K060161 · Musculoskeletal Transplant Foundation · MBP · Mar 21, 2006 · Orthopedic
Device Facts
Record ID
K060161
Device Name
AFT ALLOGRAFT FILLER TUBE
Applicant
Musculoskeletal Transplant Foundation
Product Code
MBP · Orthopedic
Decision Date
Mar 21, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Intended Use
AFT is intended for use as a bone void filler in the extremities, spine and pelvis for voids or gaps that are not intrinsic to the stability of the bony structure. AFT is indicated for use in the treatment of osseous defects caused by surgery or traumatic injury. AFT is intended for single patient use only.
Device Story
AFT Allograft Filler Tube is a bone void filler composed of human demineralized bone matrix, human non-demineralized bone, and sodium hyaluronate. Provided pre-loaded in a disposable delivery tube for single-patient use. Used by surgeons to fill osseous voids or gaps in extremities, spine, and pelvis not intrinsic to bony stability. Device is osteoconductive and possesses osteoinductive potential. Aseptically processed to ensure viral inactivation. Supports new bone growth in osseous defects.
Clinical Evidence
No clinical data. Evidence consists of bench testing and in vivo testing in an athymic mouse model to demonstrate osteoinductive potential and support for new bone growth. Biocompatibility established via ISO 10993 testing and historical clinical use. Viral inactivation validated via model virus panel.
Technological Characteristics
Composition: human demineralized bone matrix, human non-demineralized bone, sodium hyaluronate. Form factor: pre-loaded disposable delivery tube. Resorbable. Aseptically processed. Sterility tested per USP <71>. Biocompatibility per ISO 10993.
Indications for Use
Indicated for patients requiring bone void filling in extremities, spine, and pelvis for non-structural osseous defects resulting from surgery or trauma.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
K113167 — MTF NEW BONE VOILD FILLER · Musculoskeletal Transplant Foundation · Jan 24, 2012
K121313 — MTF NEW BONE VOID FILLER · Musculoskeletal Transplant Foundation · Jan 11, 2013
K080399 — DBX DEMINERALIZED BONE MATRIX PUTTY, DBX DEMINERALIZED BONE MATRIX PASTE · Musculoskeletal Transplant Foundation · Oct 10, 2008
K110003 — MTF NEW BONE VOID FILLER · Musculoskeletal Transplant Foundation · Oct 13, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
AFT 510(k) Premarket Notification
MAR 2 1 2006
K 06016/
# X. 510(k) SUMMARY OF SAFTEY & EFFECTIVENESS
| PROPREIETARY NAME: | AFT Allograft Filler Tube |
|-----------------------------------|---------------------------------------------------------------------------------------------|
| COMMON NAME: | Bone Void Filler Containing Human Demineralized<br>Bone Matrix (DBM) |
| PROPOSED REGULATORY<br>CLASS: | Class II |
| CLASSIFICATION<br>IDENTIFICATION: | 21 C.F.R. §888.3045 Resorbable calcium salt bone<br>void filler device |
| PRODUCT CODE: | 87—Orthopedic Devices |
| SPONSOR: | Musculoskeletal Transplant Foundation<br>125 May Street<br>Edison, NJ 08837<br>732-661-0202 |
#### INDICATIONS FOR USE:
AFT is intended for use as a bone void filler in the extremities, spine and pelvis for voids or gaps that are not intrinsic to the stability of the bony structure. AFT is indicated for use in the treatment of osseous defects caused by surgery or traumatic injury.
#### DEVICE DESCRIPTION:
AFT is composed of human demineralized bone matrix, human non-demineralized bone and sodium hyaluronate. All components of AFT are resorbable. AFT is aseptically processed and provided pre-loaded into a disposable delivery tube.
#### SUBSTANTIAL EQUIVALENCE INFORMATION:
AFT shares the same function and intended use and therefore is substantially equivalent to OSTEOSET® and Exactech Resorbable Bone Paste. In vivo testing in the athymic mouse model has demonstrated that AFT materials can effectively support new bone growth in osseous defects.
Confidential
{1}------------------------------------------------
## SAFETY AND EFFECTIVENESS INFORMATION:
Biocompatibility of AFT materials has been established through their long history of safe and effective clinical use, further supported by laboratory testing conducted per ISO 10993. AFT is single-donor processed using aseptic techniques and is tested for sterility per current USP <71>.
### OSTEOINDUCTIVITY POTENTIAL:
AFT is osteoconductive, and has been shown to have osteoinductivity potential in an athymic mouse. Every lot of final product will be tested to ensure the osteoinductive potential of the final product. Osteoinduction assay results in the athymic mouse model should not be interpreted to predict clinical performance in human subjects.
## VIRAL CLEARANCE AND INACTIVATION:
The method for processing the DBM and CBM (cortical cancellous) contained in the AFT was evaluated for its viral inactivation potential. A panel of model potential human viruses representing various virus types, sizes, shapes and genomes were evaluated. The DBM processing methods were determined to provide significant viral inactivation potential for a wide range of potential viruses. The CBM processing methods were determined to provide some viral inactivation potential for a wide range of viruses. In comparison, the CBM processing methods provided less viral inactivation potential than the DBM processing methods; therefore, the risk of disease transmission for the CBM component is greater than the DBM component. However, the risk for disease transmission for these components remains low due to the multiple safeguards employed, i.e., donor selection, laboratory testing, and material processing.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo features a stylized eagle with three stripes forming its wing and body. The eagle is encircled by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 1 2006
Musculoskeletal Transplant Foundation c/o Ms. Nancy Bennewitz Regulatory Affairs Submission Specialist 125 May Street Edison, New Jersey 08837
Re: K060161
Trade/Device Name: AFT Allograft Filler Tube Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: II Product Code: MBP Dated: December 6, 2005 Received: January 20, 2006
Dear Ms. Bennewitz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for abovened as to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (sec above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 – Ms. Nancy Bennewitz
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Seeting your device as described in your Section 510(k) This letter will anow you to begin manies.ing of substantial equivalence of your device to a legally premits that hourication. The PDF miamig of cation for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your ac not our and one one one the regulation entitled, and comact the Office of Compliance at (210) = 6 = 8 = 8 = 8 = 8 = 8 = 8 = 8 = 8 = 8 = 0 = 3 = 0 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = other general information on your responsibilities under the Act from the Division of Small other gelleral International on your responsibilities and its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Herbert Lemmerling
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# IV. INDICATIONS FOR USE
510(k) Number (if known):
Device Name: AFT Allograft Filler Tube
Indications for Use:
AFT is intended for use as a bone void filler in the extremities, spine and pelvis for voids or gaps that are not intrinsic to the stability of the bony structure. AFT is indicated for use in the treatment of osseous defects caused by surgery or traumatic injury.
AFT is intended for single patient use only.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ OR (Pcr 21 CFR 801 Subpart D)
Over-The-Counter Use __ (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE—CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Helent Lemmerts
Division of General, Restorative, and Neurological Devices
**510(k) Number** K060161
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.