← Product Code [LZO](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LZO) · K243029

# Origin TT Stem (K243029)

_Signature Orthopaedics Pty, Ltd. · LZO · Mar 11, 2025 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LZO/K243029

## Device Facts

- **Applicant:** Signature Orthopaedics Pty, Ltd.
- **Product Code:** [LZO](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LZO.md)
- **Decision Date:** Mar 11, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3353
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Components of the Signature Orthopaedics hip replacement range are intended to replace a hip joint where bone stock is sufficient to support the implant. When a surgeon has selected prosthetic replacement as the preferred treatment, the devices are indicated for: - Non-inflammatory degenerative joint disease including osteoarthritis or avascular necrosis - Inflammatory joint disease including rheumatoid arthritis - Correction of functional deformity including congenital hip dysplasia - Traumatic injury involving the hip joint including traumatic arthritis or femoral head or neck fracture - Failed previous hip surgery including internal fixation or joint fusion, reconstruction, hemiarthroplasty, surface replacement, or total replacement Signature Orthopaedics' Origin TT, Aria, Remedy, Origin-NS, Pegasus, Spartan, World and Everglade Hip femoral stems, SignaSure Cementless Cups, Logical and World Acetabular Cups are intended for cementless fixation only. Signature Orthopaedics' Evolve, Cemented TSI (both CoCr and HNSS variants), Origin Cemented femoral stems and SignaSure Cemented Cups are intended for cemented fixation only. Signature Orthopaedics' SignaSure Logical/World Metal Insert is indicated for use with a cementless Signature Orthopaedics' Logical/World Acetabular Cup to provide dual mobility articulation. Signature Orthopaedics' constrained liner components are indicated particularly for patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability. Signature Orthopaedics' Evolve UniPolar Head and BiPolar Head are intended for hemi- hip arthroplasty only, where the natural acetabulum does not require replacement. The Evolve UniPolar Head are indicated for bone fractures or pathologies involving only the femoral head/neck and/or proximal femur, such as: - Acute femoral head or neck fracture - Fracture dislocation of the hip - Avascular necrosis of the femoral head - Non-union of femoral neck fractures - Certain high subcapital and femoral neck fractures in the elderly - Degenerative arthritis involving only the femoral head

## Device Story

Origin™ TT Stem is a single-use, cementless femoral hip prosthesis; manufactured from titanium alloy; features proximal plasma spray coating of titanium and hydroxyapatite; utilizes 12/14 taper for modular head attachment. Device implanted by orthopedic surgeons in clinical settings to replace diseased or damaged hip joints. Functions as structural support for hip joint; restores joint function; alleviates pain; improves mobility. Surgeon selects device based on patient bone stock and pathology; implant integrates with patient bone via cementless fixation. Benefits include restoration of hip joint mechanics and stability.

## Clinical Evidence

Bench testing only. Evaluations included range of motion, stem and neck fatigue testing, and finite element analysis (FEA). Testing conducted per ISO 21535, ASTM F2996, ISO 7206-4, and ISO 7206-6. Results demonstrate sufficient strength and performance for intended use.

## Technological Characteristics

Titanium alloy (ISO 5832-3); proximal plasma spray coating of titanium (ASTM F1580) and hydroxyapatite (ISO 13779-2); 12/14 taper; cementless fixation; modular femoral stem design.

## Regulatory Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

## Predicate Devices

- Signature Orthopaedics Spartan Hip Stem ([K192883](/device/K192883.md))
- Signature Orthopaedics Encore Hip Stem ([K211505](/device/K211505.md))

## Reference Devices

- Signature Orthopaedics Origin Hip Stem ([K121297](/device/K121297.md))
- DePuy Actis DuoFix Hip Stem ([K150862](/device/K150862.md))
- DePuy Actis DuoFix Hip Stem ([K160907](/device/K160907.md))
- Signature CoCr Femoral Heads ([K121297](/device/K121297.md), [K163081](/device/K163081.md))
- Ceramic Femoral Heads ([K190704](/device/K190704.md))
- Logical™ PX, G and C Series Acetabular Cups ([K121297](/device/K121297.md), [K153131](/device/K153131.md))
- Logical™ Liners ([K121297](/device/K121297.md), [K153131](/device/K153131.md), [K241690](/device/K241690.md))
- Logical™ Constrained Liners ([K153131](/device/K153131.md))
- BiPolar Heads ([K133370](/device/K133370.md), [K163081](/device/K163081.md))
- Evolve™ UniPolar Heads ([K143184](/device/K143184.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

March 11, 2025

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Signature Orthopaedics Pty Ltd % Robert Poggie President BioVera, Inc. 65 Promenade Saint Louis Notre-Dame-de-L'Ile-Perrot, OC J7W3J6 Canada

Re: K243029

Trade/Device Name: Origin™ TT Stem Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO, MEH, OQI, LPH, OQG, KWZ, KWL, KWY Dated: January 31, 2025 Received: February 3, 2025

Dear Robert Poggie:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Limin Sun -S

Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K243029

Device Name Origin TT Stem

## Indications for Use (Describe)

Components of the Signature Orthopaedics hip replacement range are intended to replace a hip joint where bone stock is sufficient to support the implant. When a surgeon has selected prosthetic replacement as the preferred treatment, the devices are indicated for:

- · Non-inflammatory degenerative joint disease including osteoarthritis or avascular necrosis
- · Inflammatory joint disease including rheumatoid arthritis
- · Correction of functional deformity including congenital hip dysplasia
- · Traumatic injury involving the hip joint including traumatic arthritis or femoral head or neck fracture

· Failed previous hip surgery including internal fixation or joint fusion, reconstruction, hemiarthroplasty, surface replacement, or total replacement

Signature Orthopaedics' Origin TT, Aria, Remedy, Origin-NS, Pegasus, Spartan, World and Everglade Hip femoral stems, SignaSure Cementless Cups, Logical and World Acetabular Cups are intended for cementless fixation only.

Signature Orthopaedics' Evolve, Cemented TSI (both CoCr and HNSS variants), Origin Cemented femoral stems and SignaSure Cemented Cups are intended for cemented fixation only.

Signature Orthopaedics' SignaSure Logical/World Metal Insert is indicated for use with a cementless Signature Orthopaedics' Logical/World Acetabular Cup to provide dual mobility articulation.

Signature Orthopaedics' constrained liner components are indicated particularly for patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability.

Signature Orthopaedics' Evolve UniPolar Head and BiPolar Head are intended for hemi- hip arthroplasty only, where the natural acetabulum does not require replacement. The Evolve UniPolar Head are indicated for bone fractures or pathologies involving only the femoral head/neck and/or proximal femur, such as:

- · Acute femoral head or neck fracture
- · Fracture dislocation of the hip
- · Avascular necrosis of the femoral head
- · Non-union of femoral neck fractures
- · Certain high subcapital and femoral neck fractures in the elderly
- · Degenerative arthritis involving only the femoral head

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image features the logo of SignatureOrtho. The logo consists of a green circular icon with a white figure inside, followed by the text "SignatureOrtho" in gray. The letters of "Ortho" are in a lighter shade of gray, and there is a superscript "TM" next to the word.

# 510(k) SUMMARY

| Manufacturer:         | Signature Orthopaedics Pty Ltd                                                                                                         |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------|
|                       | 7 Sirius Road                                                                                                                          |
|                       | Lane Cove, NSW 2066                                                                                                                    |
|                       | Australia                                                                                                                              |
| Device Trade<br>Name: | Origin™ TT Stem                                                                                                                        |
| Common Name:          | Femoral Hip Prosthesis                                                                                                                 |
| Contact:              | Dr. Declan Brazil                                                                                                                      |
|                       | Managing Director of Signature Orthopaedics                                                                                            |
| Prepared By:          | Signature Orthopaedics Pty Ltd                                                                                                         |
|                       | 7 Sirius Road                                                                                                                          |
|                       | Lane Cove, NSW 2066                                                                                                                    |
|                       | Australia                                                                                                                              |
|                       | Phone: +61 (2) 9428 5181                                                                                                               |
|                       | Fax: +61 (2) 8456 6065                                                                                                                 |
| Date Prepared:        | January 31, 2025                                                                                                                       |
| Classification:       | Class II per 21 CFR 888.3353: Hip joint metal/ceramic/polymer<br>semi-constrained cemented or nonporous uncemented prosthesis<br>(LZO) |
|                       | Class II per 21 CFR 888.3353: Hip joint metal/ceramic/polymer<br>semi-constrained cemented or nonporous uncemented prosthesis<br>(MEH) |
|                       | Class II per 21 CFR 888.3353: Hip joint metal/ceramic/polymer<br>semi-constrained cemented or nonporous uncemented prosthesis<br>(OQI) |
|                       | Class II per 21 CFR 888.3358: Hip joint metal/polymer/metal semi-<br>constrained porous-coated uncemented prosthesis (LPH)             |
|                       | Class II per 21 CFR 888.3358: Hip joint metal/polymer/metal semi-<br>constrained porous-coated uncemented prosthesis (OQG)             |
|                       | Class II per 21 CFR 888.3310: Hip joint metal/polymer constrained<br>cemented or uncemented prosthesis (KWZ)                           |
|                       | Class II per 21 CFR 888.3360: Hip joint femoral (hemi-hip)<br>metallic cemented or uncemented prosthesis (KWL)                         |
|                       | Class II per 21 CFR 888.3390: Hip joint femoral (hemi-hip)<br>metal/polymer cemented or uncemented prosthesis (KWY)                    |
| Predicate Devices:    | Substantial equivalence to the following devices is claimed:                                                                           |
|                       | • Signature Orthopaedics Spartan Hip Stem (K192883)                                                                                    |
|                       | • Signature Orthopaedics Encore Hip Stem (K211505)                                                                                     |

- Signature Orthopaedics Origin Hip Stem (K121297)
 DePuy Actis DuoFix Hip Stem (K150862 and K160907)

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for SignatureOrtho. The logo features a green circle with a white abstract figure inside, followed by the text "SignatureOrtho" in a gray, sans-serif font. The letters are bolded, and there is a trademark symbol after the word "Ortho".

## Device Description:

The Signature Orthopaedics Origin™ TT Stem is intended for single use and cementless fixation. It is manufactured from titanium alloy as per ISO 5832-3 and has a proximal plasma spray coating of titanium as per ASTM F1580 and hydroxyapatite as per ISO 13779-2. It features a 12/14 taper and is compatible for use with Signature CoCr Femoral Heads (K121297 and K163081), Ceramic Femoral Heads (K190704), Logical™ PX, G and C Series Acetabular Cups (K121297 and K153131), Logical™ Liners (K121297, K153131 and K241690), Logical™ Constrained Liners (K153131), BiPolar Heads (K133370 and K163081) and Evolve™ UniPolar Heads (K143184).

## Indications for Use:

Components of the Signature Orthopaedics hip replacement range are intended to replace a hip joint where bone stock is sufficient to support the implant. When a surgeon has selected prosthetic replacement as the preferred treatment, the devices are indicated for:

- · Non-inflammatory degenerative joint disease including osteoarthritis or avascular necrosis
- Inflammatory joint disease including rheumatoid arthritis .
- Correction of functional deformity including congenital hip dysplasia .
- Traumatic injury involving the hip joint including traumatic arthritis or femoral head . or neck fracture
- Failed previous hip surgery including internal fixation or joint fusion, reconstruction, . hemiarthroplasty, surface replacement, or total replacement

Signature Orthopaedics' Origin TT, Aria, Remedy, Origin-NS, Pegasus, Spartan, World and Everglade Hip femoral stems, SignaSure Cementless Cups, Logical and World Acetabular Cups are intended for cementless fixation only.

Signature Orthopaedics' Evolve, Cemented TSI (both CoCr and HNSS variants), Origin Cemented femoral stems and SignaSure Cemented Cups are intended for cemented fixation onlv.

Signature Orthopaedics' SignaSure Logical/World Metal Insert is indicated for use with a cementless Signature Orthopaedics' Logical/World Acetabular Cup to provide dual mobility articulation.

Signature Orthopaedics' constrained liner components are indicated particularly for patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability.

Signature Orthopaedics' Evolve UniPolar Head and BiPolar Head are intended for hemihip arthroplasty only, where the natural acetabulum does not require replacement. The Evolve UniPolar Head and BiPolar Head are indicated for bone fractures or pathologies involving only the femoral head/neck and/or proximal femur, such as:

- · Acute femoral head or neck fracture
- · Fracture dislocation of the hip
- · Avascular necrosis of the femoral head
- · Non-union of femoral neck fractures
- · Certain high subcapital and femoral neck fractures in the elderly
- · Degenerative arthritis involving only the femoral head

### Comparison of Technological Characteristics:

The subject and predicate device have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are similar between the subject and predicates:

- . Indications for Use
- Materials of manufacture .
- Structure support mechanism .
- · Principle of operation
- Sizes .
- · Design

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for SignatureOrtho. The logo consists of a green circle with a white figure inside, followed by the text "SignatureOrtho" in gray. The "TM" symbol is located in the upper right corner of the logo.

## Performance Testing:

Non-clinical testing and engineering evaluations were conducted to verify that the performance of the Signature Orthopaedics Origin™ TT Stem is adequate for anticipated in-vivo use. The following non-clinical testing was carried out on the worst-case sizes of the subject devices:

- Range of Motion .
- Stem and Neck Fatigue FEA .
- Stem and Neck Fatigue Testing .

Testing was performed in accordance with the following testing standards:

- ISO 21535 Non-active surgical implants Joint replacement implants Specific . requirements for hip joint replacement implants
- ASTM F2996 Standard Practice for Finite Element Analysis (FEA) of Non-Modular . Metallic Orthopaedic Hip Femoral Stems
- ISO 7206-4 Partial and total hip joint prosthesis Part 4: Determination of endurance . properties and performance of stemmed femoral components
- ISO 7206-6 Partial and total hip joint prostheses Part 6: Endurance properties testing . and performance requirements of neck region of stemmed femoral components

The results of non-clinical testing show that the strength of the Signature Orthopaedics Origin™ TT Stem is sufficient for their intended use and substantially equivalent to the legally marketed predicate device.

### Substantial Equivalence:

The Origin™ TT Stem has the same intended use and same indications for use as the predicate device. The subject devices use the same operating principle, incorporate the same basic design, and are manufactured using the same materials as the predicate device.

Any differences do not raise new questions of safety and effectiveness as established with performance testing. The subject devices are at least as safe and effective as the legally marketed predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LZO/K243029](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LZO/K243029)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
