← Product Code [MNH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MNH) · K090477

# ALPINE PEDICLE SCREW SYSTEM (K090477)

_Sanacor · MNH · Jul 27, 2009 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MNH/K090477

## Device Facts

- **Applicant:** Sanacor
- **Product Code:** [MNH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MNH.md)
- **Decision Date:** Jul 27, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Alpine Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (Pseudoarthrosis). In addition, this device is a pedicle screw system indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to S1) with removal of the implants after the attainment of a solid fusion.

## Device Story

Alpine Pedicle Screw System is a spinal fixation device consisting of screws, rods, and associated instruments. Used by surgeons via posterior approach to insert pedicle screws into vertebral bodies for immobilization and stabilization of spinal segments. System provides mechanical support as an adjunct to fusion. Clinical benefit includes stabilization of acute instabilities or deformities. Device is intended for use in clinical settings by trained surgeons.

## Clinical Evidence

Bench testing only. Testing conducted in accordance with ASTM F1717 and F1798, including static compression bending, static torsion, dynamic compression bending, and axial grip. Results met or exceeded all product specifications.

## Technological Characteristics

Spinal fixation system comprising screws, rods, and instruments. Materials and design are consistent with predicate pedicle screw systems. Mechanical testing performed per ASTM F1717 and F1798. No software or electronic components.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Mesa Pedicle Screw System
- OvalTwist Pedicle Screw System ([K061577](/device/K061577.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KC90427

IJUL 2 7 2009

# 510(k) Summary

pl 082

.

| Submitter:                          | Sanacor LLC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                     | Mr. Michael Ensign, Regulatory Affairs Manager<br>Sanacor LLC<br>P.O. Box 1196<br>Pleasant Grove, UT 84062<br>Telephone: (801) 358-9986<br>Fax: (801) 785-9415                                                                                                                                                                                                                                                                                                                                   |
| Date Prepared:                      | February 19, 2009                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Trade Name:                         | Alpine Pedicle Screw System                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Classification, Name<br>and Number: | Class II<br>Pedicle Screw System<br>21 CFR 888.3070                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Product Code:                       | MNI and MNH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Predicate Device(s):                | The subject device is substantially equivalent to the following<br>devices:                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                     | Mesa Pedicle Screw System<br>Marketed and distributed by K2M, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                     | OvalTwist Pedicle Screw System (K061577)<br>Marketed and distributed by Signus Medical LLC                                                                                                                                                                                                                                                                                                                                                                                                       |
| Device Description:                 | The Alpine Pedicle Screw System is a spinal system that consists<br>of screws, rods, and associated instruments. Fixation is provided<br>by bone (pedicular) screws inserted into the vertebral body of the<br>spine using a posterior approach.                                                                                                                                                                                                                                                 |
| Intended Use:                       | The Alpine Pedicle Screw System is intended to provide<br>immobilization and stabilization of spinal segments in skeletally<br>mature patients as an adjunct to fusion in the treatment of the<br>following acute instabilities or deformities of the thoracic, lumbar,<br>and sacral spine: degenerative spondylolisthesis with objective<br>evidence of neurological impairment, fracture, dislocation,<br>scoliosis, kyphosis, spinal tumor, and failed previous fusion<br>(Pseudoarthrosis). |
|                                     | In addition, this device is a pedicle screw system indicated for the<br>treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S<br>vertebra in skeletally mature patients receiving fusion by                                                                                                                                                                                                                                                                                          |

.

{1}------------------------------------------------

autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to S1) with removal of the implants after the attainment of a solid fusion.

Mechanical testing of the subject device consisted of static compression bending, static torsion, dynamic compression bending, and axial grip. All testing was conducted in accordance with ASTM F1717 and F1798. The device performed as designed and met, or exceeded, all product specifications.

Sanacor LLC considers the Alpine Pedicle Screw System to be equivalent to the predicate devices listed above. This conclusion is based on the devices' similarities in function, design, materials, and indications for use.

5

Functional and Safety Testing:

Conclusion:

{2}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with outstretched wings, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## JUL 2 7 2009

Sanacor LLC % Mr. Michael Ensign Director of Engineering PO Box 1196 Pleasant Grove, UT, 84062

Re: K090477

Trade/Device Name: Alpine Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: II Product Code: MNH, MNI Dated: July 20, 2009 Received: July 22, 2009

Dear Mr. Ensign:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Michael Ensign

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bárbara Rivero

Mark N. Melkerson Director Division of General. Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known):

Device Name: Alpine Pedicle Screw System

Indications for Use:

The Alpine Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (Pseudoarthrosis).

In addition, this device is a pedicle screw system indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to S1) with removal of the implants after the attainment of a solid fusion.

Prescription Use X (Part 21 CFR 801 Subpart D)

. AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

\$\varepsilon\$ \$\sim\$ (EXT for axM)

б

(Division Sign-Off) Division of Surgical, Orthor care, and Restorauve Devices

510(k) Number K090477

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MNH/K090477](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MNH/K090477)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
