← Product Code [MNH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MNH) · K012273

# SPIRAL RADIUS 90-D SST SYSTEM (K012273)

_United States Surgical, A Division of Tyco Healthc · MNH · Aug 15, 2001 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MNH/K012273

## Device Facts

- **Applicant:** United States Surgical, A Division of Tyco Healthc
- **Product Code:** [MNH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MNH.md)
- **Decision Date:** Aug 15, 2001
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The components of the Spiral Radius 90-D™ SST System are indicated for spinal fixation. When used as an anterolateral/anterior system consisting of rods ands screws, the levels of attachment are the lumbar, thoracic and cervical spine. The points of attachment and methods are screw fixation to the anterolateral vertebral bodies of the lumbar and thoracic spine (T6-L5) and the anterior vertebral bodies of the cervical spine. The Indications are degenerative disc disease defined as back pain of discogenic origin with degeneration of the history and radiographic studies. confirmed ه disc spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion (pseudoarthrosis). When used as a nonpedicle posterior system consisting of hooks, crosslinks and sacral/iliac screws the levels of attachments are the lumbar, thoracic and cervical spine and sacrum and ilium. Intended uses include hook and sacral screw fixation to the lumbar spine, noncervical spine and to the T1-S1 spine; and hook and sacral/iliac screw fixation to the noncervical spine. The indications are degenerative disc disease defined as back pain of discogenic origin with degenerative of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumors, pseudoarthrosis, or failed previous fusion (pseudoarthrosis). When used as a pedicle screw system, in the non-cervical spine of skeletally mature patients, The Spiral Radius 90-D™ SST System is indicated for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of solid fusion mass. In addition, when used as a pedicle screw system, in the non-cervical spine of skeletally mature patients, the Spiral Radius 90-D™ SST System is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in the treatment of the following acute and chronic instabilities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

## Device Story

Stainless steel spinal rod system; components include rods, screws, hooks, and crosslinks. Used for spinal fixation, immobilization, and stabilization. Implanted by surgeons in clinical settings to treat spinal instabilities and deformities. Provides mechanical support to spinal segments to facilitate fusion. Benefits include stabilization of the spine in patients with degenerative conditions, fractures, or deformities.

## Clinical Evidence

No clinical data provided; substantial equivalence based on design and material characteristics.

## Technological Characteristics

Medical grade stainless steel conforming to ASTM F138. System components include rods, screws, hooks, and crosslinks. Mechanical fixation device.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Spiral Radius 90-D™ System ([K992784](/device/K992784.md))
- Spiral Radius 90-D™ System ([K000009](/device/K000009.md))
- Rogozinski Spinal Rodding System ([K954696](/device/K954696.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### Summary of Safety and Effectiveness IX.

United States Surgical SUBMITTER: A division of Tyco Healthcare Group, LP 150 Glover Avenue Norwalk, CT 06856

Luis Nesprido CONTACT PERSON:

July 10, 2001 DATE PREPARED:

Spinal Intervertebral Body Fixation Orthosis CLASSIFICATION NAME: Spinal Interlaminal Fixation Orthosis Spondylolisthesis Spinal Fixation Device System

Stainless Steel Spinal Rod System COMMON NAME:

Spiral Radius 90-D™ SST System PROPRIETARY NAME:

USS's Spiral Radius 90-D™ System – K992784 & PREDICATE DEVICES: K000009 & the Rogozinski Spinal Rodding System -K954696

The components of the Spiral Radius 90-D™ SST System are INTENDED USE: indicated for spinal fixation.

> When used as an anterolateral/anterior system consisting of rods ands screws, the levels of attachment are the lumbar, thoracic and cervical spine. The points of attachment and methods are screw fixation to the anterolateral vertebral bodies of the lumbar and thoracic spine (T6-L5) and the anterior vertebral bodies of the cervical spine. The Indications are degenerative disc disease defined as back pain of discogenic origin with degeneration of the history and radiographic studies. confirmed ه disc spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion (pseudoarthrosis).

> When used as a nonpedicle posterior system consisting of hooks, crosslinks and sacral/iliac screws the levels of attachments are the lumbar, thoracic and cervical spine and sacrum and ilium. Intended uses include hook and sacral screw fixation to the lumbar spine, noncervical spine and to the T1-S1 spine; and hook and sacral/iliac screw fixation to the noncervical spine. The indications are degenerative disc disease defined as back pain of discogenic origin with degenerative of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumors, pseudoarthrosis, or failed previous fusion (pseudoarthrosis).

> When used as a pedicle screw system, in the non-cervical spine of skeletally mature patients, The Spiral Radius 90-D™ SST

{1}------------------------------------------------

System is indicated for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of solid fusion mass.

In addition, when used as a pedicle screw system, in the noncervical spine of skeletally mature patients, the Spiral Radius 90-D™ SST System is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in the treatment of the following acute and chronic instabilities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

MATERIALS: The material used is medical grade stainless steel material that conforms to ASTM F138.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular emblem with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized design featuring three curved lines that resemble a bird in flight or a symbolic representation of human services.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 1 5 2001

Ms. Luis Nesprido Senior Associate, Regulatory Affairs United States Surgical A Division of Tyco Healthcare Group, LP 150 Glover Avenue Norwalk, Connecticut 06856

Re: K012273

Trade Name: Spiral Radius 90-DTM SST System Frade Name: "Spiral Pass.3050, 888.3060, and 888.3070 Regulatory Class: II Product Code: MNH, KWP, KWQ, and MNI Dated: July 18, 2001 Received: July 19, 2001

## Dear Ms. Nesprido:

We have reviewed your Section 510(k) notification of intent to market the device referenced we have reviewed your Section 3 rQxy nearstantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the stated in the encrosure) to device Amendments, or to devices that have been reclassified in enacificin date of the Medical Doviet Faderal Food, Drug, and Cosmetic Act (Act). You may, accordance with the provisions of the general controls provisions of the Act. The general therefore, market the device, subject to the genirements for annual registration, listing of devices, Controls provisions of the rice network and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III IT your device is classified (see associated to such additional controls. Existing major regulations (1 renarce rippival), it hay of the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good A substantially equirement, as set forth in the Quality System Regulation (QS) for Martination ing Practice requiration (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to inspections, the Food and Drug realitines and in regulatory action. In addition, FDA may publish contiply with the Grill Teganation and our device in the Federal Register. Please note: this rurner almounteenting concerning your submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

## Page 2 - Ms. Luis Nesprido

This letter will allow you to begin marketing your device as described in your 110k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed notification. The I Dri Inding of successification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CER Part 801 and If you desire specific advice for your do real not be and the Office of Compliance at additionally 607.10 for m viro diagnestions on the promotion and advertising of your device, (201) 594-4037. Traditionally, for quese at (301) 594-4639. Also, please note the regulation
 Please contact the Office of Compliance at (301) 594-4639. Also, please of Com prease connect the Office or Comphane as (301) Dotification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Information on your responsionnees and its toll of 100 miles (800) 638-2041 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

AAMephullerm for
Justin M. White, Ph.D., M.D.

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### Indications for Use V. 510(k) Number (if known):

Ko12273

Spiral Radius 90-D™ SST System Device Name:

# Indications For Use:

The components of the Spiral Radius 90-D™ SST System are indicated for spinal fixation.

When used as an anterolateral/anterior system consisting of rods and screws, the levels of attachment are the lumbar, thoracic and cervical spines. The points of attachment and methods are screw fixation to the anterolateral bodies of the lumbar and thoracic spine (T6-L5) and the anterior vertebral bodies of the cervical spine. The indications are degenerative disc disease defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion (pseudoarthrosis).

When used as a nonpedicle posterior system consisting of hooks, crosslinks and sacral/iliac screws the levels of attachments are the lumbar, thoracic and cervical spine and sacrum and ilium. Intended uses include hook and sacral screw fixation to the lumbar spine, noncervical spine and to the T1-S1 spine; and hook and sacraliliac screw fixation to the noncervical spine. The indications are degenerative disc disease defined as back pain of discogenic origin with degenerative of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumors, pseudoarthrosis, or failed previous fusion (pseudoarthrosis).

When used as a pedicle screw system, in the non-cervical spine of skeletally mature patients, the Spiral Radius 90-D™ SST System is indicated for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of solid fusion mass.

In addition, when used as a pedicle screw system, in the non-cervical spine of skeletally mature patients, the Spiral Radius 90-D™ SST System is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in the treatment of the following acute and chronic instabilities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

(Please do not write below this line - continue on another page if needed)

| Prescription Use:            | <div style="display:inline-block;">✓</div> |
|------------------------------|--------------------------------------------|
| Use:<br>(Per 21 CFR 801.109) |                                            |

Concurrence of CDRH, Office of Device Evaluation (ODE)

OR Over-The-Counter  
(Division Sign-Off)  
Division of General, Restorative  
and Neurological Devices

| 510(k) Number | k012273 |
|---------------|---------|
|---------------|---------|

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MNH/K012273](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MNH/K012273)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
