← Product Code [MEH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MEH) · K220495

# SignaSure Dual Mobility System (K220495)

_Signature Orthopaedics Pty, Ltd. · MEH · Jan 13, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MEH/K220495

## Device Facts

- **Applicant:** Signature Orthopaedics Pty, Ltd.
- **Product Code:** [MEH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MEH.md)
- **Decision Date:** Jan 13, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3353
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Components of the Signature Orthopaedics hip replacement range are intended to replace a hip joint where bone stock is sufficient to support the implant. When a surgeon has selected prosthetic replacement as the preferred treatment, the devices are indicated for: - Non-inflammatory degenerative joint disease including osteoarthritis or avascular necrosis - Inflammatory joint disease including rheumatoid arthritis - Correction of functional deformity including congenital hip dysplasia - Traumatic injury involving the hip joint including traumatic arthritis or femoral head or neck fracture - Failed previous hip surgery including internal fixation or joint fusion, reconstruction, hemiarthroplasty, surface replacement, or total replacement - Dislocation risks (when used with SignaSure range) Signature Orthopaedics' Origin, Aria, Remedy, Origin-NS, Pegasus, Spartan, World and Everglade Hip femoral stems, SignaSure Cementless Cups, Logical and World Acetabular Cups are intended for cementless fixation only. Signature Orthopaedics' Evolve, Cemented TSI (both CoCr and HNSS variants), and Cemented Origin femoral stems and SignaSure Cemented Cups are intended for cemented fixation only. Signature Orthopaedics' SignaSure Logical/World Metal Insert is indicated for use with a cementless Signature Orthopaedics' Logical/World Acetabular Cup to provide dual mobility articulation. Signature Orthopaedics' constrained liner components are indicated particularly for patients at hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability. Signature Orthopaedics' Evolve UniPolar Head and BiPolar Head are intended for hemi-hip arthroplasty only, where the natural acetabulum does not require replacement. The Evolve UniPolar Head are indicated for bone fractures or pathologies involving only the femoral head/neck and/or proximal femur, such as: - Acute femoral head or neck fracture - Fracture dislocation of the hip - Avascular necrosis of the femoral head - Non-union of femoral neck fractures - Certain high subcapital and femoral neck fractures in the elderly - Degenerative arthritis involving only the femoral head

## Device Story

SignaSure Dual Mobility System provides hip joint replacement via Monobloc or Modular configurations. Modular system consists of mobile polyethylene insert and metallic insert for use with cemented or cementless acetabular cups. Polyethylene insert articulates within metallic insert/cup to provide dual mobility. Used by orthopedic surgeons in clinical settings for total hip arthroplasty or hemi-arthroplasty. Device restores joint function, improves stability, and reduces dislocation risk in patients with degenerative or traumatic hip conditions. Surgeon selects appropriate components based on patient bone stock and clinical history.

## Clinical Evidence

No clinical data. Evidence consists of bench testing and engineering evaluations including liner-shell offset pull-out strength, rotational stability, push-out strength, fretting corrosion, impingement, range of motion, and head-poly assembly/lever-out testing per ASTM F1820, ASTM F1875, ASTM F2582-14, and ISO 21535:2007.

## Technological Characteristics

Modular dual mobility hip prosthesis. Materials: Highly crosslinked polyethylene (ASTM F648), Cobalt Chrome alloy (ASTM F1537, ISO 5832-12). Cementless cups feature plasma-sprayed titanium (ASTM F1580) and hydroxyapatite (ISO 13779-1/2). Cemented cups are grit-blasted. Mechanical fixation via cement or press-fit. No software or electronic components.

## Regulatory Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

## Predicate Devices

- Stryker Modular Dual Mobility (MDM) Insert and Polyethylene ([K103233](/device/K103233.md), [K112556](/device/K112556.md))
- Serf Novae Dual Mobility Acetabular Cup ([K111572](/device/K111572.md))
- Serf CI../..X liner for Novae Dual Mobility Acetabular Cup ([K142675](/device/K142675.md))
- Stryker Restoration Anatomic Dual Mobility (ADM) System ([K072020](/device/K072020.md), [K093644](/device/K093644.md))
- Signature Orthopaedics SignaSure Dual Mobility System ([K211742](/device/K211742.md))

## Reference Devices

- Signature Orthopaedics Logical C-Series Cup ([K153131](/device/K153131.md))
- Signature Orthopaedics Logical XLPE Liner ([K121297](/device/K121297.md))
- Signature Orthopaedics CoCr Femoral Head ([K121297](/device/K121297.md))
- Biomet G7 Dual Mobility System ([K150522](/device/K150522.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.

January 13, 2023

Signature Orthopaedics Pty Ltd. Declan Brazil Managing Director 7 Sirius Road Lane Cove, NSW 2066 Australia

Re: K220495

Trade/Device Name: SignaSure Dual Mobility System Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: MEH, LZO Dated: December 8, 2022 Received: December 14, 2022

Dear Declan Brazil:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Limin

Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K220495

Device Name SignaSure Dual Mobility System

#### Indications for Use (Describe)

Components of the Signature Orthopaedics hip replacement range are intended to replace a hip joint where bone stock is sufficient to support the implant. When a surgeon has selected prosthetic replacement as the preferred treatment, the devices are indicated for:

- · Non-inflammatory degenerative joint disease including osteoarthritis or avascular necrosis
- · Inflammatory joint disease including rheumatoid arthritis
- Correction of functional deformity including congenital hip dysplasia
- · Traumatic injury involving the hip joint including traumatic arthritis or femoral head or neck fracture

· Failed previous hip surgery including internal fixation or joint fusion, reconstruction, hemiarthroplasty, surface replacement, or total replacement

- · Dislocation risks (when used with SignaSure range)
Signature Orthopaedics' Origin. Aria, Remedy, Origin-NS, Pegasus, Spartan, World and Everglade Hip femoral stems, SignaSure Cementless Cups, Logical and World Acetabular Cups are intended for cementless fixation only.

Signature Orthopaedics' Evolve, Cemented TSI (both CoCr and HNSS variants), and Cemented Origin femoral stems and SignaSure Cemented Cups are intended for cemented fixation only.

Signature Orthopaedics' SignaSure Logical/World Metal Insert is indicated for use with a cementless Signature Orthopaedics' Logical/World Acetabular Cup to provide dual mobility articulation.

Signature Orthopaedics' constrained liner components are indicated particularly for patients at hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability.

Signature Orthopaedics' Evolve UniPolar Head and BiPolar Head are intended for hemi-hip arthroplasty only, where the natural acetabulum does not require replacement. The Evolve UniPolar Head are indicated for bone fractures or pathologies involving only the femoral head/neck and/or proximal femur, such as:

- Acute femoral head or neck fracture
- Fracture dislocation of the hip
- Avascular necrosis of the femoral head
- Non-union of femoral neck fractures
- · Certain high subcapital and femoral neck fractures in the elderly
- · Degenerative arthritis involving only the femoral head

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

## 510(K) SUMMARY - SIGNASURE DUAL MOBILITY SYSTEM

| Main<br>Manufacturer:       | Signature Orthopaedics Pty Ltd<br>7 Sirius Road<br>Lane Cove, NSW 2066<br>Australia                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Additional<br>Manufacturer: | Signature Orthopaedics Europe Ltd<br>Unit A, IDA Business & Technology Park Garrycastle<br>Athlone Westmeath N37 DY26<br>Ireland                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Contact:                    | Dr. Declan Brazil<br>CEO of Signature Orthopaedics                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Prepared By:                | Signature Orthopaedics Pty Ltd<br>7 Sirius Road<br>Lane Cove, NSW 2066<br>Australia<br>Phone: +61 (2) 9428 5181<br>Fax: +61 (2) 8456 6065                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Date Prepared:              | April 20th, 2022                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Device Trade<br>Name:       | SignaSure Dual Mobility System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Common Name:                | Hip Replacement Prosthesis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Classification:             | Class II per 21 CFR 888.3353: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis (MEH)<br>Class II per 21 CFR 888.3353: Prosthesis, Hip, Semi-Constrained,<br>Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented (LZO)                                                                                                                                                                                                                                                                                                                                                                                               |
| Predicate Devices:          | Primary Predicate<br>• Stryker Modular Dual Mobility (MDM) Insert and Polyethylene (K103233, K112556)<br>Additional Predicates<br>• Serf Novae Dual Mobility Acetabular Cup (K111572)<br>• Serf CI../..X liner for Novae Dual Mobility Acetabular Cup (K142675)<br>• Stryker Restoration Anatomic Dual Mobility (ADM) System (K072020, K093644)<br>• Signature Orthopaedics SignaSure Dual Mobility System (K211742)<br>Reference Predicate<br>• Signature Orthopaedics Logical C-Series Cup (K153131)<br>• Signature Orthopaedics Logical XLPE Liner (K121297)<br>• Signature Orthopaedics CoCr Femoral Head (K121297)<br>• Biomet G7 Dual Mobility System (K150522) |

{5}------------------------------------------------

#### Device Description:

The SignaSure Dual Mobility System is available in two configurations, a Monobloc configuration and a Modular configuration. The Monobloc configuration is a Dual Mobility acetabular system consisting of a metallic shell, cemented, and a mobile polyethylene insert to articulate within the shell as a femoral head. The Modular configuration (subject device) is a dual mobility acetabular system consisting of 2 components; a mobile polyethylene component and a metallic insert which can be used with interfacing cemented or cementless cups.

The SignaSure Poly, used in both configurations, is manufactured from highly crosslinked polyethylene (as per ASTM F648) and is spherical in geometry with outer and inner spherical conforming articular surfaces. The inner articular surface mates and retains a femoral head. The outer articular surface articulates within the highly polished inner diameter of the SignaSure Cementless or Cemented Cup to be used, or the metallic insert. The SignaSure Poly components are available in size 37 to 41 mm (outer diameter) allowing use with 22 mm and 39 to 59 mm femoral heads.

The SignaSure Logical Insert is manufactured from Cobalt Chrome (CoCr) alloy (as per ASTM F1537 and ISO 5832-12) and mates with Signature Orthopaedics' Logical Shell to be implanted without bone cement and the SignaSure World Insert is manufactured from CoCr alloy (as per ASTM F1537 and ISO 5832-12) and mates with Signature Orthopaedic's World Cup to be implant without bone cement. The SignaSure Logical Insert is available in 34 to 48 mm sizes whereas the SignaSure World Insert is available in 38 to 46 mm (internal diameter) sizes.

The SignaSure Cementless Cup is manufactured from CoCr alloy (as per ASTM F1537 and ISO 5832-12) and is sequentially plasma sprayed with titanium coating (as per ASTM F1580) and hydroxyapatite (as per ISO 13779-1 and ISO 13779-2) to gain cementless fixation. The SignaSure Cemented Cup is manufactured from CoCr alloy (as per ASTM F1537 and ISO 5832-12) and is grit blasted and macro textured to aid in fixation via bone cement. These cups, as from K211742, underwent no changes and were included in this submission as they interact with the subject SignaSure Poly devices.

#### Indications for Use:

Components of the Signature Orthopaedics hip replacement range are intended to replace a hip joint where bone stock is sufficient to support the implant. When a surgeon has selected prosthetic replacement as the preferred treatment, the devices are indicated for:

- Non-inflammatory degenerative joint disease including osteoarthritis or avascular necrosis
- Inflammatory joint disease including rheumatoid arthritis
- Correction of functional deformity including congenital hip dysplasia ●
- Traumatic injury involving the hip joint including traumatic arthritis or femoral head or ● neck fracture
- . Failed previous hip surgery including internal fixation or joint fusion, reconstruction, hemiarthroplasty, surface replacement, or total replacement
- . Dislocation risks (when used with SignaSure range)

Signature Orthopaedics' Origin, Aria, Remedy, Origin-NS, Pegasus, Spartan, World and Everglade Hip femoral stems, SignaSure Cementless Cups, Logical and World Acetabular Cups are intended for cementless fixation only.

{6}------------------------------------------------

Signature Orthopaedics' Evolve. Cemented TSI (both CoCr and HNSS variants), and Cemented Origin femoral stems and SignaSure Cemented Cups are intended for cemented fixation only.

Signature Orthopaedics' SignaSure Logical/World Metal Insert is indicated for use with a cementless Signature Orthopaedics' Logical/World Acetabular Cup to provide dual mobility articulation.

Signature Orthopaedics' constrained liner components are indicated particularly for patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability.

Signature Orthopaedics' Evolve UniPolar Head and BiPolar Head are intended for hemi-hip arthroplasty only, where the natural acetabulum does not require replacement. The Evolve UniPolar Head and BiPolar Head are indicated for bone fractures or pathologies involving only the femoral head/neck and/or proximal femur, such as:

- Acute femoral head or neck fracture ●
- Fracture dislocation of the hip
- Avascular necrosis of the femoral head ●
- Non-union of femoral neck fractures ●
- Certain high subcapital and femoral neck fractures in the elderly .
- Degenerative arthritis involving only the femoral head .

#### Performance Testing:

Non-clinical testing and engineering evaluations were conducted to verify that the performance of the SignaSure Dual Mobility System is adequate for anticipated in-vivo use. The following non-clinical tests were carried out:

- . Liner-Shell Offset Pull-Out Strength per ASTM F1820
- Liner-Shell Rotational Stability (torque-out) per ASTM F1820 ●
- Liner-Shell Push-Out Strength per ASTM F1820
- Fretting Corrosion as per ASTM F1875 and Engineering ● Evaluation
- Impingement per ASTM F2582-14
- Range of Motion Testing per ISO 21535:2007 ●
- Head Poly Assembly Testing
- Head Poly Pull-Out Testing
- . Head Lever Out Testing
- Post Impingement Lever-Out Testing per ASTM F1820 ●

#### Substantial Equivalence:

The Signature Orthopaedics SignaSure Dual Mobility System has the same intended use, indications for use, materials and similar design features to the predicate devices. Nonclinical testing results support the substantial equivalence claim.

#### Comparison of technological characteristics

The subject devices share the same characteristics and the predicate devices as follows:

SignaSure Dual Mobility System (Modular Insert configuration):

- . The SignaSure Modular Insert Dual Mobility System shares the same intended use and indications for use as Stryker's MDM Insert (K103233, K112556).
- The SignaSure Poly material is the same as Stryker's MDM Insert (K103233, K112556).
- The SignaSure Insert material is the same as Stryker's MDM Insert (K103233, ● K112556).

{7}------------------------------------------------

#### SignaSure Dual Mobility System (Monobloc configuration):

The materials are the same as the Stryker ADM Cup (K072020, K093644), and Logical XLPE Liner (K121297). The SignaSure Cup's geometry is similar to the Serf Novae Dual Mobility Acetabular Cup (K111572). The SignaSure Cementless Cup's coating is identical to the Logical C-Series Cup (K121297). The SignaSure Cemented Cup's outer surface is similar to the Serf Novae Stick Dual Mobility Acetabular Cup (K111572).

The SignaSure Dual Mobility System's Monobloc Configuration is unchanged when compared to the Monobloc Configuration cleared in K211742.

#### Conclusions:

Technical comparison, engineering evaluation, and review of the documentation provided for the subject and predicate devices demonstrates substantial equivalence in device design, intended use, indications for use and material. Non-clinical data supports the safety and effectiveness of the Signature Orthopaedics SignaSure Dual Mobility System.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MEH/K220495](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MEH/K220495)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
