HYDRALOK SYSTEM

K051216 · Altiva Corp. · MNH · Jul 19, 2005 · Orthopedic

Device Facts

Record IDK051216
Device NameHYDRALOK SYSTEM
ApplicantAltiva Corp.
Product CodeMNH · Orthopedic
Decision DateJul 19, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The HydraLok System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine including degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion (pseudoarthrosis). In addition, when used as a pedicle screw fixation system, the HydraLok System is intended for skeletally mature patients with severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra, who are receiving fusion by autogenous bone graft only having the device attached to the lumbar and sacral spine (L3 and below), who are having the device removed after the development of a solid fusion.

Device Story

HydraLok System is a top-loading spinal fixation system consisting of titanium alloy screws, rods, and connectors. Used by surgeons in clinical settings to provide immobilization and stabilization of spinal segments as an adjunct to fusion. Implants are provided clean and non-sterile in various sizes. Device is attached to the spine to support bone fusion; intended for removal after solid fusion is achieved in severe spondylolisthesis cases. Benefits include stabilization of spinal segments in patients with acute/chronic instabilities or deformities.

Clinical Evidence

Bench testing only.

Technological Characteristics

Top-loading spinal fixation system; materials: titanium alloy; components: screws, rods, and connectors; provided clean and non-sterile; requires sterilization by the user.

Indications for Use

Indicated for skeletally mature patients requiring spinal immobilization and stabilization as an adjunct to fusion for thoracic, lumbar, and sacral spine conditions including degenerative spondylolisthesis with neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and pseudoarthrosis. Specifically indicated for severe spondylolisthesis (Grades 3 and 4) at L5-S1 when using autogenous bone graft with attachment at L3 or below.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K05/216 1/2 ## PREMARKET NOTIFICATION 510(K) SUMMARY JUL 1 9 2005 | Company: | Director RA/QA<br>Altiva Corporation<br>9800 Suite 1, Southern Pines Blvd<br>Charlotte, NC 28273<br>Telephone: 704/409-1754<br>Fax: 704/409-1771 | |------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Company Contact: | John Kapitan | | Date: | May 10, 2005 | | Trade Name: | HydraLok™ System | | Common Name<br>Classification:<br>FDA Product Code : | Pedicle Screw Spinal Fixation System<br>Orthopedics, 888.3070, Class II<br>MNI, MNH | | Device Description: | The HydraLok System is a top-loading spinal fixation<br>system including screws, rods, and connectors. The<br>titanium alloy components are provided clean and non-<br>sterile. Various sizes of the implants are provided. | | Intended Use: | The HydraLok System is intended to provide<br>immobilization and stabilization of spinal segments as<br>an adjunct to fusion in the treatment of the following<br>acute and chronic instabilities or deformities of the<br>thoracic, lumbar and sacral spine including<br>degenerative spondylolisthesis with objective evidence<br>of neurological impairment, fracture, dislocation,<br>scoliosis, kyphosis, spinal tumor and failed previous<br>fusion (pseudoarthrosis) | | | In addition, when used as a pedicle screw fixation<br>system, the HydraLok System is intended for skeletally<br>mature patients with severe spondylolisthesis (Grades 3<br>and 4) of the L5-S1 vertebra, who are receiving fusion<br>by autogenous bone graft only having the device<br>attached to the lumbar and sacral spine (L3 and below),<br>who are having the device removed after the<br>development of a solid fusion. | | Predicate Device: | Predicate device information is included. | {1}------------------------------------------------ KOS/216 2/2 Performance Data: Performance data were submitted to characterize the HydraLok System. : {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The logo is surrounded by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" in a circular arrangement. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## JUL 1 9 2005 Mr. John Kapitan Director RA/QA Altiva Corporation 9800-I Southern Pine Boulevard Charlotte, North Carolina 28273 Re: K051216 Trade/Device Name: Altiva HydraLokTM System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: II Product Code: MNH, MNI Dated: May 10, 2005 Received: May 12, 2005 Dear Mr. Kapitan: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becaon 310(s) pressed is substantially equivalent (for the indications felerenced above and nave decembergally marketed predicate devices marketed in interstate for use stated in the encrosule) to regars and the Medical Device Amendments, or to commerce prior to May 20, 1976, the excordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). alle Cosmetic Act (71ct) that do not requent of the general controls provisions of the Act. The You may, merciole, manol the Act include requirements for annual registration, listing of general controls provisions of the Free labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (300 a0070) als. Existing major regulations affecting your device can may be subject to such additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean I least be advised that I DT is lessants of the complies with other requirements of the Act that I Dri has intado a actuations administered by other Federal agencies. You must of any I cuttar statures and regaranents, including, but not limited to: registration and listing (21 Compry with an the Aot 8 requirements, 11); good manufacturing practice requirements as set CFR Part 807), adomig (21 OFF Part 820); and if applicable, the electronic forth in the quality bystelly (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Mr. John Kapitan This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to begin marketing your antial equivalence of your device to a legally premarket notheation. The PDA miding of bactification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your acon 276-0120. Also, please note the regulation entitled, comact the Office of Compuners as (2 + + n = (21CFR Part 807.97). You may obtain "Misbranding by reference to premanteenoulties. Souther the Act from the Division of Small other geteral information on your responsible and its toll-free number (800) 638-2041 or 1 Manufacturers, International and Consumer Prosistanter Prosis.org/vcdrh/dsma/dsmamain.html Sincerely yours, Miriam C. Provost Miriam C. Provost, Ph.D Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indication for Use Statement 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: Altiva HydraLok™ System Indications for Use: · ·- · · · · نخست تي تي تي تي The HydraLok System is intended to provide immobilization and stabilization The Hydracok System is intended to free treatment of the following of spinal segments as an adjunct to fusion in the treasing lymber ond socra of spinal segments as an adjunct to racion in the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological . impairment, - fracture, . - dislocation, . - scoliosis, . - kyphosis, . - spinal tumor and . - failed previous fusion (pseudoarthrosis) . In addition, when used as a pedicle screw fixation system, with acquere System is intended for skeletally mature patients with severe System - is - intended - for - Skolotally - marketra, who are receiving spondylolistiesis (Craces o ana f) or life drive the device attached to the fusion by adtogenous bone grait only having the device removed after the development of a solid fusion. Prescription Use _____________________________________________________________________________________________________________________________________________________________ or Over-The-Counter Use (21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) iriam C. Provost (Division Sign-Off) (Division of General, Restorative and Neurological Devices 510(k) Number K051216
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...