← Product Code [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI) · K965111

# STABILIZER SOFT TISSUE ANCHOR MODEL NUMBER TA1005, TA1008, ANCHOR INSERTER/SUTURE ORGANIZER MODEL NUMBERS AI1005, AI1008 (K965111)

_R. Thomas Grotz, M.D., Inc. · MBI · Feb 18, 1997 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K965111

## Device Facts

- **Applicant:** R. Thomas Grotz, M.D., Inc.
- **Product Code:** [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI.md)
- **Decision Date:** Feb 18, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Stabilizer Soft Tissue Anchor is intended for use with USP Sutures as an attachment means for soft tissue and bone for the indications listed below: Shoulder 1. Bankart repair 2. SLAP lesion repair 3. Acromioclavicular separation 4. Rotator cuff repair 5. Capsule shift\capsulolabral reconstruction 6. Biceps tenodesis 7. Deltoid repair Elbow 1. Tennis elbow repair 2. Biceps tendon reattachment 3. Medial and lateral repairs Ankle 1. Lateral instability 2. Medial instability 3. Achilles tendon repair/ reconstruction Knee 1. Extra-capsular repairs and attachment of: - medial collateral ligament - lateral collateral ligament - posterior oblique ligament and joint capsule to tibia - joint capsule closure to the anterior proximal tibia 2. Extra-capsular reconstruction, ITB tenodesis 3. Patellar ligament and tendon avulsions repairs

## Device Story

Stabilizer Soft Tissue Anchor is a 316L stainless steel implant for attaching soft tissue to bone during joint reconstruction. Device consists of an anchor available in 5mm and 8mm diameters, a dedicated drill, an inserter, and a crimper. Procedure involves drilling a hole in bone, inserting the anchor, and using the inserter to expand the anchor's teeth for fixation. Three USP grade surgical sutures are used to secure soft tissue to the anchor. The inserter also functions as a suture organizer. Used by surgeons in clinical settings. Provides mechanical fixation to facilitate healing; benefits include stable attachment of tendons and ligaments to bone.

## Clinical Evidence

bench testing only

## Technological Characteristics

Material: 316L stainless steel. Design: Expandable tooth anchor. Sizes: 5mm and 8mm diameter. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Mitek Superanchor

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K965111
510(k) Premarket Notification
Stabilizer™ Soft Tissue Anchor
for Human Joints
R. Thomas Grotz, M.D.
- Confidential -
510(k) SUMMARY
FEB 18 1997

## SUBMITTED BY:

R. Thomas Grotz, M.D.
530 Bush Street, 10th Floor
San Francisco, California 94108
Telephone: (415) 398-2332
Fax: (415) 398-2614

Date Submitted: December 19, 1996

## CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME

Classification Name: Fastener, fixation, nondegradable, soft tissue
Common/Usual Name: Soft Tissue Anchor
Proprietary Name: Stabilizer™ Soft Tissue Anchor

## PREDICATE DEVICE

Mitek Superanchor™ manufactured by Mitek Surgical Products, Inc.

## DEVICE DESCRIPTION

The Stabilizer Soft Tissue Anchor is a 316L Stainless Steel implant designed to provide an attachment means for soft tissue to bone using USP grade surgical sutures in joint reconstruction procedures. It is available in two sizes: 5 and 8 mm diameter. Attachment is accomplished by drilling an appropriately sized hole in uncompromised bone with a specifically designed drill, inserting the soft tissue anchor into the bone, expanding the stabilizer teeth of the implant to secure the anchor into bone using the anchor inserter, and securing soft tissue to the implanted anchor by using three sutures. The anchor inserter (which spreads the stabilizer teeth of the implant) also serves as a suture organizer for delivery of sutures to the implant site during the implantation procedure. A crimper is also included to help secure the suture of choice to the Stabilizer and to prepare the stabilizer for entry into the predrilled hole.

## INDICATIONS FOR USE:

The Stabilizer Soft Tissue Anchor is intended for use with USP Sutures as an attachment means for soft tissue and bone for the indications listed below:

|  Shoulder | Elbow  |
| --- | --- |
|  1. Bankart repair | 1. Tennis elbow repair  |
|  2. SLAP lesion repair | 2. Biceps tendon reattachment  |
|  3. Acromioclavicular separation | 3. Medial and lateral repairs  |
|  4. Rotator cuff repair |   |

21

{1}

510(k) Premarket Notification
Stabilizer™ Soft Tissue Anchor
for Human Joints
R. Thomas Grotz, M.D.
- Confidential -

## Shoulder (continued)

5. Capsule shift\capsulolabral reconstruction
6. Biceps tenodesis
7. Deltoid repair

## Ankle

1. Lateral instability
2. Medial instability
3. Achilles tendon repair/ reconstruction

## Knee

1. Extra-capsular repairs and attachment of:
- medial collateral ligament
- lateral collateral ligament
- posterior oblique ligament and joint capsule to tibia
- joint capsule closure to the anterior proximal tibia
2. Extra-capsular reconstruction, ITB tenodesis
3. Patellar ligament and tendon avulsions repairs

## CONTRAINDICATIONS, WARNINGS, PRECAUTIONS AND COMPLICATIONS

### 1. Contraindications

The Stabilizer Soft Tissue Anchor is contraindicated in the presence of pathological conditions such as severe osteopenia, cystic degeneration, or comminution of bone which would compromise fixation. The Stabilizer should not be used in compromised bone or in the presence of pathological soft tissue conditions which would compromise fixation. It should also not be used in the presence of pathophysiological conditions such as infection, osteonecrosis, or bone disease. The product should not be used in patients with known allergies to stainless steel.

### 2. Warnings

#### 2.1

The Stabilizer Soft Tissue Anchor is intended to assist in the fixation of soft tissue to bone. Each case should be carefully analyzed to assure that the appropriate sized anchor and suture are being used and that the appropriate number of anchors are present to support the reconstruction. Excessive tension on the suture or anchor may result in suture breakage or implant pull-out from the bone. In some cases, revisions may require explant of the bone anchor.

#### 2.2

The drill is stainless steel. To assure proper bone cutting characteristics, the drill should be replaced after every 10 uses. If the drill should break during use, remnants should be removed from the surgical site prior to proceeding.

22

{2}

510(k) Premarket Notification
Stabilizer™ Soft Tissue Anchor
for Human Joints
R. Thomas Grotz, M.D.
- Confidential -

## 3. Precautions

The Stabilizer Soft Tissue Anchor is intended for use by surgeons familiar with soft tissue and bone attachment techniques. The patient must be cautioned against early weightbearing and/or premature ambulation as this could lead to loosening or failure of the implant or suture attachments. Standard postoperative practices for the treatment and rehabilitation of repaired joints must be followed.

## 4. Complications

Potential complications with the The Stabilizer Soft Tissue Anchor include, but are not limited to, the following: infection, osteomyelitis, suture breakage, implant breakage, implant pull-out, reoperation, revision or explant.

## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

The product design, material of construction, and function as a soft tissue anchor is substantially equivalent to the FDA cleared predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K965111](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K965111)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
