← Product Code [MQP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP) · K231134

# VerteLoc Spinal System (K231134)

_Signature Orthopaedics Pty, Ltd. · MQP · Feb 13, 2024 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP/K231134

## Device Facts

- **Applicant:** Signature Orthopaedics Pty, Ltd.
- **Product Code:** [MQP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP.md)
- **Decision Date:** Feb 13, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The VerteLoc Spinal System is indicated for use as a cervical vertebral body replacement (VBR) fusion device in the cervical spine (C3-C7) in skeletally mature patients for partial or total replacement of diseased, collapsed or unstable vertebral body due to tumour, trauma (i.e., fracture), or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues. The VerteLoc Spinal System is to be used with bone graft material. The Vertel.oc Spinal System is intended for use with supplemental internal fixation that has been cleared by the FDA.

## Device Story

Modular vertebral body replacement (VBR) system for cervical spine; replaces diseased, collapsed, or unstable vertebral bodies. Comprises two end bodies and one mid-body; assembled to match patient-specific spinal deficits. Features inner hollow for autograft/allograft bone graft material. Used in clinical settings by surgeons; requires supplemental internal fixation. Provides structural support to spinal column; facilitates fusion. Benefits patients by restoring spinal stability following corpectomy or trauma.

## Clinical Evidence

Bench testing only. No clinical or animal data required. Testing included static and dynamic axial compression, static and dynamic torsion, and subsidence testing to verify performance.

## Technological Characteristics

Modular VBR system; PEEK OPTIMA LT1 (ASTM F2026); titanium pins (ASTM F136, ISO 5832-3); tantalum marker beads (ASTM F560, ISO 13782). Features snap-fit locking mechanism. Mechanical device; no software or energy source.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- DePuy Synthes Spine Bengal Stackable Cage System ([K190284](/device/K190284.md))

## Reference Devices

- K2M Santorini Corpectomy Cage Systems ([K180665](/device/K180665.md))
- Signus Medizintechnik GmbH Athlet VBR System ([K081332](/device/K081332.md))
- Signature Orthopaedics Arlington Cage ([K172020](/device/K172020.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out next to it. The full name reads "U.S. Food & Drug Administration".

February 13, 2024

Signature Orthopaedics Pty Ltd Declan Brazil, Ph.D. Managing Director 7 Sirius Road Lane Cove West, NSW 2066 Australia

Re: K231134

Trade/Device Name: VerteLoc Spinal System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: PLR Dated: November 30, 2023 Received: January 12, 2024

Dear Dr. Brazil:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Brent Showalter -S

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health 

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known) K231134

Device Name VertLoc Spinal System

#### Indications for Use (Describe)

The VerteLoc Spinal System is indicated for use as a cervical vertebral body replacement (VBR) fusion device in the cervical spine (C3-C7) in skeletally mature patients for partial or total replacement of diseased, collapsed or unstable vertebral body due to tumour, trauma (i.e., fracture), or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues.

The VerteLoc Spinal System is to be used with bone graft material. The Vertel.oc Spinal System is intended for use with supplemental internal fixation that has been cleared by the FDA.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

PSC Publishing Services (301) 443-6740

{3}------------------------------------------------

#### 5 510(K) SUMMARY

| Manufacturer:      | Signature Orthopaedics Pty Ltd<br>7 Sirius Road<br>Lane Cove, NSW 2066<br>Australia<br><br>Signature Orthopaedics Europe Ltd<br>Unit A, IDA Business & Technology Park Garrycastle<br>Athlone Westmeath N37 DY26<br>IRELAND                                                                       |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name: | Verteloc Spinal System                                                                                                                                                                                                                                                                            |
| Common Name:       | Vertebral Body Replacement Device – Cervical                                                                                                                                                                                                                                                      |
| Contact:           | Dr. Declan Brazil<br>Managing Director of Signature Orthopaedics                                                                                                                                                                                                                                  |
| Prepared By:       | Signature Orthopaedics Pty Ltd<br>7 Sirius Road<br>Lane Cove, NSW 2066<br>Australia<br>Phone: +61 (2) 9428 5181<br>Fax: +61 (2) 8456 6065                                                                                                                                                         |
| Date Prepared:     | Monday, October 9, 2023                                                                                                                                                                                                                                                                           |
| Classification:    | Class II as per 21 CFR 888.3060 Spinal Vertebral Body<br>Replacement Device (PLR)                                                                                                                                                                                                                 |
| Predicate Devices: | Primary Predicate Device<br>• DePuy Synthes Spine Bengal Stackable Cage System (K190284)<br><br>Reference Predicate Devices<br>• K2M Santorini Corpectomy Cage Systems (K180665)<br>• Signus Medizintechnik GmbH Athlet VBR System (K081332)<br>• Signature Orthopaedics Arlington Cage (K172020) |

## Device Description:

The Verteloc Spinal System is a modular vertebral body replacement (VBR) system intended to replace a diseased, collapsed, damaged or unstable cervical vertebral body following partial or total corpectomy. The VerteLoc device comprise of three components; two end bodies and one mid body, which when used in combination is used to replace a measured deficit in the patient's spine. The VerteLoc device is offered in a variety of footprints and heights to accommodate the needs of patients. 

{4}------------------------------------------------

When assembled, the central portion of the end and mid bodies create an inner hollow for the placement of autograft and/or autograft bone graft material. The VerteLoc device is manufactured from unreinforced PEEK OPTIMA LT1 as per ASTM F2026 with titanium pins as per ASTM F136 and ISO 5832-3 as well as tantalum marker beads as per ASTM F560 and ISO 13782.

The VerteLoc Spinal System VBR is to be used in conjunction with supplemental fixation systems.

## Indications for Use:

The VerteLoc Spinal System is indicated for use as a cervical vertebral body replacement (VBR) fusion device in the cervical spine (C3-C7) in skeletally mature patients for partial or total replacement of diseased, collapsed, damaged or unstable vertebral body due to tumour, trauma (i.e., fracture), or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues.

The VerteLoc Spinal System is to be used with bone graft material. The VerteLoc Spinal System is intended for use with supplemental internal fixation that has been cleared by the FDA.

## Summary of Technological Characteristics:

The VerteLoc Spinal System share the following technological characteristics as the predicate devices:

- . The VerteLoc Spinal System has the same intended use as the DePuy Synthes Bengal Stackable Cage System.
- . The VerteLoc Spinal System has a shared indications for use as the DePuv Synthes Bengal Stackable Cage System.
- The VerteLoc Spinal System Cages are manufactured from the same material as the DePuy Synthes Bengal Stackable Cage System as well as Signature Orthopaedics Arlington Cages with the exception of the titanium pins.
- . The VerteLoc Spinal System Cages have the same height range as the DePuy Synthes Bengal Stackable Cage System and has the available length and widths fall within the Depuy Synthes Bengal Stackable Cage System' range.
- o The Verteloc Spinal System Cages are packaged, transported and sterilised through the same method(s) as Signature Orthopaedics Arlington Cages.

The following are the technological differences between the VerteLoc Spinal System and the predicate devices:

- . The VerteLoc Spinal Cage utilises titanium pins to assist in initial fixation where the Depuy Synthes Bengal Stackable Cage System does not include additional titanium pins and utilises the surface finish on the outer cages top and bottom surface areas (the same as the subject device) only.
- . The VerteLoc Spinal Cage utilises a snap-fit lock between stackable cages where as the DePuy Synthes Bengal Stackable Cage System uses screw and nut fasteners.

## Performance Data:

Non-clinical testing and engineering evaluations were conducted to verify that the performance of the Verteloc Spinal System is adequate for anticipated in-vivo use. No animal or clinical testing was required to support substantial equivalence. Nonclinical testing carried out included:

{5}------------------------------------------------

- . Static Axial Compression Testing
- Static Torsional Testing
- Dynamic Axial Compression Testing
- Dynamic Torsional Testing
- . Subsidence Testing

The results of verification and validation analyses demonstrated the Verteloc Spinal System to be substantially equivalent to the predicate devices.

## Conclusion:

The subject devices are substantially equivalent to the predicate devices since it has the same intended use, indications for use, materials, design features, and sterilisation to either all or one of the predicate devices. Non-clinical testing support the substantial equivalence claim.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP/K231134](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP/K231134)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
