K031265 · Spine Next SA · MNH · Oct 9, 2003 · Orthopedic
Device Facts
Record ID
K031265
Device Name
SHIRAZ-JAVA SPINAL FIXATION SYSTEM
Applicant
Spine Next SA
Product Code
MNH · Orthopedic
Decision Date
Oct 9, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3070
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When used as a pedicle screw fixation system, the SHIRAZ JAVA Spinal Fixation System is intended for treatment of severe spondylolisthesis (Grades 3 and 4) at the fifth lumbar - first sacral (L.5-S1) vertebral joint in skeletally mature patients receiving fusion by autogenous bone grafting having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. When used as a pedicle screw fixation system, the JAVA Spinal Fixation System is also intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).
Device Story
Spinal fixation system comprising screws, rods, and connectors; designed for surgical stabilization of thoracic, lumbar, and sacral spine segments during bone graft fusion. Intended for use by surgeons in clinical/OR settings. Implants provide temporary immobilization; removed after solid fusion mass development. Benefits include structural support for spinal correction and fusion in patients with severe instabilities or deformities.
Clinical Evidence
Bench testing only. Biomechanical static and dynamic testing performed in accordance with ASTM F1717 and ASTM F1798.
Technological Characteristics
Materials: Titanium Alloy (Ti6A4V) per ASTM F136. Components: Screws, rods, and connectors. Mechanical fixation system; no energy source or software.
Indications for Use
Indicated for skeletally mature patients requiring spinal fusion for severe spondylolisthesis (Grades 3-4) at L5-S1, or acute/chronic instabilities/deformities of the thoracic, lumbar, and sacral spine including degenerative spondylolisthesis with neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and pseudarthrosis.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
{0}------------------------------------------------
031265 14
# OCT - 9 2003
# 510(k) SUMMARY
## SUBMITTED BY
Diane Johnson On behalf of Spine Next America 4720 Salisbury Road Suite 135 Jacksonville, FL 32256
April 14, 2003 Date Submitted:
## CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME
Trade/Proprietary Name:
Common/Usual Name: Classification Name:
Spine Next SHIRAZ JAVA Spinal Fixation System Pedicle Screw System Orthosis, Spondylolisthesis Spinal Fixation or Orthosis, Spinal Pedicle Fixation
## PREDICATE DEVICE
Moss Miami [reference 510(k) K021880 cleared June 26, 2002]
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92.
## DEVICE DESCRIPTION AND MATERIALS OF CONSTRUCTION
The Spine Next SHIRAZ JAVA Spinal Fixation System is designed to aid the surgical correction of several types of spinal conditions. This system is intended only to provide stabilization during the development of a solid fusion with a bone graft. These implants are intended to be removed after development of a solid fusion mass. The system includes screws, rods, and connectors.
All implants are manufactured from Titanium Alloy (Ti6A4V) meeting the requirements of ASTM F136.
## INDICATIONS FOR USE
When used as a pedicle screw fixation system, the SHIRAZ JAVA Spinal Fixation System is intended for treatment of severe spondylolisthesis (Grades 3 and 4) at the fifth lumbar - first sacral (L.5-S1) vertebral joint in skeletally mature patients receiving fusion
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains handwritten text. The text on the top line reads "k0 31 265". The text on the bottom line reads "2012".
by autogenous bone grafting having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
When used as a pedicle screw fixation system, the JAVA Spinal Fixation System is also intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).
### PERFORMANCE DATA
Biomechanical testing, including static and dynamic testing, was performed in accordance with ASTM F1717 and ASTM 1798.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features an eagle with its wings spread, symbolizing the agency's mission to protect the health of all Americans. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT - 9 2003
Spine Next SA c/o Ms. Diane Johnson Director, Regulatory Affairs Spine Next America Corporation 104 Greenwood Creek Road Queenstown, MD 21658
Re: K031265
Trade/Device Name: SHIRAZ JAVA Spinal Fixation System Regulation Number: 21 CFR 888.3050, 888.3070, Regulation Name: Spinal interlaminal fixation orthosis, Pedicle screw spinal system Regulatory Class: II Product Code: MNH, MNI Dated: July 24, 2003 Received: July 25, 2003
Dear Ms. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Diane Johnson
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594- 4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sincerely, yours,
Mark A. Milliken
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): ___________________________________________________________________________________________________________________________________________________
Device Name: Spine Next SHIRAZ JAVA Spinal Fixation System
## Indications for Use:
NEEDED)
When used as a pedicle screw fixation system, the SHIRAZ JAVA Spinal Fixation System is intended for treatment of severe spondylolisthesis (Grades 3 and 4) at the fifth lumbar - first sacral (L5-S1) vertebral joint in skeletally mature patients receiving fusion by autogenous bone grafting having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
When used as a pedicle screw fixation system, the JAVA Spinal Fixation System is also intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (PER 21 CFR 801.109) OR
#### Over-The-Counter Use
Optional Format 1-2-96
Mark N. Milken
Divisio ign-Off Division General. Restorative and Neurological Devices
510(k) Number K031265
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.