ATTUNE® Porous Fixed Bearing Tibial Base, Medialized Dome Patella and Medialized Anatomic Patella with AFFIXIUM 3DP Technology
Applicant
Depuy Ireland UC
Product Code
MBH · Orthopedic
Decision Date
Jun 18, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3565
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ATTUNE Porous Fixed Bearing Tibial Base, Medialized Dome Patella and Medialized Anatomic Patella with AFFIXIUM 3DP Technology are intended for cementless use within the ATTUNE™ Total Knee Replacement System. Porous coated implants may be used with or without cement. Candidates for total knee replacement include patients with a severely painful and/or impaired knee function resulting from osteoarthritis, post-traumatic arthritis, or a failed previous implant (provided that adequate bone is present).
Device Story
The ATTUNE Porous Fixed Bearing Tibial Base and Medialized Patellae are components of the ATTUNE Total Knee Replacement System. These implants replace natural articular surfaces of the knee joint. The device utilizes AFFIXIUM 3DP technology, which is a 3D-printed titanium porous structure designed to facilitate biological fixation to bone. The implants are intended for cementless use, though they may be used with cement. The device is used by orthopedic surgeons in a clinical/surgical setting to restore knee function in patients with arthritis or previous implant failure. The primary change from the predicate is the use of Grade 5 titanium alloy chemistry (ASTM F-2924) instead of Grade 23 (ASTM F-3001).
Clinical Evidence
No clinical data was required. Substantial equivalence was demonstrated through non-clinical bench testing, including an oxygen content study and a 3D printer recycling study, in accordance with FDA guidance for porous-coated uncemented knee prostheses.
Technological Characteristics
Implants are manufactured from Titanium alloy (Ti-6Al-4V). Subject device uses Grade 5 chemistry (ASTM F-2924); predicate uses Grade 23 (ASTM F-3001). Features AFFIXIUM 3DP technology (3D-printed porous structure) for biological fixation. Designed for cementless or cemented use. No software or electronic components.
Indications for Use
Indicated for patients with severely painful and/or impaired knee function due to osteoarthritis, post-traumatic arthritis, or failed previous implant, provided adequate bone is present.
Regulatory Classification
Identification
A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.
Predicate Devices
ATTUNE® Porous Fixed Bearing Tibial Base, Medialized Dome Patella, and Medialized Anatomic Patella with AFFIXIUM™ 3DP Technology (K202194)
ATTUNE® Porous Fixed Bearing Tibial Base, Medialized Dome Patella, and Medialized Anatomic Patella with AFFIXIUM™ 3DP Technology (K232303)
Related Devices
K202194 — ATTUNE® Porous Fixed Bearing Tibial Base, Medialized Dome Patella, and Medialized Anatomic Patella with AFFIXIUM 3DP Technology · Depuy Ireland UC · Nov 25, 2020
K232303 — ATTUNE® Porous Fixed Bearing Tibial Base, Medialized Dome Patella and Medialized Anatomic Patella with AFFIXIUM 3DP Technology · Depuy Ireland UC · Dec 11, 2023
K213781 — ATTUNE® Porous Fixed Bearing Tibial Base with AFFIXIUM 3DP · Depuy Ireland UC · Jan 18, 2022
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 18, 2024
Depuy Ireland UC Catherine O'Regan Regulatory Affairs Project Lead Loughbeg, Ringaskiddy Cork, Ireland
Re: K240678
Trade/Device Name: ATTUNE® Porous Fixed Bearing Tibial Base, Medialized Dome Patella and Medialized Anatomic Patella with AFFIXIUM™ 3DP Technology Regulation Number: 21 CFR 888.3565 Regulation Name: Knee Joint Patellofemorotibial Metal/Polymer Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: MBH, JWH Dated: April 24, 2024 Received: April 24, 2024
Dear Catherine O'Regan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Lixin Liu -S
Lixin Liu, Ph. D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K240678
Device Name
ATTUNE Porous Fixed Bearing Tibial Base, Medialized DomePatella and Medialized Anatomic Patella with AFFIXIUM 3DP Technology
Indications for Use (Describe)
The ATTUNE Porous Fixed Bearing Tibial Base, Medialized DomePatella and Medialized Anatomic Patella with AFFIXIUM 3DP Technology are intended for cementless use within the ATTUNE™ Total Knee Replacement System. Porous coated implants may be used with or without cement. Candidates for total knee replacement include patients with a severely painful and/or impaired knee function resulting from osteoarthritis, post-traumatic arthritis, or a failed previous implant (provided that adequate bone is present).
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) SUMMARY
(As required by 21 CFR 807.92 and 21 CFR 807.93)
| Contact Details | |
|------------------------------------|-----------------------------------------------------------------------|
| Applicant Name | DePuy Ireland UC |
| Applicant Address | Loughbeg, Ringaskiddy |
| | Co. Cork Munster, IRELAND |
| Applicant Contact Telephone | +35321491400 |
| Applicant Contact | Catherine O'Regan |
| Applicant Contact Email | coregan1@its.jnj.com |
| Correspondent Name | DePuy Ireland UC |
| Correspondent Address | Loughbeg, Ringaskiddy |
| | Co. Cork Munster, IRELAND |
| Correspondent Contact<br>Telephone | +35321491400 |
| Correspondent Contact | Catherine O'Regan |
| Correspondent Contact Email | coregan1@its.jnj.com |
| Date prepared | June 18th, 2024 |
| Device Name | |
| Device Trade Name | ATTUNE® Porous Fixed Bearing Tibial Base, Medialized Dome Patella and |
| | Medialized Anatomic Patella with AFFIXIUM™ 3DP Technology |
| Common Name | Total Knee Prosthesis |
| Classification Name | Knee joint patellofemorotibial metal/polymer porous-coated uncemented |
| | prosthesis |
| | Knee joint patellofemorotibial polymer/metal/polymer semi constrained |
| | cemented prosthesis |
| Regulation Number | 21 CFR 888.3565; 21 CFR 888.3560 |
| Product Code(s) | Primary: MBH |
| | Secondary: JWH |
| Regulatory Class | Class II |
{5}------------------------------------------------
| Legally Marketed Primary Predicate Device | |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| K202194 | ATTUNE® Porous Fixed Bearing Tibial Base, Medialized Dome Patella, and Medialized Anatomic Patella with AFFIXIUM™ 3DP Technology |
| Legally Marketed Secondary Predicate Device | |
| K232303 | ATTUNE® Porous Fixed Bearing Tibial Base, Medialized Dome Patella, and Medialized Anatomic Patella with AFFIXIUM™ 3DP Technology |
| Device Description Summary | |
The ATTUNE® Porous Fixed Bearing Tibial Base, Medialized Dome Patella and Medialized Anatomic Patella with AFFIXIUM™ 3DP Technology are compatible with the ATTUNE Knee System composed of individually packaged femoral, tibial and patellar components designed to replace the natural articular surface of the knee joint. The femoral component is a metal implant with or without porous coating. The tibial component may be comprised of a metal tibial base with or without porous coating, and a polyethylene insert and locking components, or be an all polyethylene device. The patella component may be of an all polyethylene design or a polyethylene patella with porous metal backing.
## Intended Use/Indications for Use
The ATTUNE Porous Fixed Bearing Tibial Base, Medialized Dome Patella and Medialized Anatomic Patella with AFFIXIUM 3DP Technology are intended for cementless use within the ATTUNE™ Total Knee Replacement System. Porous coated implants may be used with or without cement.
Candidates for total knee replacement include patients with a severely painful and/or impaired knee function resulting from osteoarthritis, post-traumatic arthritis, or a failed previous implant (provided that adequate bone is present).
## Indications for use Comparison
The indications for use are the same as the predicate device.
#### Technological Comparison
The subject device ATTUNE Porous Fixed Bearing Tibial Base, Medialized Dome Patella and Medialized Anatomic Patella with AFFIXIUM 3DP Technology are the same as the predicate DePuy ATTUNE Porous Fixed Bearing Tibial Base, Medialized Dome Patella and Medialized Anatomic Patella with AFFIXIUM 3DP Technology (K202194 & K232303) in principle of operation, intended use, classification, design, raw material, and fixation. The only difference between the two devices is that the subject device is output material grade 5 chemistry (ASTM F-2924), whereas the predicate devices is output material grade 23 chemistry (ASTM F-3001). The subject ATTUNE Porous Fixed Bearing Tibial Base, Medialized Dome Patella and Medialized Anatomic Patella with AFFIXIUM 3DP Technology and predicate ATTUNE Porous Fixed Bearing Tibial Base, Medialized Dome Patella and Medialized Anatomic Patella with AFFIXIUM 3DP Technology (K202194 & K232303) are both manufactured from Titanium alloy (Ti-6A1-4V). The subject device are identical to the shape and size configurations for the predicate ATTUNE Porous Fixed Bearing Tibial Base, Medialized Dome Patella and Medialized Anatomic Patella with AFFIXIUM 3DP Technology. The ATTUNE Porous Fixed Bearing Tibial Base, Medialized Dome Patella and Medialized Anatomic Patella with AFFIXIUM 3DP Technology are intended for cementless or cemented use, as is the predicate.
The subject devices and the predicate ATTUNE Porous Fixed Bearing Tibial Base, Medialized Dome Patella and Medialized Anatomic Patella with AFFIXIUM 3DP Technology (K202194 & K232303) both utilize
{6}------------------------------------------------
AFFIXIUM™ 3DP Technology printed titanium porous structure for biological fixation.
K232303 is added as a secondary predicate device for MR compatibility testing and labeling.
Non-Clinical and/or Clinical Tests Summary & Conclusions
The following tests were performed (per FDA's Class II Special Controls Guidance Document: Knee Joint Patell femorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA) to demonstrate substantial equivalence of safety and effectivenss with the predicate devices:
Oxygen Content Study EOS M290 3D Printer Recycling Study
Clinical testing was not required to demonstrate substantial equivalence.
The subject ATTUNE Porous Fixed Bearing Tibial Base, Medialized Dome Patella and Medialized Anatomic Patella with AFFIXIUM 3DP Technology are substantially equivalent to the predicate devices; DePuy ATTUNE Porous Fixed Bearing Tibial Base, Medialized Dome Patella and Medialized Anatomic Patella with AFFIXIUM 3DP Technology (K202194 & K232303).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.