K233758 · Ortho Development Corp. · MEH · Mar 8, 2024 · Orthopedic
Device Facts
Record ID
K233758
Device Name
Trivicta Hip Stem
Applicant
Ortho Development Corp.
Product Code
MEH · Orthopedic
Decision Date
Mar 8, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is intended for use in total and hemi-hip arthroplasty. The device is intended for uncemented, press-fit use only in cases of: 1. Notably impaired hip joint due to osteoarthritis, rheumatoid arthritis, and/or post traumatic arthritis. 2. Previously failed hip surgery. 3. Proximal femoral neck fractures or dislocation. 4. Idiopathic avascular necrosis of the femoral head. 5. Non-union of proximal femoral neck fractures. 6. Treatment of fractures that are unmanageable using other forms of therapy. 7. Benign or malignant bone tumors, congenital dysplasia or other structural abnormalities where sufficient bone stock exists to properly seat the prosthesis.
Device Story
Trivicta Hip Stem is a single-piece, tapered femoral hip stem for uncemented, press-fit fixation in the medullary canal. Device features a 12/14 trunnion taper for modular attachment to femoral heads; available in various lengths (101-119mm), offsets, and collar/collarless configurations. Manufactured from titanium alloy with hydroxyapatite (HA) and sintered bead porous coating to promote osseointegration. Used by orthopedic surgeons in clinical settings for total or hemi-hip arthroplasty. Device provides structural support and joint articulation when paired with compatible femoral heads and acetabular components. Benefits include restoration of hip joint function and stability in patients with degenerative or traumatic hip conditions.
Clinical Evidence
No clinical testing was performed. Safety and effectiveness were established through non-clinical bench testing, including range of motion (ISO 21535:2023), shear fatigue (ASTM F1160-14), neck fatigue (ISO 7206-6:2013), distal stem fatigue (ISO 7206-4:2010), and engineering analysis of impingement (ASTM F2582-20). Biocompatibility was confirmed per ISO 10993-1.
Technological Characteristics
Titanium alloy femoral stem; hydroxyapatite (HA) and sintered bead porous coating. 12/14 trunnion taper. Lengths 101-119mm; neck angle 132°. Sterilized via gamma radiation (ISO 11137). Packaging validated per ISO 11607. Mechanical standards: ISO 21535, ASTM F1160, ISO 7206-6, ISO 7206-4, ASTM F2582.
Indications for Use
Indicated for patients requiring total or hemi-hip arthroplasty due to osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, failed prior hip surgery, proximal femoral neck fractures/dislocation, idiopathic avascular necrosis, non-union of femoral neck fractures, or benign/malignant bone tumors/dysplasia where sufficient bone stock exists.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Predicate Devices
DepuySynthes Actis DuoFix Hip Prosthesis (K150862)
Reference Devices
DepuySynthes Actis DuoFix Hip Prosthesis (K210581)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 8, 2024
Ortho Development Corp. Drew Weaver Director of Regulatory Affairs 12187 S. Business Park Drive Draper, Utah 84020
Re: K233758
Trade/Device Name: Trivicta™ Hip Stem Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: MEH, LPH, LZO, KWL Dated: February 12, 2024 Received: February 12, 2024
Dear Drew Weaver:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Limin Sun -S
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
K233758
Device Name
Trivicta™ Hip Stem
#### Indications for Use (Describe)
This device is intended for use in total and hemi-hip arthroplasty. The device is intended for uncemented, press-fit use only in cases of:
- 1. Notably impaired hip joint due to osteoarthritis, rheumatoid arthritis, and/or post traumatic arthritis.
- 2. Previously failed hip surgery.
- 3. Proximal femoral neck fractures or dislocation.
- 4. Idiopathic avascular necrosis of the femoral head.
- 5. Non-union of proximal femoral neck fractures.
6. Treatment of fractures that are unmanageable using other forms of therapy.
7. Benign or malignant bone tumors, congenital dysplasia or other structural abnormalities where sufficient bone stock exists to properly seat the prosthesis.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:
| Name of the Sponsor: | Ortho Development® Corporation<br>12187 South Business Park Drive<br>Draper, Utah 84020 | | |
|-------------------------------------------------|-----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|--|
| 510(k) Contact: | Name: | Drew Weaver | |
| | Position: | Director of Regulatory Affairs | |
| | Address: | 12187 S. Business Park Drive<br>Draper, UT 84020 USA | |
| | Telephone: | (801) 553-9991 | |
| | Email: | dweaver@orthodevelopment.com | |
| Date Prepared: | March 7, 2024 | | |
| Submission Type: | Traditional | | |
| Proprietary Name: | Trivicta™ Hip Stem | | |
| Common Name: | Uncemented Hip Prosthesis | | |
| Product Code /<br>Classification: | MEH | 21 CFR 888.3353: Hip joint metal/ceramic/polymer<br>semi-constrained cemented or nonporous uncemented<br>prosthesis | |
| Associated Product<br>Code(s) / Classification: | LPH | 21 CFR 888.3358: Hip joint metal/polymer/metal<br>semi-constrained porous-coated uncemented<br>prosthesis | |
| | LZO | 21 CFR 888.3353: Hip joint metal/ceramic/polymer<br>semi-constrained cemented or nonporous uncemented<br>prosthesis | |
| | KWL | 21 CFR 888.3360: Hip joint femoral (hemi-hip)<br>metallic cemented or uncemented prosthesis | |
| Device Class: | Class II | | |
| Primary Predicate<br>Device: | DepuySynthes Actis DuoFix Hip Prosthesis (K150862) | | |
| Reference Predicate<br>Device: | DepuySynthes Actis DuoFix Hip Prosthesis (K210581)<br>Entrada™ Hip Stem (K171249) | | |
{5}------------------------------------------------
## 1.0 Device Description:
Trivictal is a single-piece, tapered, collared and non-collared, hydroxyapatite (HA) and sintered bead commercially pure (Cp) Titanium coated femoral hip stem designed for single use. The stem has a neck with a 12/14 trunnion taper for modular attachment to femoral heads.
Trivicta is manufactured from titanium alloy and device fixation is achieved through uncemented press-fit in the medullary canal and through the use of biocompatible HA coating and porous sintered bead coating.
The stem has a variety of sizes to accommodate most patients encountered: lengths (101-119mm), horizontal offsets (38-53mm), vertical offsets (29-36mm), resection angle of 41°, and neck angle of 132°. The stem is offered with both standard (STD) and extended (EXT) offsets and with and without a collar.
Trivicta™ is compatible with the following Ortho Development devices: CoCr Femoral Heads, Biolox Delta Ceramic Femoral Heads, Solitude™ Unipolar Head, Escalade Acetabular Cup System, Legend® Acetabular Liner, Escalade Legend® Acetabular Shell, and Tri-plus™ DCM Liner.
## 2.0 Indication for Use:
This device is intended for use in total and hemi-hip arthroplasty. The device is intended for uncemented, press-fit use only in cases of:
- 1. Notably impaired hip joint due to osteoarthritis, rheumatoid arthritis, and/or posttraumatic arthritis.
- Previously failed hip surgery. 2.
- 3. Proximal femoral neck fractures or dislocation.
- 4. Idiopathic avascular necrosis of the femoral head.
- 5. Non-union of proximal femoral neck fractures.
- Treatment of fractures that are unmanageable using other forms of therapy. 6.
- 7. Benign or malignant bone tumors. congenital dysplasia or other structural abnormalities where sufficient bone stock exists to properly seat the prosthesis.
## 3.0 Comparison of Technological Characteristic:
Trivicta is the same as the previously cleared predicate device DepuySynthes Actis (K150862, K210581) in terms of materials, fixation, taper/trunnion, offsets, collar/collarless, and sterilization, and is similar to Actis in terms of indications for use, stem sizes and lengths, neck angle, and packaging.
Trivicta is technologically the same as the already cleared predicate device Entrada (K171249) in terms of indications for use, taper/trunnion, neck angle, offsets, collar/collarless, sterilization method, and shelf life. Trivicta is similar to Entrada in terms of packaging.
{6}------------------------------------------------
## 4.0 Performance Data:
#### Sterilization
Trivicta is gamma radiation sterilized and was validated to a sterility assurance level of 10th in accordance with the ISO 11137.
#### Shelf Life
The packaging for Trivicta was validated in accordance with ISO 11607.
#### Biocompatibility
Trivicta's biocompatibility was established according to the requirements of ISO 10993-1 and found to be safe for its intended use.
#### Mechanical Testing
The following non-clinical mechanical tests and analyses were conducted on the subject device.
- Range of Motion Test (ISO 21535:2023) ●
- Shear Fatigue Test (ASTM F1160-14) ●
- Neck Fatigue Test (ISO 7206-6:2013) ●
- Distal Stem Fatigue Test (ISO 7206-4:2010)
- Engineering analysis of impingement performance (ASTM F2582-20) ●
## Clinical Testing
No clinical testing is required to establish the safety and effectiveness of Trivicta.
## 5.0 Substantial Equivalence Conclusion:
Verification and validation activities were conducted to establish the performance of Trivicta. The results of verification and validation activities demonstrate that Trivicta performs as well as the legally marketed predicates.
Based on similarities in indication for use/intended use, technological characteristic, basic design, device material, and principle of operation, Trivicta is considered substantially equivalent to the previously cleared predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.