K103742 · Integra Life Sciences · MBP · Mar 17, 2011 · Orthopedic
Device Facts
Record ID
K103742
Device Name
ACCELL EVO3 (FORMERLY ACCELL A2I)
Applicant
Integra Life Sciences
Product Code
MBP · Orthopedic
Decision Date
Mar 17, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Accell Evo3 is intended to fill voids and gaps in the skeletal system that are not intrinsic to the bony structure. The product is indicated for use as a bone graft extender in the spine, extremities and pelvis. Accell Evo3 may also be used as a bone void filler in the posterolateral spine, extremittes and pelvis. The voids or gaps may be surgically created defects or the result of traumatic injury to the bone.
Device Story
Accell Evo3 is a moldable, osteoconductive bone void filler putty; composed of ground, cortical de-mineralized bone matrix (DBM) in particulate and solubilized forms; includes poloxamer reverse phase medium for handling. Packaged in prefilled open-bore polycarbonate syringe; terminally sterilized via e-beam radiation. Used by surgeons in clinical settings to fill skeletal voids/gaps not intrinsic to bony stability. Device resorbs over time, providing a scaffold for bone growth and remodeling. Osteoinductive potential assayed in vitro and validated via in vivo athymic mouse model. Clinical benefit derived from supporting bone fusion and repair in orthopedic procedures.
Clinical Evidence
No human clinical data provided. Evidence based on in vitro assays for bone morphogenetic protein-2 and in vivo testing in a postero-lateral rabbit spine model. Testing demonstrated effective bone formation and fusion, with radiographic, histological, and biomechanical results comparable to autograft controls.
Technological Characteristics
Moldable putty containing cortical de-mineralized bone matrix (DBM) and poloxamer reverse phase medium. Packaged in prefilled open-bore polycarbonate syringe. Sterilized via e-beam radiation. Osteoconductive material properties.
Indications for Use
Indicated for patients requiring bone graft extension or void filling in the spine, extremities, and pelvis due to surgically created defects or traumatic bone injury.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
# 510(k) SUMMARY
#### Date of Preparation December 20, 2010
Company Integra LifeSciences 2 Goodyear Irvine, CA 92618
#### Contact
Jenny Fam Director of Regulatory Affairs Integra Orthobiologics 949-595-8710
Proprietary Name Accell Evo3
# Common Name
Bone Void Filler
## Classification
21 CFR 888.3045; Resorbable calcium salt bone void filler
## Product Codes
MQV: Filler, Bone Void, Calcium Compound MBP: Filler, Bone Void, Osteoinduction (Without Human Growth Factor)
## Predicate Devices
The device is substantially equivalent to Integra Accell DBM Family (K061880) and Ostectech Grafton DBM Putty (K051195).
#### Indications for Use
Accell Evo3 is intended to fill voids and gaps in the skeletal system that are not intrinsic to the bony structure. The product is indicated for use as a bone graft extender in the spine, extremities and pelvis. Accell Evo3 may also be used as a bone void filler in the posterolateral spine, extremittes and pelvis. The voids or gaps may be surgically created defects or the result of traumatic injury to the bone.
#### Device Description
Accell Evo3 is a moldable putty that contains ground, cortical de-mineralized bone marrix (DBM) in particulate and solubilized forms as well as a poloxamer reverse phase medium for proper handling. The device is packaged in a prefiled open-bore polycarbonate syringe and terminally sterilized by e-beam radiation. The DBM used to manufacture Accell Evo3 is only obtained from AATB-accredited facilities.
#### Device Performance
Accell Evo3 is an osteoconductive bone void filler. The device is resorbed over time and provides a favorable environment that stimulates bone growth and remodeling. Accell Evo3 has been assayed in vitro for bone morphogenetic protein-2. The in vitro assay has been validated to an in vivo (athymic mouse) osteoinductive assay. It is unknown how osteoinductive polential measured via the in vivo assays will correlate with human clinical performance.
#### Viral Inactivation
The methods for processing the DBM contained in Accell Evo3 have been evaluated for viral inactivation potential. A panel of model potential human viruses representing various wirus types and genomes were evaluated. The DBM processing methods have been determined to provide significant viral inactivation for a wide range of potential viruses.
## Determination of Substantial Equivalence
Accell Evo3 demonstrated effective bone formation through fusion when tested in the postero-lateral rabbit spine model. Radiographic, histological and biomechanical evidence of the device's equivalence to autograft The in vivo and in vitro tests performed on Accell Evo3 demonstrate substantial equivalence in terms of controls. intended use, technological characteristics and performance to its predicate devices.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three wing-like strokes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Integra LifeSciences IsoTis Orthobiologics % Ms. Jenny Fam Director of Regulatory Affairs 2 Goodyear Irvine, California 92672
MAR 1 7 2011
## Re: K103742
Trade/Device Name: Accell Evo36 Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MBP, MQV Dated: December 20, 2010 Received: December 28, 2010
# Dear Ms. Fam:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Ms. Jenny Fam
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/industry/default.htm.
Sincerely vours.
Az. B. M. h
Mark N. Melkerson Director Division of Surgical. Orthonedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## INDICATIONS FOR USE
## 510(k) Number:
Device Name: Accell Evo38
#### Indications for Use:
Accell Evo3 is intended for filling voids and gaps in the skeletal system that are not intrinsic to the stability of the bony structure. The product is indicated for use as a bone graft extender in the spine, extremities and pelvis. Accell Evo3 may also be used as a bone void filler in the posterolateral spine, extremities and pelvis. The voids or gaps may be surgically created defects or the result of traumatic injury to the bone.
Prescription Use X (Part 21 CFR 801 Sub-part D) AND/OR
Over-The-Counter Use (Part 21 CFR 807 Sub-part C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sign Off
(Division Sign-Off) (Division Sign-Off)
Division of Surgical, Orthopedic.
Division torotive Devices Division of Surgical
and Restorative Devices
Page 1 of 1
510(k) Number K103742
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.