K013962 · Medtronic Sofamor Danek, Inc. · MNH · Jun 20, 2002 · Orthopedic
Device Facts
Record ID
K013962
Device Name
EQUATION FIXATION SYSTEM
Applicant
Medtronic Sofamor Danek, Inc.
Product Code
MNH · Orthopedic
Decision Date
Jun 20, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3070
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EQUATION™ Fixation System is a temporary implant system used for correction and stabilization of the posterior spine for the development of a solid spinal fusion. When used as a pedicle screw fixation system of the non-cervical posterior spine in skeletally mature patients, the EQUATION™ Fixation System is indicated for one or more of the following: (1) degenerative spondylolisthesis with objective evidence of neurologic impairment, (2) fracture, (3) dislocation, (4) scoliosis, (5) kyphosis, (6) spinal tumor, and/or (7) failed previous fusion (pseudarthrosis). In addition, when used as a pedicle screw fixation system, the EQUATION™ Fixation System is indicated for skeletally mature patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of a solid fusion mass.
Device Story
EQUATION™ Fixation System is a temporary posterior spinal implant system; consists of titanium alloy screws, nuts, 3.6mm rods, and cross connectors. Used by surgeons in clinical settings to provide stabilization during spinal fusion procedures. Components are configured and locked to create a construct tailored to individual patient anatomy. System provides mechanical support to the spine to facilitate fusion; removed after solid fusion mass development. Benefits include correction and stabilization of the spine in patients with degenerative conditions, fractures, or deformities.
Clinical Evidence
Bench testing only. Mechanical testing was performed and compared to test data from the previously cleared CD HORIZON® Spinal System to demonstrate substantial equivalence.
Technological Characteristics
Materials: Titanium alloy (Ti-6Al-4V) per ASTM F136 or ISO 5832-3. Components: Screws, nuts, 3.6mm rods, cross connectors. Design: Rigidly lockable construct. Connectivity: None. Energy source: None. Sterilization: Not specified.
Indications for Use
Indicated for skeletally mature patients requiring posterior spinal stabilization/fusion for degenerative spondylolisthesis with neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, or pseudarthrosis. Also indicated for severe L5-S1 spondylolisthesis (Grades 3-4) using autogenous bone graft, with fixation at L3 or below.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
Predicate Devices
CD HORIZON® Spinal System
Related Devices
K033495 — MODIFICATION TO EQUATION FIXATION SYSTEM · Medtronic Sofamor Danek, Inc. · Dec 3, 2003
K042453 — MODIFICATION TO EQUATION FIXATION SYSTEM · Medtronic Sofamor Danek, Inc. · Oct 7, 2004
K070367 — ENDIUS TITLE 2 POLYAXIAL SPINAL SYSTEM · Endius, Inc. · Apr 12, 2007
K103091 — LANX SPINAL FIXATION SYSTEM · Lanx, Inc. · Apr 15, 2011
{0}------------------------------------------------
K013962 p½
# Medtronic Sofamor Danek EQUATION™ Fixation System 510(k) Summary May 2002
| Submitter: | Medtronic Sofamor Danek USA, Inc.<br>1800 Pyramid Place<br>Memphis, TN 38132 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Richard Treharne<br>Sr. Vice President Regulatory Affairs |
| Trade Name: | EQUATION™ Fixation System |
| Classification Name: | Spinal Intervertebral Body Fixation Orthosis, Class II |
| Predicate Device(s): | The EQUATION™ Fixation System is substantially equivalent to the CD<br>HORIZON® Spinal System. |
| Device Description: | The Medtronic Sofamor Danek EQUATION™ Fixation System consists of a<br>variety of shapes and sizes of screws, nuts, and 3.6mm rods and cross<br>connectors. The implant components can be rigidly locked in a variety of<br>configurations, with each construct being tailor-made for the individual case.<br>The implants are made of titanium alloy (Ti-6Al-4V) described by ASTM<br>Standard F136 or ISO 5832-3. Stainless steel and titanium implant components<br>must not be used together in a construct. |
| Intended Use: | The EQUATION™ Fixation System is a temporary implant system used for<br>correction and stabilization of the posterior spine for the development of a<br>solid spinal fusion. When used as a pedicle screw fixation system of the non-<br>cervical posterior spine in skeletally mature patients, the EQUATION™<br>Fixation System is indicated for one or more of the following: (1) degenerative<br>spondylolisthesis with objective evidence of neurologic impairment, (2)<br>fracture, (3) dislocation, (4) scoliosis, (5) kyphosis, (6) spinal tumor, and/or (7)<br>failed previous fusion (pseudarthrosis).<br>In addition, when used as a pedicle screw fixation system, the EQUATION™<br>Fixation System is indicated for skeletally mature patients: (a) having severe<br>spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral (L5-S1)<br>vertebral joint; (b) who are receiving fusions using autogenous bone graft only;<br>(c) who are having the device fixed or attached to the lumbar and sacral spine |
{1}------------------------------------------------
CO i3962 p²/2
(L3 and below); and (d) who are having the device removed after the development of a solid fusion mass.
Functionality &
Mechanical testing was performed on the EQUATION™ Fixation System and Safety Testing: was compared to test data on the previously cleared CD HORIZON® Spinal System. The test results were provided in this submission.
The EQUATION™ Spinal System is substantially equivalent to the CD Conclusion: HORIZON® Spinal System.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an eagle or bird in flight, composed of three curved lines that suggest the wings and body of the bird.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 2-0-2002
Richard W. Treharne, Ph.D. Senior Vice President, Research and Regulatory Affairs Medtronic Sofamor Danek 1800 Pyramid Place Memphis. Tennessee 38132
Re: K013962
Trade/Device Name: EQUATION" Fixation System Regulatory Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: II Product Code: MNH, MNI Dated: April 16, 2002 Received: April 17, 2002
Dear Dr. Treharne:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Richard W. Treharne, Ph.D.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
R. Mark McMellon
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## 510(k) Number (if known): KO13962-
### Device Name: EQUATION™ Fixation System
### Indications for Use:
The EQUATION™ Fixation System is a temporary implant system used for correction and stabilization of the posterior spine for the development of a solid spinal fusion. When used as a pedicle screw fixation system of the non-cervical posterior spine in skeletally mature patients, the EQUATION™ Fixation System is indicated for one or more of the following: (1) degenerative spondylolisthesis with objective evidence of neurologic impairment, (2) fracture, (3) dislocation, (4) scoliosis, (6) spinal tumor, and/or (7) failed previous fusion (pseudarthrosis).
In addition, when used as a pedicle screw fixation system, the EQUATION™ Fixation System is indicated for skeletally mature patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of a solid fusion mass.
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Evaluation (ODE)
| Prescription Use<br>(Per 21 CFR 801.109) | |
|------------------------------------------|--|
|------------------------------------------|--|
OR
| Over-The-Counter Use | (Optional 1-2-96) |
|----------------------|-------------------|
|----------------------|-------------------|
General, Restorative and Neurological Devices 510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.