← Product Code [MAI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAI) · K250526

# Arthrex 4.75 mm Double Loaded Knotless Corkscrew Suture Anchor (K250526)

_Arthrex, Inc. · MAI · Mar 14, 2025 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAI/K250526

## Device Facts

- **Applicant:** Arthrex, Inc.
- **Product Code:** [MAI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAI.md)
- **Decision Date:** Mar 14, 2025
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Arthrex 4.75 mm Double Loaded Knotless Corkscrew Suture Anchor is intended for fixation of suture (soft tissue) to bone in the shoulder. Specifically, rotator cuff repairs.

## Device Story

The Arthrex 4.75 mm Double Loaded Knotless Corkscrew Suture Anchor is a sterile, single-use orthopedic implant designed for soft tissue-to-bone fixation in the shoulder. The device consists of a fully threaded anchor body, available in BioComposite (PLLA/βTCP) or PEEK, pre-loaded with sutures on a disposable inserter. It is used by orthopedic surgeons during rotator cuff repair procedures. The anchor is inserted into bone to secure soft tissue without the need for manual knot tying. By providing a secure fixation point, the device facilitates the healing of the rotator cuff to the humeral head, potentially improving patient outcomes in shoulder repair surgeries.

## Clinical Evidence

Bench testing only. Tensile (pull-out) testing demonstrated the device performs statistically equivalent to the predicate. Bacterial endotoxin testing per EP 2.6.14/USP <85> confirmed compliance with pyrogen limit specifications.

## Technological Characteristics

Materials: BioComposite (PLLA/βTCP) or PEEK. Design: Fully threaded knotless suture anchor. Form factor: 4.75 mm diameter. Sterilization: Ethylene Oxide. Connectivity: None (mechanical implant).

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Arthrex Corkscrew FT ([K173788](/device/K173788.md))

## Reference Devices

- Arthrex 3.9 mm Corkscrew Suture Anchor ([K193157](/device/K193157.md))
- Arthrex Double Loaded Knotless FiberTak® Suture Anchor ([K230433](/device/K230433.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

March 14, 2025

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym is in a blue square, and the full name, "U.S. Food & Drug Administration," is in blue text.

Arthrex, Inc. Ruth Segall Regulatory Affairs Specialist II 1370 Creekside Boulevard Naples, Florida 34108

Re: K250526

Trade/Device Name: Arthrex 4.75 mm Double Loaded Knotless Corkscrew Suture Anchor Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: MAI, MBI Dated: February 21, 2025 Received: February 24, 2025

Dear Ms. Segall:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

## CHRISTOPHER FERREIRA -S

Christopher Ferreira, MS Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K250526

Device Name

Arthrex 4.75 mm Double Loaded Knotless Corkscrew Suture Anchor

Indications for Use (Describe)

The Arthrex 4.75 mm Double Loaded Knotless Corkscrew Suture Anchor is intended for fixation of suture (soft tissue) to bone in the shoulder. Specifically, rotator cuff repairs.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary

| Date Prepared                             | 03/14/2025                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number                             | K250526                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Submitter                                 | Arthrex Inc.<br>1370 Creekside Boulevard<br>Naples, FL 34108-1945                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Contact Person                            | Name: Ruth Segall<br>Title: Regulatory Affairs Specialist II<br>Phone: 239-598-4302 x 71764<br>Email: Ruth.Segall@arthrex.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Trade Name                                | Arthrex 4.75mm Double Loaded Knotless Corkscrew®<br>Suture Anchor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Common Name                               | Suture Anchor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Product Code                              | MAI, MBI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Classification Name                       | 21 CFR 888.3030: Single/multiple component metallic<br>bone fixation appliances and accessories<br>21 CFR 888.3040: Smooth or threaded metallic bone<br>fixation fastener                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Regulatory Class                          | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Primary Predicate Device                  | K173788: Arthrex Corkscrew FT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Additional Predicate Devices              | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Reference Devices                         | K193157: Arthrex 3.9 mm Corkscrew Suture Anchor<br>K230433: Arthrex Double Loaded Knotless FiberTak®<br>Suture Anchor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Purpose of Submission                     | This Special 510(k) premarket notification is submitted<br>to obtain clearance for the Arthrex 4.75 mm Double<br>Loaded Knotless Corkscrew Suture Anchor as a line<br>extension to the Arthrex Corkscrew Suture Anchors<br>cleared via K173788.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Device Description                        | The Arthrex 4.75 mm Double Loaded Knotless<br>Corkscrew Suture Anchor is a fully threaded knotless<br>suture anchor that is pre-loaded on a disposable<br>inserter with Arthrex sutures. The anchor may be<br>manufactured from either BioComposite (PLLA/βTCP<br>TriCalcium Phosphate) or PEEK (Polyether-ether-<br>ketone)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                           | The Arthrex 4.75 mm Double Loaded Knotless<br>Corkscrew Suture Anchor is provided sterile (Ethylene<br>Oxide), single-use, and is packaged in a single-pack or 5-<br>pack.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Indications for Use                       | The Arthrex 4.75 mm Double Loaded Knotless<br>Corkscrew Suture Anchor is intended for fixation of<br>suture (soft tissue) to bone in the shoulder. Specifically,<br>rotator cuff repairs.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Performance Data                          | Tensile (pull-out) testing was conducted on the<br>proposed Arthrex 4.75 mm Doubled Loaded Knotless<br>Corkscrew Suture Anchor submitted in this Special<br>510(k). The test data demonstrates that the proposed<br>device performs statistically equivalent to the predicate<br>device for the intended indications for use.<br>Bacterial endotoxin per EP 2.6.14/USP <85> was<br>conducted to demonstrate that the device meets<br>pyrogen limit specifications.                                                                                                                                                                                                                                                                                                                                                                                            |
| Technological and Predicate<br>Comparison | The Arthrex 4.75 mm Double Loaded Knotless<br>Corkscrew Suture Anchor are a line extension to the<br>predicate device K173788. The indications for use for<br>the subject device are a subset of the indications of the<br>predicate device. The proposed and predicate devices<br>have identical basic design, indications for use, intended<br>use, packaging, shelf-life, biocompatibility profile,<br>manufacturing, and sterilization processes. In<br>comparison to the predicate device, the proposed<br>modifications include minor modifications to the<br>method of fixation of the suture to the anchor, creating<br>a closed loop within the anchor body, and the pre-<br>loaded suture. Any differences between the proposed<br>and predicate devices are considered minor and do not<br>raise any questions concerning safety or effectiveness. |
| Conclusion                                | The Arthrex 4.75 mm Double Loaded Knotless<br>Corkscrew Suture Anchor is substantially equivalent to<br>the predicate devices in which the basic design features<br>and intended use are identical. The proposed device is<br>indicated for a subset of the indications for use of the                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |

{5}------------------------------------------------

{6}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAI/K250526](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAI/K250526)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
