← Product Code [MNH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MNH) · K051044

# ARCTEC SEMIRIGID PLATING SYSTEM (K051044)

_Altiva Corp. · MNH · Oct 7, 2005 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MNH/K051044

## Device Facts

- **Applicant:** Altiva Corp.
- **Product Code:** [MNH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MNH.md)
- **Decision Date:** Oct 7, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The ArcTecTM Semi-Rigid Plating System is intended for posterior, noncervical, pedicle fixation in order to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine, including degenerative spondylolisthesis with objective evidence of neurological impairment; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudarthrosis). In addition, when used as a pedicle screw fixation system, the ArcTec™ Semi-Rigid Plating System is intended for skeletally mature patients with severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra, who are receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 and below), with removal of the implants after the attainment of solid fusion.

## Device Story

ArcTec Semi-Rigid Plating System consists of plates, bolts, washers, and locking nuts; used for posterior spinal fixation. Features low-profile, low-stiffness plate; patent-pending washer design with polyaxial bandclamp allows unlimited placement and 10-degree screw angulation. System combines plate simplicity with polyaxial screw adaptability. Used by surgeons trained in spinal decompression and bone grafting; intended for OR environment. Provides spinal segment immobilization to facilitate fusion. Benefits patients by stabilizing acute/chronic spinal instabilities or deformities.

## Clinical Evidence

Bench testing only. Mechanical performance evaluated in accordance with ASTM F1717 to demonstrate substantial equivalence.

## Technological Characteristics

Constructed from titanium alloy (Ti-6Al-4V per ASTM F136). System includes plates, bolts, washers, and locking nuts. Features polyaxial bandclamp design allowing 10-degree screw angulation. Single-use only. Mechanical performance validated per ASTM F1717.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Simmons Stainless Steel Plating System ([K930353](/device/K930353.md))
- Ultium Plating System ([K962784](/device/K962784.md))
- HydraLok System ([K051216](/device/K051216.md))
- Rogozinski Spinal System ([K983899](/device/K983899.md))
- Titanium Moss Miami Spinal System ([K964024](/device/K964024.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K a51044 p 1/2

### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (21CFR 807.92)

OCT 7 - 2005

- I. DATE PREPARED: October 03, 2005
- II. SUBMITTED BY: Altiva Corporation 9800 Southern Pines Blvd. Suite I Charlotte, NC 28273
- III. CONTACT PERSON: John Kapitan

Director of Regulatory Affairs & Quality Assurance (704) 409-1754

- IV. DEVICE NAME: Trade/Proprietary Name: ArcTecTM Semi-Rigid Plating System
- V. DEVICE CLASSIFICATION NAME: Pedicle screw spinal system 21 CFR 888.3070, Class II MNH, MNI
- PREDICATE DEVICE INFORMATION: VI.

Simmons Stainless Steel Plating System (K930353) Ultium Plating System (K962784) HydraLok System (K051216) Rogozinski Spinal System (K983899) Titanium Moss Miami Spinal System (K964024)

#### VII. DEVICE DESCRIPTION:

The ArcTec™ Semi-Rigid Plating System includes plates, bolts, washers and locking nuts in addition to standard surgical instrumentation.

{1}------------------------------------------------

KOSIO44 p½

The semi-rigid low profile plate has a low stiffness and is easy to contour. The patent pending washer design incorporates a unique polyaxial bandclamp feature that allows unlimited range of placement on the plate and screw angulation of up to 10 degrees in any plane. The system combines the simplicity of a plate with the adaptability of a polyaxial screw connection.

#### VIII. MATERIALS:

The implant components are manufactured from titanium alloy (Ti-6Al-4V per ASTM F136). The implant components are for single use only.

### IX. INDICATION FOR USE:

The ArcTec™ Semi-Rigid Plating System is intended for posterior, noncervical, pedicle fixation in order to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine, including degenerative spondylolisthesis with objective evidence of neurological impairment; fracture; dislocation; scoliosis; spinal turnor; and failed previous fusion (pseudarthrosis).

In addition, when used as a pedicle screw fixation system, the ArcTec™ Semi-Rigid Plating System is intended for skeletally mature patients with severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra, who are receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 and below), with removal of the implants after the attainment of solid fusion.

#### X. SAFETY INFORMATION:

Only surgeons trained and experienced in spinal decompression and bone grafting techniques should use the Arc Tec TM Semi-Rigid Plating System. In addition, thorough familiarization with the implants, instrumentation, and surgical technique is essential. Federal law (U.S.A.) restricts this device to sale by or on the order of a physician.

#### XI. PERFORMANCE DATA

Mechanical testing in accordance with ASTM F1717 was conducted to evaluate performance as a basis for substantial equivalence.

#### XII. CONCLUSION

Altiva believes sufficient information is included to reach a determination of substantial equivalence to the predicate devices based on intended use, function, design, material and mechanical performance.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo features a stylized eagle with three stripes forming its body and wings. The eagle is enclosed within a circle, and the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is written around the upper portion of the circle.

OCT 7 - 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

John Kapitan Director of Regulatory Affairs and Quality Altiva Corporation 9800 Southern Pine Boulevard, Suite 1 Charlotte, North Carolina 28273

Re: K051044

Trade/Device Name: ArcTec™ Semi-Rigid Plating System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: II Product Code: MNH, MNI Dated: September 9, 2005 Received: September 12, 2005

Dear Mr. Kapitan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2- John Kapitan

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to logal|y marketed predicate device results in a classification for your device and thus, permits wourderice to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation onitible, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html .

Sincerely yours,

2

Mark N. Melkerson, Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): KOS1044

Device Name: ArcTecTM Semi-Rigid Plating System

Indications for Use:

The ArcTecTM Semi-Rigid Plating System is intended for posterior, noncervical, pedicle fixation in order to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine, including degenerative spondylolisthesis with objective evidence of neurological impairment; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudarthrosis).

In addition, when used as a pedicle screw fixation system, the ArcTec™ Semi-Rigid Plating System is intended for skeletally mature patients with severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra, who are receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 and below), with removal of the implants after the attainment of solid fusion.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

"Caution: Federal law restricts this device to sale by or on the order of a "physician"

# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| (Division Sign-Off) |
|---------------------|
|---------------------|

Division of General, Restorative,
and Neurological Devices

| 510(k) Number | K051044 |
|---------------|---------|
|---------------|---------|

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MNH/K051044](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MNH/K051044)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
