to provide data to support a substantial equivalence determination. These tests were performed to support system related performance under consideration of well-established acceptance criteria. Performance testing included:

- ASTM F1717: .
	- Static Compression
	- Dynamic Compression ●
	- Static Torsion .
- = ASTM F1798:
	- . Static Flexion Extension
	- Dynamic Flexion Extension .
	- . Static Axial Grip
	- Static Axial Pull-Off .

Result: The results demonstrate that the MOSS VRS Spinal System is substantially equivalent to the identified predicate devices.

# Substantial Equivalence Summary and Conclusion

The subject MOSS VRS Spinal System can be considered substantially equivalent to legally marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K203607](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K203607)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
