K123604 · Howmedica Osteonics Corp. · MAY · Aug 16, 2013 · Orthopedic
Device Facts
Record ID
K123604
Device Name
ABG III MONOLITHIC HIP STEM
Applicant
Howmedica Osteonics Corp.
Product Code
MAY · Orthopedic
Decision Date
Aug 16, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Anato™ Hip Stem is a sterile, single-use device intended for use in primary and revision total and hemi-hip arthroplasty to alleviate pain and restore function. The stem is designed for use with the currently available compatible Howmedica Osteonics' V40™ femoral heads and their compatible acetabular components. Head compatibility with the stems includes: V40130 Biolox Delta, Biolox Delta Universal Taper Heads and Sleeves, V40™ CoCr Heads, V40™ LFITM CoCr Heads, C-Taper Alumina Heads when used with the V40TM /C-taper Adapter, C-Taper Delta Heads when used with C-taper Adaptor, UHR® Universal Head, Unitrax® Heads when used with the Unitrax® V40™ Modular Adapter.
Device Story
Anato™ Hip Stem is a monolithic, non-porous coated femoral hip prosthesis for cementless, press-fit application. Device features anatomic design with 130-degree neck angle and proximal normalization patterns (scales) to convert shear stresses to compressive forces for improved load transmission. Manufactured from Ti-6Al-4V alloy with CP titanium and PureFix® hydroxylapatite (HA) coatings. Available in 8 sizes with neutral and anteverted neck versions; requires specific right/left components. Used by orthopedic surgeons in hospital settings for primary or revision total/hemi-hip arthroplasty. Device integrates with compatible V40™ femoral heads and acetabular components to restore joint function and alleviate pain. Clinical benefit derived from stable fixation and anatomical load distribution.
Clinical Evidence
No clinical data. Substantial equivalence established via non-clinical bench testing, including distal stem fatigue testing (ISO 7206-4) and neck fatigue testing (ISO 7206-6).
Indicated for patients requiring total or hemi-hip arthroplasty due to noninflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, failed prior procedures, or proximal femur fractures (nonunions, femoral neck, trochanteric) unmanageable by other techniques. When used with constrained liners, indicated for patients at high risk of dislocation due to prior history, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability. Intended for cementless use only.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
K993768 — CITATION TMZF HA STEM · Howmedica Osteonics Corp. · Jan 21, 2000
K172684 — Smith & Nephew, Inc. ANTHOLOGY AFIT Hip Stem · Smith & Nephew, Inc. · Oct 3, 2017
K122853 — NOVEL FIT AND FILL STEM · Howmedica Osteonics Corp. · Feb 1, 2013
K051738 — SECUR-FIT MAX AND PLUS MAX HIP STEMS · Howmedica Osteonics Corp. · Jul 19, 2005
K980766 — OSTEONICS ANTEVERTED NECK HA HIP STEMS · Osteonics Corp. · May 27, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
| | 510(k) Summary | | |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Sponsor: | Howmedica Osteonics Corp.<br>325 Corporate Drive<br>Mahwah, NJ 07430 | | |
| Contact Person: | Colleen Burdel<br>Regulatory Affairs Associate<br>t. 201-831-5969<br>f. 201-831-4969 | | |
| Date Prepared: | November 20 , 2012 | | |
| Proprietary Name: | ABG® III Monolithic Hip Stem (also known as<br>Anato™ Hip Stem) | | |
| Common Name: | Hip prosthesis | | |
| Classification Name and Reference: | Hip joint metal/ceramic/polymer semi-constrained<br>cemented or nonporous uncemented prosthesis, 21<br>CFR §888.3353 (Product Codes MAY, LZO, MEH) | | |
| | Hip joint metal/polymer/metal semi-constrained<br>porous coated uncemented prosthesis, 21 CFR<br>§888.3358 (Product Codes LPH, MBL) | | |
| Legally Marketed Device to Which Substantial Equivalence is Claimed:<br>ABG® II Monolithic Hip Stem | K110807 | | |
| ABG® II Monolithic Hip Stem | K110807 |
|-----------------------------|------------------|
| Accolade® II Hip Stem | K103479, K120578 |
#### Device Description:
Howmedica Osteonics is introducing a monolithic, non-porous coated femoral hip prosthesis, which is intended for cementless, press-fit application. The basic design of the Anato™ Hip Stem is similar to two other commerically distributed total hip systems, which are Howmedica Osteonics' ABG® II Monolithic and Accolade® II Hip Stems.
The Anato™ and the Accolade® II stem are both made from the same material combinations. The subject device is similar in design to the ABG® II Monolithic stem as they are both anatomic stems with a 130 degree neck angle. There are normalization patterns (scales) on the anterior and posterior aspects of the proximal end of the stem to facilitate press-fit stability and load transmission. Based on 3-Dimensional simulations, these normalizations are designed to convert medial and lateral shear stresses to compressive forces, which may facilitate proximal loading.
The subject stem is manufactured from titanium (Ti-6Al-4V) alloy, commercially pure (CP) titanium coating, and PureFix® hydroxylapatite (HA) coating identical to the previously cleared Accolade® II Hip Stem.
The Anato™ Hip Stem has a shot peened neck and will be available in 8 sizes ranging from size 1 through 8 with two different neck versions (neutral and anteverted). As the stem is anatomic, there will be a right and left component for each size. The stem is designed only for
{1}------------------------------------------------
use with compatible Howmedica Osteonics' femoral heads and acetabular components.
#### Intended Use:
The Anato™ Hip Stem is a sterile, single-use device intended for use in primary and revision total and hemi-hip arthroplasty to alleviate pain and restore function. The stem is designed for use with the currently available compatible Howmedica Osteonics' V40™ femoral heads and their compatible acetabular components. Head compatibility with the stems includes: V40130 Biolox Delta, Biolox Delta Universal Taper Heads and Sleeves, V40™ CoCr Heads, V40™ LFITM CoCr Heads, C-Taper Alumina Heads when used with the V40TM /C-taper Adapter, C-Taper Delta Heads when used with C-taper Adaptor, UHR® Universal Head, Unitrax® Heads when used with the Unitrax® V40™ Modular Adapter.
## Indications:
The indications for use for total hip arthroplasty include:
- noninflammatory degenerative joint disease, including osteoarthritis . and avascular necrosis;
- rheumatoid arthritis; .
- correction of functional deformity; .
- revision procedures where other treatments or devices have failed; and, ●
- nonunions, femoral neck fractures, and trochanteric fractures of the . proximal femur with head involvement that are unmanageable using other techniques.
Additional indication specific to use of Anato™ Hip Stem with compatible Howmedica Osteonics Constrained Liners:
- When the stem is to be used with compatible Howmedica Osteonics . Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
The Anato™ Hip Stem is intended for cementless use only and is intended for total and hemiarthroplasty procedures.
#### Summary of Technologies:
Device comparison and mechanical testing showed that the proposed device is substantially equivalent in intended use, materials and performance characteristics to the predicate devices.
## Non-Clinical Testing:
The following tests were conducted:
- 1 ) Distal Stem Fatigue Testing in compliance with ISO 7206-4
- 2) Neck Fatigue Testing in compliance with ISO 7206-6
The results of the above testing verify that the new device is substantially equivalent to devices currently cleared for marketing.
# Clinical Testing:
Clinical testing was not required as a basis for substantial equivalence.
{2}------------------------------------------------
.
:
# Conclusion:
. .
·
1. 1. 1. 1.
. . . .
:
The Anato™ Hip Stem is substantially equivalent to the predicate devices identified in this premarket notification.
. . . .
.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 16, 2013
Ms. Colleen Burdel Regulatory Affairs Associate Howmedica Osteonics Corporation 325 Corporate Drive Mahwah, New Jersey 07430
Re: K123604
Trade/Device Name: ABG III Monolithic Hip Stem (also known as Anato™ Hip Stem) Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: MAY. LZO, MEH, LPH, MBL Dated: July 15, 2013 Received: July 16, 2013
Dear Ms. Burdel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRII does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act `s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 – Ms. Colleen Burdel
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours,
# Erin I. Keith
For
Mark N. Melkerson Director Division of Orthopedie Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K123604
Device Name: ABG® III Monolithic Hip Stem (also known as Anato™ Hip Stem)
The indications for use for total hip arthroplasty include:
- 1. noninflammatory degenerative joint disease, including
- osteoarthritis and avascular necrosis;
- 2. rheumatoid arthritis;
- 3. correction of functional deformity;
- 4. revision procedures where other treatments or devices have failed; and.
- 5. nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Additional indication specific to use of Anato™ Hip Stem with compatible Howmedica Osteonics Constrained Liners:
- 1. When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
The Anato™ Hip Stem is intended for cementless use only and is intended for total and hemi-arthroplasty procedures.
| Prescription Use | X |
|-----------------------------|--------|
| (Part 21 CFR 801 Subpart D) | AND/OR |
| Over-The-Counter Use | |
| (21 CFR 807 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Elizabeth L. Frank -S
Division of Orthopedic Devices
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.