← Product Code [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI) · K190446

# Responsive Arthroscopy Wedge Push-In Suture Anchors (K190446)

_Responsive Arthroscopy, LLC · MBI · Mar 27, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K190446

## Device Facts

- **Applicant:** Responsive Arthroscopy, LLC
- **Product Code:** [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI.md)
- **Decision Date:** Mar 27, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Responsive Arthroscopy Wedge Push-In Suture Anchors are intended to be used for fixation of soft tissue to bone in the shoulder, foot/ankle, hip, knee, hand/wrist, and elbow in the following procedures. Shoulder: Rotator Cuff Repair, Bankart Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction. Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair, Bunionectomy, Digital Tendon Transfers. Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Vastus Medialis Obliquus Advancement, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis, Joint Capsule Closure. Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction, Carpal Ligament Reconstruction, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP Joints for all Digits, Digital Tendon Transfers. Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis Repair. Hip: Capsular Repair, Acetabular Labral Repair, Gluteus Medius Repair.

## Device Story

Knotless suture anchor system for soft tissue-to-bone fixation. Components include PEEK anchors, repair/passing sutures, inserters, drills, and guide tubes. Used in orthopedic surgical procedures (shoulder, foot/ankle, hip, knee, hand/wrist, elbow). Operated by surgeons in clinical/OR settings. Device provides mechanical fixation of soft tissue to bone; surgeon uses instrumentation to place anchor and secure tissue. Benefits include stable fixation for tendon/ligament repairs.

## Clinical Evidence

Bench testing only. Performed suture diameter, tensile, and cyclic suture locking force testing. Results demonstrate performance equivalent to predicate.

## Technological Characteristics

Anchors machined from PEEK (ASTM F2026). Repair sutures made from Dyneema ultrahigh molecular weight polyethylene. Mechanical fixation principle. Includes various inserter/instrumentation geometries.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Responsive Arthroscopy Wedge Push-In Suture Anchor ([K181076](/device/K181076.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 27, 2019

Responsive Arthroscopy LLC % Benjamin Arnold Managing Member Cor Medical Ventures LLC 215 S. Highway 101, Suite 200 Solana Beach, California 92075

Re: K190446

Trade/Device Name: Responsive Arthroscopy Wedge Push-In Suture Anchors Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: February 21, 2019 Received: February 25, 2019

Dear Mr. Arnold:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurence D. Coyne -S

For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K190446

Device Name

Responsive Arthroscopy Wedge Push-In Suture Anchors

Indications for Use (Describe)

The Responsive Arthroscopy Wedge Push-In Suture Anchors are intended to be used for fixation of soft tissue to bone in the shoulder, foot/ankle, hip, knee, hand/wrist, and elbow in the following procedures.

Shoulder: Rotator Cuff Repair, Bankart Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.

Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair, Bunionectomy, Digital Tendon Transfers.

Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Vastus Medialis Obliquus Advancement, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis, Joint Capsule Closure.

Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction, Carpal Ligament Reconstruction, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP Joints for all Digits, Digital Tendon Transfers.

Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis Repair.

Hip: Capsular Repair, Acetabular Labral Repair, Gluteus Medius Repair.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| <span style="text-decoration: underline;">Prescription Use</span> (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------------|---------------------------------------------|
|-----------------------------------------------------------------------------------------------|---------------------------------------------|

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) SUMMARY

### SUBMITTER:

Submitted By:

| Company Name:<br>Address: | Responsive Arthroscopy LLC<br>701 N. 3rd Street, Suite 208<br>Minneapolis, MN 55401 |
|---------------------------|-------------------------------------------------------------------------------------|
| Telephone:                | 858-720-1847                                                                        |
| CONTACT PERSON:           | Benjamin Arnold                                                                     |
| DATE PREPARED:            | February 21, 2019                                                                   |
| TRADE NAME:               | Responsive Arthroscopy Wedge Push-In Suture Anchor System                           |
| COMMON NAME:              | RA Wedge Push-In Suture Anchor System                                               |
| CLASSIFICATION NAME:      | Smooth or threaded metallic bone fixation fastener.<br>(21 CFR 888.3040)            |
| PRODUCT CODE:             | MBI                                                                                 |

## SUBSTANTIAL EQUIVALENCE:

The Responsive Arthroscopy (RA) Wedge Push-In Suture Anchor System is substantially equivalent to the predicate devices in all facets including: function, design, performance, material, and intended use.

Responsive Arthroscopy Wedge Push-In Suture Anchor: K181076 Primary Predicate:

The Wedge Push-In Suture Anchor is of identical size and made from the same basic materials as K181076. The device has identical surgical technique and tissue tensioning to K181076. Anv differences between the Wedge Push-In Suture Anchor and the predicate are considered minor and does not raise questions concerning safety and effectiveness.

### DEVICE DESCRIPTION:

The Responsive Arthroscopy (RA) Wedge Push-In Suture Anchor is a knotless suture anchor for the fixation of soft tissue to bone. The system includes suture anchors made of polyether ether ketone per ASTM F2026 (PEEK) along with repair sutures, passing sutures, inserters, drills, and guide tubes. The suture anchor inserters and instrumentation are available in a variety of geometries and configurations to accommodate various procedures and patient anatomies.

### MATERIALS:

The RA Wedge Push-In Suture Anchor System anchors are machined from extruded PEEK per ASTM F2026. Implantable repair sutures are tapered #2 suture made from Dyneema® ultrahigh molecular weight polyethylene.

{4}------------------------------------------------

## INDICATIONS FOR USE:

The Responsive Arthroscopy Wedge Push-In Suture Anchors are intended to be used for fixation of soft tissue to bone in the shoulder, foot/ankle, hip, knee, hand/wrist, and elbow in the following procedures:

Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.

Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valqus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair/Tendon Repair, Bunionectomy, Digital Tendon Transfers.

Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Vastus Medialis Obliguus Advancement. Patellar Tendon Repair. Posterior Oblique Ligament Repair. Iliotibial Band Tenodesis, Joint Capsule Closure.

Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction, Carpal Ligament Reconstruction, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP Joints for all Digits, Digital Tendon Transfers.

Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis Repair.

Hip: Capsular Repair, Acetabular Labral Repair, Gluteus Medius Repair.

# PERFORMANCE TESTING:

The following bench testing was performed on the Responsive Arthroscopy Wedge Push-In Suture Anchor System:

Suture Diameter Testing Suture Tensile Testing Cyclic Suture Locking Force Testing

In summary, mechanical testing of the Responsive Arthroscopy Wedge Push-In Suture Anchor System indicated no new risks and demonstrated substantial equivalence in performance compared to a legally marketed predicate.

# CONCLUSIONS:

The Responsive Arthroscopy Wedge Push-In Suture Anchor System has shown to be substantially equivalent to legally marketed predicates based on indications for use, technological characteristics, performance testing, and comparison to predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K190446](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K190446)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
