ZIMMER PERSONA THE PERSONALIZED KNEE SYSTEM

K122745 · Zimmer, Inc. · MBH · Dec 5, 2012 · Orthopedic

Device Facts

Record IDK122745
Device NameZIMMER PERSONA THE PERSONALIZED KNEE SYSTEM
ApplicantZimmer, Inc.
Product CodeMBH · Orthopedic
Decision DateDec 5, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3565
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Persona™ Personalized Knee System is a semiconstrained modular knee prosthesis designed to resurface the articulating surface of the femoral, tibial and patellar bones. The Persona Knee System utilizes a modular design between the tibial plates and articular surfaces. The addition of the Trabecular Metal™ femoral component will provide surgeons with the ability to implant with, or without cement (biological fixation). This device is indicated for patients with severe knee pain and disability due to: - Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. - Collagen disorders, and/or avascular necrosis of the femoral condyle. - Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. - Moderate valgus, varus, or flexion deformities. - The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. Porous coated components may be used cemented or uncemented (biological fixation). All other femoral, tibial baseplate, and all-polyethylene (UHMWPE and VEXLPE) patella components are indicated for cemented use only.

Device Story

The Persona™ Personalized Knee System is a modular, semiconstrained knee prosthesis designed to resurface femoral, tibial, and patellar articulating surfaces. It incorporates a Trabecular Metal™ femoral component, allowing for either cemented or uncemented (biological fixation) implantation. The system is intended for use by orthopedic surgeons in clinical settings to address severe knee pain and disability resulting from arthritis, necrosis, or previous surgical failure. By restoring joint configuration and stability, the device aims to improve patient mobility and reduce pain. The system is used during total knee arthroplasty procedures; surgeons select appropriate modular components to match patient anatomy. The device is compatible with MRI environments.

Clinical Evidence

No clinical data was required or provided for this device. Substantial equivalence was established through non-clinical bench testing, including posterior condyle fatigue testing and mechanical/chemical characterization of the Trabecular Metal™ porous surface.

Technological Characteristics

Modular knee prosthesis featuring Trabecular Metal™ porous-coated femoral components. Materials include UHMWPE and VEXLPE for patella components. Designed for cemented or uncemented (biological fixation) use. Testing conducted per FDA guidance for knee joint metal/polymer porous-coated uncemented prostheses and modified metallic surfaces. MRI compatible.

Indications for Use

Indicated for patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle, post-traumatic joint loss, or moderate valgus/varus/flexion deformities. Also indicated for salvage of failed surgical attempts or knees lacking stability in flexion.

Regulatory Classification

Identification

A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.

Special Controls

*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for Zimmer Biomet. The logo consists of a stylized letter "Z" inside of a circle, with the word "zimmer" written in lowercase letters below the circle. The logo is black and white. K122745 1/4 · P.O. Box 708 Warsaw, IN 46581-0708 574 267-6131 ### DEC 0 5 2012 #### Summary of Safety and Effectiveness Sponsor: Contact Person: Date: Trade Name: Product Codes / Device: Regulation Numbers / Description: Predicate Device: Zimmer, Inc. P.O. Box 708 Warsaw, IN 46581-0708 Pauline A. Shand Sr. Specialist, Regulatory Affairs Telephone: (574) 372-4765 Fax: (574) 372-4605 September 20, 2012 Zimmer® Persona™ Personalized Knee System MBH, JWH, OIY 21 CFR & 888.3565 - Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis 21 CFR § 888.3560 - Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented Persona™ Personalized Knee System (K113369, cleared March 27, 2012) NexGen® Trabecular Metal™ Metal Tibial Tray (K072160, cleared September 5, 2007) NexGen® Complete Knee Solution Cruciate Retaining Flex Femoral (CR Flex) Components, manufactured by Zimmer, Inc. (K023211, cleared October 17, 2002) {1}------------------------------------------------ د" - "" - "" K122745₂/₄ NexGen® LPS-Flex Porous Femoral Component, NexGen Gender Solutions Female (GSF), CR-Flex and LPS-Flex Porous Femoral components, manufactured by Zimmer, Inc. (K072619, cleared November 21, 2007 #### Device Description: Intended Use: The Persona™ Personalized Knee System is a semiconstrained modular knee prosthesis designed to resurface the articulating surface of the femoral, tibial and patellar bones. The Persona Knee System utilizes a modular design between the tibial plates and articular surfaces. The addition of the Trabecular Metal™ femoral component will provide surgeons with the ability to implant with, or without cement (biological fixation). This device is indicated for patients with severe knee pain and disability due to: - Rheumatoid arthritis, osteoarthritis, traumatic arthritis, | polyarthritis. - Collagen disorders, and/or avascular necrosis of the । femoral condyle. - Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. - Moderate valgus, varus, or flexion deformities. - - । The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. Porous coated components may be used cemented or uncemented (biological fixation). All other femoral, tibial baseplate, and all-polyethylene (UHMWPE and VEXLPE) patella components are indicated for cemented use only. Comparison to Predicate Device: The proposed Zimmer® Persona™ Personalized Knee System components are similar or identical in intended use, materials, sterility, and performance characteristics to the predicate devices. {2}------------------------------------------------ Page 3 しい! . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Performance Data (Nonclinical and/or Clinical): Clinical Performance and Conclusions: Clinical data and conclusions were not needed for this device. K122745 34 Non-Clinical Performance and Conclusions: Bench testing outlined below was conducted according to FDA guidance documents: FDA Guidance: Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses FDA Guidance: Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement FDA Guidance: Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment | Property or<br>Characteristic | Test Results | |----------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Posterior Condyle Fatigue Testing in a<br>Cantilever Loading Condition of the<br>PersonaTM Tracebular MetalTM<br>femoral component | Demonstrated adequate posterior condyle fatigue strength of<br>the uncemented PersonaTM Tracebular MetalTM femoral<br>component in the cantilever loading condition | | Posterior Condyle Fatigue of<br>PersonaTM Tracebular MetalTM<br>femoral component in three-point<br>Bend Loading Condition | Demonstrated adequate posterior condyle fatigue<br>performance of the PersonaTM Tracebular MetalTM femoral<br>component in a 3-point bend loading condition | | Modified Metallic Surface<br>Characterization for the PersonaTM<br>Tracebular MetalTM Porous femoral<br>knee component with Trabecular<br>MetalTM. | Mechanical, physical, and chemical analyses of Trabecular<br>MetalTM were assessed in accordance to the<br>Guidance Document for Testing Orthopedic Implants with<br>Modified Metallic Surfaces Apposing Bone or Bone Cement | 11 {3}------------------------------------------------ 1999 - 1999 - 1999 | Static Tensile, Static Shear and Shear<br>Fatigue of Trabecular Metal™<br>Diffusion Bonded to Zimaloy-One<br>Hour Cycles, 0.28mm Ti Sheet<br>Interlayer | Demonstrated that one hour diffusion bonding cycles for<br>Trabecular Metal™ produces a bond that met the 20 MPa<br>static tensile test and static shear acceptance criteria. | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Evaluation of Interactions of the<br>Zimmer Legacy Knee and Persona<br>Primary Implant Systems with the<br>Magnetic Fields in the Magnetic<br>Resonance Imaging (MRI)<br>Environment | Demonstrated safety and compatibility within the MRI<br>environment. | and the comments of the comments of . : {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a bold, serif font. The text is centered on the image and is the only element present. Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or a stylized human figure. The seal is black and white. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 Letter dated: December 5, 2012 Zimmer, Incorporated % Ms. Pauline A. Shand Senior Specialist. Regulatory Affairs P.O. Box 708 Warsaw, Indiana 46581-0708 Re: K122745 Trade/Device Name: Zimmer® Persona™ Personalized Knee System Regulation Number: 21 CFR 888.3565 Regulation Name: Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis Regulatory Class: II Product Codes: MBH, JWH, OIY Dated: September 6, 2012 Received: September 7, 2012 Amended: September 24, 2012 Dear Ms. Shand: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {5}------------------------------------------------ Page 2 - Ms. Pauline A. Shand device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Pari 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ## Mark N. Melkerson Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ #### Indications for Use #### 510(k) Number (if known): K122745 #### Device Name: Zimmer® Persona™ Personalized Knee System #### Indications for Use: This device is indicated for patients with severe knee pain and disability due to: - Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. - - Collagen disorders, and/or avascular necrosis of the femoral condyle. l - | Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. - -Moderate valgus, varus, or flexion deformities. - The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. Porous coated components may be used cemented (biological fixation). All other femoral, tibial baseplate, and all-polyethylene (UHMWPE and VEXLPE) patella components are indicated for cemented use only. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (Please do not write below this line - Continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE) ## Krishna Asundi for (Division Sign-Off) Division of Orthopedic Devices # 2012.12.05 11:29:29 -05'00' Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...