OSSIOfiber® Interference Screw

K252022 · OSSIO , Ltd. · MAI · Aug 19, 2025 · Orthopedic

Device Facts

Record IDK252022
Device NameOSSIOfiber® Interference Screw
ApplicantOSSIO , Ltd.
Product CodeMAI · Orthopedic
Decision DateAug 19, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The OSSIOfiber® Interference Screws, are indicated for fixation of tissue, including ligament or tendon to bone, or a bone/tendon to bone in adults and children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by fixation. Interference fixation is appropriate for surgeries of the knee, shoulder, elbow, ankle, foot, and hand/wrist where the sizes offered are patient appropriate; Specifically: • Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction • Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair, Flexor Hallucis Longus for Achilles Tendon reconstruction, tendon transfers in the foot and ankle • Knee: Repair/Reconstruction of the Anterior Cruciate Ligament, Posterior Cruciate Ligament, Medial Collateral Ligament, Lateral Collateral Ligament, Patellar Tendon, Posterior Oblique Ligament, Iliotibial Band Tenodesis, MPFL • Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction • Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty), Carpal Ligament Reconstructions and repairs, tendon transfer in the hand/wrist

Device Story

Cannulated, fully threaded interference screw; composed of poly (L-lactide-co-D,L-lactide) (PLDLA) reinforced with continuous mineral fibers. Used for orthopedic soft tissue-to-bone fixation. Implant degrades via hydrolysis into alpha-hydroxy acids and mineral fibers; load transfers to anatomy during healing; eliminates need for hardware removal. Supplied sterile for single-patient use. Used by orthopedic surgeons in clinical settings. Mechanical performance verified via pull-out, torsional strength, and driving torque testing. Biocompatibility established via reference to previously cleared OSSIOfiber devices.

Clinical Evidence

Bench testing only. Includes pull-out strength, torsional strength, and driving torque testing at time zero and following in-vitro degradation. Biocompatibility established via ISO 10993 data from reference devices.

Technological Characteristics

Material: PLDLA polymer reinforced with continuous mineral fibers. Design: Cannulated, fully threaded screw (6-12 mm diameter, 23-28 mm length). Sterilization: Ethylene Oxide (EtO). Principle: Interference fixation with bioresorbable material. No software/electronics.

Indications for Use

Indicated for fixation of tissue (ligament/tendon to bone or bone/tendon to bone) in adults, children (2-12 years), and adolescents (12-21 years) where growth plates are fused or not crossed by fixation. Applicable for knee, shoulder, elbow, ankle, foot, and hand/wrist surgeries.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION August 19, 2025 OSSIO Ltd. % Dave McGurl Vice President, Regulatory Affairs- Orthopedics MCRA, LLC 803 7th Street NW, Third Floor Washington, District of Columbia 20001 Re: K252022 Trade/Device Name: OSSIOfiber® Interference Screw Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI Dated: June 30, 2025 Received: June 30, 2025 Dear Dave McGurl: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of {1} K252022 - Dave McGurl Page 2 Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the {2} K252022 - Dave McGurl Page 3 Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Christopher Ferreira, M.S. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} OSSIOfiber® Interference Screw Page 10 of 36 | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252022 | ? | | Please provide the device trade name(s). | | ? | | OSSIOfiber® Interference Screw | | | | Please provide your Indications for Use below. | | ? | | The OSSIOfiber® Interference Screws, are indicated for fixation of tissue, including ligament or tendon to bone, or a bone/tendon to bone in adults and children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by fixation. Interference fixation is appropriate for surgeries of the knee, shoulder, elbow, ankle, foot, and hand/wrist where the sizes offered are patient appropriate; Specifically: • Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio- Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction • Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair, Flexor Hallucis Longus for Achilles Tendon reconstruction, tendon transfers in the foot and ankle • Knee: Repair/Reconstruction of the Anterior Cruciate Ligament, Posterior Cruciate Ligament, Medial Collateral Ligament, Lateral Collateral Ligament, Patellar Tendon, Posterior Oblique Ligament, Iliotibial Band Tenodesis, MPFL • Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction • Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty), Carpal Ligament Reconstructions and repairs, tendon transfer in the hand/wrist | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} Page 1 of 3 # 510(k) SUMMARY ## OSSIOfiber® Interference Screw ## Submitter: **Ossio Ltd.** 8 HaTochen Street, Caesarea, Israel, 3079861 Phone: +972-4-9986600 Facsimile: +972-4-9986601 Contact Person: Taly Lindner Date Prepared: August 19, 2025 ## Regulatory Contact: Dave McGurl Vice President, Regulatory Affairs – Orthopedics MCRA, LLC 803 7th St NW, Floor 3 Washington, DC 20001 Office: 202.552.5800 **Name of Device:** OSSIOfiber® Interference Screw **Common or Usual Name:** Fastener, Fixation, Biodegradable, Soft Tissue **Classification Name:** Single/multiple component metallic bone fixation appliances and accessories **Regulatory Class:** Class II, 21 C.F.R. § 888.3030 **Product Code:** MAI ## Primary Predicate: Arthrex FastThread Interference Screw (K202535) ## Additional Predicate: MILAGRO BR Interference Screws (5x23mm, 5x30mm, 6x23mm, 6x30mm) (K240441) ## Reference Devices: OSSIOfiber® Pin Product Family, OSSIOfiber® Compression Screw, OSSIOfiber® Trimmable Fixation Nail (K231272) OSSIOfiber® Small Suture Anchor 2.5-3.5 mm (K243760) OSSIOfiber® Suture Anchor (K251309) ## Purpose of Submission This traditional 510(k) premarket notification is being submitted to obtain clearance for the OSSIOfiber® Interference Screw. ## Device Description The OSSIOfiber® Interference Screw is an orthopedic implant made of poly (L-lactide-co-D,L-lactide) (PLDLA) reinforced with continuous mineral fibers. The OSSIOfiber® Interference Screws are cannulated and fully threaded, available in diameters ranging from 6 to 12 mm and lengths from 23 to 28 mm. The polymer content degrades by hydrolysis into alpha-hydroxy acids that are metabolized by the body. The fibers are made from minerals that are found in natural bone. As the OSSIOfiber® implants degrade, the load transfers to the surrounding anatomy throughout the healing period of the implantation site. Substantial degradation takes place within approximately 18 months as shown in pre-clinical studies, thus eliminating the requirement for future hardware removal surgery. {5} Page 2 of 3 The OSSIOfiber® Interference Screw is supplied sterile, for single patient use only. The OSSIOfiber® Interference Screw is designed to be used with commonly available orthopedic surgical tools such as ISO 9714 compatible instrumentation such as ISO 9714 compatible instrumentations. ## Indications For Use The OSSIOfiber® Interference Screws, are indicated for fixation of tissue, including ligament or tendon to bone, or a bone/tendon to bone in adults and children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by fixation. Interference fixation is appropriate for surgeries of the knee, shoulder, elbow, ankle, foot, and hand/wrist where the sizes offered are patient appropriate; Specifically: - Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction - Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair, Flexor Hallucis Longus for Achilles Tendon reconstruction, tendon transfers in the foot and ankle - Knee: Repair/Reconstruction of the Anterior Cruciate Ligament, Posterior Cruciate Ligament, Medial Collateral Ligament, Lateral Collateral Ligament, Patellar Tendon, Posterior Oblique Ligament, Iliotibial Band Tenodesis, MPFL - Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction - Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty), Carpal Ligament Reconstructions and repairs, tendon transfer in the hand/wrist ## Summary of Technological Characteristics The OSSIOfiber® Interference Screws have the same intended use, and principles of operation, and similar indications for use, material composition and design characteristics as the predicate devices (K202535, K240441). The indications of the subject device are identical to the primary predicate (K202535), except slightly different wording relating to the repair/reconstruction of the knee, which is similar to the wording of the additional predicate (K240441). The OSSIOfiber® Interference Screws include indications for use in pediatric patients, which are not present in the predicate device (K202535, K240441). The difference in the identified patient population does not alter the intended use, as the design, materials, and performance characteristics are appropriate for both adult and pediatric applications. The OSSIOfiber® Interference Screws have identical intended use, material composition, principles of operation, manufacturing and sterilization methods (sterilized by EtO), and similar indications for use as their reference devices (K251309, K243760). The material, manufacturing and sterilization methods of the subject device are also the same as that of the cleared K231272 reference device. The OSSIOfiber® subject devices are available in sizes appropriate for the children and adolescents patient population. Although there are slight design and material differences compared to the primary predicate, mechanical testing demonstrated at least equivalent performance both initially and after in-vitro degradation. Thus, any differences between the subject devices and their predicates do not raise different questions of safety and effectiveness. ## Non-Clinical Data Pull-out testing was performed to verify the strength and fixation properties of the OSSIOfiber® Interference Screws, and to compare them to those of the primary predicate device (K202535). Testing was done initially {6} Page 3 of 3 and following in-vitro degradation. The in-vitro degradation profile was characterized. Torsional strength and driving torque testing at time zero were also conducted for the OSSIOfiber® Interference Screws. Biocompatibility for the subject device was established primarily based on the referenced ISO 10993 data from the previously cleared reference devices (K231272, K251309, K243760) and a rationale. ## Conclusions The OSSIOfiber® Interference Screw is as safe and effective as its primary predicate device, Arthrex FastThread Interference Screw (K202535). The OSSIOfiber® Interference Screw have the same intended use, and principles of operation, and similar indications for use, material composition and design characteristics as the predicate devices. The OSSIOfiber® Interference Screw have identical intended use, material composition, principles of operation, manufacturing and sterilization methods (sterilized by EtO), and similar indications for use as their reference devices (K251309, K243760). Any minor differences do not alter the intended surgical use of the device and do not affect its safety and effectiveness when used as labeled. Non-clinical testing data demonstrate that the OSSIOfiber® Interference Screw is at least as safe and effective as the primary predicate device. Thus, the OSSIOfiber® Interference Screw is substantially equivalent to its predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...