SF Push- in Anchor

K240288 · Surgical Fusion Technologies GmbH · MAI · Apr 1, 2024 · Orthopedic

Device Facts

Record IDK240288
Device NameSF Push- in Anchor
ApplicantSurgical Fusion Technologies GmbH
Product CodeMAI · Orthopedic
Decision DateApr 1, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The SF Push-in Anchor is intended to be used for suture or tissue fixation in the foot, ankle, hand, wrist, elbow, hip, knee and shoulder. The SF Push-in Anchor is designed only to be inserted with the SupraFuser Generator.

Device Story

SF Push-in Anchor is a bioresorbable orthopedic fixation system for soft tissue-to-bone reattachment. System comprises implantable anchors (1.6mm–3.6mm), SupraFuser ultrasonic generator, sonotrode, and dedicated drills. During surgery, surgeon inserts anchor into pre-drilled bone hole; ultrasonic energy from generator liquefies Poly-L-lactide-co-D,L-lactide (PLDLLA) polymer at bone interface. Liquid polymer flows into cancellous bone marrow space, quenching immediately to provide mechanical anchorage. Suture fixation is achieved via manual knotting. Used in clinical settings by orthopedic surgeons for various ligament and tendon repairs across multiple joints. Benefits include secure fixation via polymer integration and full bioresorbability of the implant.

Clinical Evidence

Bench testing only. Testing included software validation, IEC 60601-1/60601-1-2 electrical safety/EMC, material characterization (GPC, viscosity, GCMS, ICP), static and dynamic pullout testing (pre/post-aging), creep testing, insertion testing, degradation testing, and biocompatibility risk assessment. Lot-to-lot endotoxin testing performed per USP <85>.

Technological Characteristics

Materials: Poly-L-lactide-co-D,L-lactide (PLDLLA). Sizes: 1.6, 2.3, 3.0, 3.6mm diameters. Principle: Ultrasonic energy-induced polymer liquefaction for bone anchorage. Components: Anchor, sonotrode, SupraFuser generator, drills, stoppers, guide. Connectivity: System-specific generator. Sterilization: Sterile. Software: Embedded firmware for ultrasonic energy control.

Indications for Use

Indicated for suture or tissue fixation in foot, ankle, hand, wrist, elbow, hip, knee, and shoulder. Specific procedures include ligament reconstruction (scapholunate, carpal, collateral), tendon repair/reattachment (flexor, extensor, biceps, Achilles, patellar), Bankart/SLAP lesion repair, capsular shift/reconstruction, stabilization (lateral, medial), hallux valgus/bunionectomy, and iliotibial band tenodesis.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the date April 1, 2024. The text is in a simple, sans-serif font. The date is written out in full, with the month, day, and year clearly visible. The text is black on a white background. Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Surgical Fusion Technologies GmbH % Kelliann Payne Regulatory Counsel Hogan Lovells US LLP 1735 Market Street, 23rd Floor Philadelphia, Pennsylvania 19103 Re: K240288 Trade/Device Name: SF Push- in Anchor Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: MAI, HTY, GAT Dated: January 31, 2024 Received: February 1, 2024 Dear Kelliann Payne: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Image /page/1/Picture/7 description: The image shows a closing statement followed by a name. The closing statement is "Sincerely,". The name below the closing statement is "Maziar Shah Mohammadi". The text is left-aligned and uses a serif font. gitally signed by Maziar Shal Date: 2024.04.01 10:52:53 -04'00 For: Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ Enclosure {3}------------------------------------------------ # Indications for Use Submission Number (if known) K240288 Device Name SF Push- in Anchor ## Indications for Use (Describe) The SF Push-in Anchor is intended to be used for suture or tissue fixation in the foot, ankle, hand, wrist, elbow, hip, knee and shoulder. The SF Push-in Anchor is designed only to be inserted with the SupraFuser Generator. SF Push-in Anchor 1.6mm: Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/ Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers Elbow: Ulnar or Radial Collateral Ligament Reconstruction Shoulder: Bankart Repair, SLAP Lesion Repair, Capsular Shift or Capsulolabral Reconstruction Foot/Ankle: Lateral Stabilization, Medial Stabilization, Metatarsal Ligament Repair, Hallux Valgus reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair SF Push-in Anchor 2.3mm: Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/ Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction, Lateral epicondylitis repair Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction Foot/Ankle: Lateral Stabilization, Metatarsal Ligament Repair, Hallux Valgus reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction, Achilles Tendon Repair, Bunionectomy Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis Hip: Acetabular labral repair, capsular repair SF Push-in Anchor 3.0mm and 3.6mm: Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/ Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and {4}------------------------------------------------ MCP joints for all digits, digital tendon transfers Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction, lateral epicondylitis repair Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction, Metatarsal tendon repair, Bunionectomy Knee: Medial collateral ligament repair, lateral collateral ligament repair, patellar tendon repair, posterior oblique ligament repair, and iliotibial band tenodesis Hip: Acetabular labral repair, capsular repair Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ ## Surgical Fusion Technologies GmbH's SF-Push-in Anchors ## 510(k) Summary ## Surgical Technology's SF Push-in Anchor Portfolio ## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared Surgical Fusion Technologies GmbH Wagistrasse 6 Schlieren, Switzerland Phone: +41 44 204 60 18 Facsimile: +41 44 204 61 2820 Contact Person: Joerg Mayer, CTO Date Prepared: March 29, 2024 #### Name of Device and Name SF Push-in Anchor ### Common or Usual Name Fastener, fixation, biodegradable, soft tissue (MAI), class II Pin, fixation, smooth (HTY), class II Suture, nonabsorbable, synthetic, polyethylene (GAT), class II #### Classification Name Single/multiple component metallic bone fixation appliances and accessories (21 CFR 888.3030) Smooth or threaded metallic bone fixation fastener (21 CFR 888.3040) Nonabsorbable poly(ethylene terephthalate) surgical suture (21 CFR 878.5000) #### Predicate Devices - K171228 SportWelding's Fiji Anchor (Primary Predicate) - K063479 Arthrex's Bio-PushLock (Secondary Predicate) - K082810 Arthrex's Bio Composite Pushlock (Secondary Predicate) - K183091 SpineWelding's Elaris Pedicle Screw System (Reference Device) . #### Indications for Use The SF Push-in Anchor is intended to be used for suture or tissue fixation in the foot, ankle, hand, wrist, elbow, knee, hip and shoulder. The SF Push-in Anchor is designed only to be inserted with the SupraFuser Generator. {6}------------------------------------------------ ## SF Push-in Anchor 1.6mm: Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP. DIP and MCP ioints for all digits, digital tendon transfers Elbow: Ulnar or Radial Collateral Ligament Reconstruction Shoulder: Bankart Repair, SLAP Lesion Repair, Capsular Shift or Capsulolabral Reconstruction Foot/Ankle: Lateral Stabilization, Medial Stabilization, Metatarsal Ligament Repair, Hallux Valqus reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair ## SF Push-in Anchor 2.3mm: Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction, Lateral epicondylitis repair Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction Foot/Ankle: Lateral Stabilization, Medial Stabilization, Metatarsal Ligament Repair, Hallux Valqus reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction, Achilles Tendon Repair, Bunionectomy Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis Hip: Acetabular labral repair, capsular repair #### SF Push-in Anchor 3.0mm and 3.6mm: Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction, lateral epicondylitis repair Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction {7}------------------------------------------------ Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus and Varus reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction, Metatarsal tendon repair, Bunionectorny Knee: Medial collateral ligament repair, lateral ligament repair, patellar tendon repair, posterior oblique ligament repair, and iliotibial band tenodesis Hip: Acetabular labral repair, capsular repair #### Technological Characteristics The SF Push-in Anchor system consists of implantable anchors with 1.6, 2.3, 3.0, and 3.6mm diameter, an ultrasonic system, a sonotrode, dedicated anchor size specific drills and stoppers, a handpiece front cover, and a guide with integrated wrench. The SF Push-in Anchors are made of biocompatible and fully bioresorbable Poly-L-lactide-co-D,L-lactide. The SF Push-in Anchors are fully bioresorbable implants designed for soft tissue reattachment to bone by means of suture materials. The implantation process employs ultrasonic energy to liquefy the polymeric components of the Push-in Anchors at the interface with bone tissue. The liquid polymer flows into the marrow space of the surrounding cancellous bone, where it is immediately quenched and provides anchorage of the implant. The ultrasonic energy for the implantation of the Push-in Anchors is produced by the SupraFuser® ultrasonic generator and applied via the attached handpiece. The sonotrode is mounted on the handpiece. It transmits the ultrasonic vibrations to the Push-in Anchor. The drills and the stoppers are dedicated to be used with the Push-in Anchor of the respective size. #### Performance Data The following tests were performed to demonstrate the substantial equivalence of the Fiji Anchor: - Software documentation and validation per FDA guidance ● - IEC 60601-1 and 60601-1-2 testing - Material characterization using GPC, inherent viscosity, GCMS, and ICP - Static pullout testing before and after aging ● - Dynamic pullout testing before and after aging - Creep testing - Insertion testing - . Degradation testing - Biocompatibility risk assessment . - Lot-to-lot endotoxin testing is also performed per USP <85> ● {8}------------------------------------------------ ### Substantial Equivalence The SF Push-in Anchors are as safe and effective as the primary predicate device by the same manufacturer (Fiji Anchors). The SF Push-in Anchors have the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices (Arthrex's 2.5mm Bio PushLock, Bio-Composite PushLock). The minor technological differences between the SF Push-in Anchors and its predicate devices, summarized in the table below, raise no new issues of safety or effectiveness. Performance data demonstrate that the SF Push-in Anchors are as safe and effective as the Fiji Anchor as well as Arthrex's 2.5mm Bio PushLock or Bio-Composite PushLock. Thus, the SF Push-in Anchors are substantially equivalent. The reference device is cited to support testing methods and results. | Property | SF Push- in Anchor | SportWelding Fiji<br>Anchor/SF Push-in anchor<br>2.3 | Arthrex 2.5 mm Bio-<br>PushLock | Arthrex Bio-Composite<br>PushLock | |------------------|------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------| | Predicate | Subject | Primary Predicate | Secondary Predicate | Secondary Predicate | | 510(k) | Subject | K171228 | K063497 | K082810 | | Components | Anchor, suture, Sonotrode,<br>US generator | Anchor, suture, sonotrode,<br>US generator | PEEK eyelet, suture, anchor | PEEK eyelet, suture, anchor | | Size | 1.6mm/5.2mm<br>2.3mm/7.2mm<br>3.0mm/8.7mm<br>3.6mm/10.2mm | 2.3 x 7.2 mm | 2.5 x 8 mm | Diameter: 2.4-6.5mm/<br>Length: 5-18.5mm | | Suture Sizes | 1.6mm: #4-0 to #0<br>2.3mm: #4-0 to #2<br>3.0mm: #0 to #2, tape up to<br>size 2mm<br>3.6mm: #2, tape up to size<br>2mm | #4-0 to #2 | #2-0. #0 | #2-0. #0, #2,tape up to size<br>2mm | | Insertion Method | Insert anchor into predrilled<br>hole while applying US<br>energy | Insert anchor into predrilled<br>hole while applying US<br>energy | Insert eyelet into predrilled<br>hole, then secure with<br>bioresorbable anchor | Insert eyelet into predrilled<br>hole, then secure with<br>bioresorbable anchor | {9}------------------------------------------------ | Property | SF Push- in Anchor | SportWelding Fiji<br>Anchor/SF Push-in anchor<br>2.3 | Arthrex 2.5 mm Bio-<br>PushLock | Arthrex Bio-Composite<br>PushLock | |--------------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------|-------------------------------------|-------------------------------------| | Fixation Method | Ultrasonic melting of the<br>polymer into the porous<br>cancellous bone | Ultrasonic melting of the<br>polymer into the porous<br>cancellous bone | Barbs in a mechanical press-<br>fit | Barbs in a mechanical press-<br>fit | | Length of<br>ultrasound energy<br>delivery | Not more than 6 seconds | Not more than 6 seconds | N/A | N/A | | Suture fixation | Manual knotting | Manual knotting | Knotless | Knotless | | Material | PLDLLA | PLDLLA | PLLA and PEEK | PLLA, PLDLA/TCP, PEEK | | Resorbable | Yes | Yes | Partly | Partly | | Sterilization | Sterile | Sterile | Sterile | Sterile | ## Conclusions The SF Push-in Anchors are substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...