← Product Code [LPH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH) · K242375

# OR3O Dual Mobility System (K242375)

_Smith & Nephew, Inc. · LPH · Oct 7, 2024 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K242375

## Device Facts

- **Applicant:** Smith & Nephew, Inc.
- **Product Code:** [LPH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH.md)
- **Decision Date:** Oct 7, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3358
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The OR30 Dual Mobility System is intended for use in primary and revision total hip arthroplasty in skeletally mature patients. Indications · Advanced degeneration of the hip joint as a result of degenerative, post-traumatic, or rheumatoid arthritis. - · Fracture or avascular necrosis of the femoral head. - · Failure of previous hip surgery: ioint reconstruction, arthrodesis, hemiarthroplasty, - surface replacement arthroplasty, or total hip replacement. - · All forms of osteoarthritis. Patients with hips at risk of dislocation. - · Femoral neck fracture or proximal hip joint fracture. The OR30 Dual Mobility System is intended for single use only. The modular OR30 Liners and Inserts are to be implanted without bone cement.

## Device Story

Modular dual mobility hip implant system; consists of metal acetabular shell, diffusion-hardened oxidized zirconium (OXINIUM DH) alloy liner, cross-linked polyethylene (XLPE) insert, and femoral head. Used in primary and revision total hip arthroplasty; implanted without bone cement. Device provides two articulating surfaces to address hip joint degeneration, fractures, or previous surgical failure. Operated by orthopedic surgeons in clinical/OR settings. Subject submission updates MR safety labeling and package inserts to define MR Conditional status.

## Clinical Evidence

Bench testing only. MRI compatibility testing performed per ASTM F2052 (displacement force), ASTM F2213 (torque), ASTM F2182 (RF heating), and ASTM F2119 (image artifact) to support MR Conditional labeling.

## Technological Characteristics

Modular hip prosthesis; materials include metal acetabular shell, diffusion-hardened oxidized zirconium (OXINIUM DH) alloy liner, and cross-linked polyethylene (XLPE) insert. Uncemented fixation. MR Conditional. Sterilized; single-use.

## Regulatory Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

## Predicate Devices

- OR3O Dual Mobility System ([K220959](/device/K220959.md))
- OR3O Dual Mobility System ([K191002](/device/K191002.md))
- OR3O Dual Mobility Liners ([K232667](/device/K232667.md))
- Smith & Nephew Hip Systems - MR Labeling and Design Change ([K211176](/device/K211176.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

October 7, 2024

Smith & Nephew, Inc. Dana Hartlein Senior Regulatory Affairs Specialist 1450 East Brooks Road Memphis, Tennessee 38116

Re: K242375

Trade/Device Name: OR30 Dual Mobility System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH Dated: August 9, 2024 Received: August 9, 2024

Dear Dana Hartlein:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Limin Sun -S

Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

#### Indications for Use

Submission Number (if known)

K242375

Device Name

OR30 Dual Mobility System

Indications for Use (Describe)

The OR30 Dual Mobility System is intended for use in primary and revision total hip arthroplasty in skeletally mature patients.

Indications

· Advanced degeneration of the hip joint as a result of degenerative, post-traumatic, or rheumatoid arthritis.

- · Fracture or avascular necrosis of the femoral head.
- · Failure of previous hip surgery: ioint reconstruction, arthrodesis, hemiarthroplasty,
- surface replacement arthroplasty, or total hip replacement.
- · All forms of osteoarthritis.
- Patients with hips at risk of dislocation.
- · Femoral neck fracture or proximal hip joint fracture.

The OR30 Dual Mobility System is intended for single use only. The modular OR30 Liners and Inserts are to be implanted without bone cement.

Type of Use (Select one or both, as applicable)

× Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K232375

Page 1 of 3

| 510(k) Summary                     |                                                                                       | Prepared on: 2024-10-07 |
|------------------------------------|---------------------------------------------------------------------------------------|-------------------------|
| Contact Details                    |                                                                                       | 21 CFR 807.92(a)(1)     |
| Applicant Name                     | Smith & Nephew, Inc.                                                                  |                         |
| Applicant Address                  | 1450 East Brooks Road Memphis TN 38116 United States                                  |                         |
| Applicant Contact Telephone        | 484-557-4768                                                                          |                         |
| Applicant Contact                  | Ms. Dana Hartlein                                                                     |                         |
| Applicant Contact Email            | dana.hartlein@smith-nephew.com                                                        |                         |
| Device Name                        |                                                                                       | 21 CFR 807.92(a)(2)     |
| Device Trade Name                  | OR3O Dual Mobility System                                                             |                         |
| Common Name                        | Hip Prothesis                                                                         |                         |
| Classification Name                | Hip joint metal/polymer/metal semi-constrained porous-coated<br>uncemented prosthesis |                         |
| Regulation Number                  | 888.3358                                                                              |                         |
| Product Code(s)                    | LPH                                                                                   |                         |
| Legally Marketed Predicate Devices |                                                                                       | 21 CFR 807.92(a)(3)     |
| Predicate #                        | Predicate Trade Name (Primary Predicate is listed first)                              | Product Code            |
| K220959                            | OR3O Dual Mobility System                                                             | LPH                     |
| K191002                            | OR3O Dual Mobility System                                                             | LPH                     |
| K232667                            | OR3O Dual Mobility Liners                                                             | LPH                     |
| K211176                            | Smith & Nephew Hip Systems - MR Labeling and Design Change                            | JDI                     |

{5}------------------------------------------------

#### K232375

Page 2 of 3

### Device Description Summary

The OR3O Hip System was originally cleared via K220959. The OR3O Hip System is a modular dual mobility implant system containing two distinct articulating surfaces. The hip system consists of a metal acetabular shell, diffusion hardened oxdized zirconium (OXINIUM DH) alloy liner, a cross-linked polyethylene (XLPE) Insert and a femoral head.

The purpose of this Taditional 510(k) is to add the MR safety information to the product labels and update the MR information within the package insert for the Snith & Nephew OR3O Dual Mobility System included within the scope of this 510(k). The subject OR30 Dual Mobility System is identical in design, technological characteristics, function, packaging, and sterilization to the commercially available predicate devices.

## Intended Use/Indications for Use

## 21 CFR 807.92(a)(5)

The OR 30 Dual Mobility System is intended for use in primary and revision total hip arthroplasty in skeletally mature patients.

Indications

· Advanced degeneration of the hip joint as a result of degenerative, post-traumatic, or rheumatoid arthritis.

· Fracture or avascular necrosis of the femoral head.

· Failure of previous hip surgery; joint reconstruction, arthrodesis, hemiathroplasty, surface replacement athroplasty, or total hip replacement.

- · All forms of osteoarthritis.
· Patients with hips at risk of dislocation.

· Femoral neck fracture or proximal hip joint fracture.

The OR3O Dual Mobility System is intended for single use only. The modular OR3O Liners and Inserts are to be implanted without bone cement.

## Indications for Use Comparison

The subject OR3O Dual Mobility System has identical indications for use as the commercially available predicate cleared via the premarket notifications listed above.

## Technological Comparison

The subject Smith & Nephew OR3O Dual Mobility System is identical in design, material, technological characteristics, function of the devices, sterile packaging, and sterilization to the commercially available predicate devices cleared in K220959 (08/18/2022), K191002 (10/31/2019), and K232667(10/27/2023). The subject 510(k) is being submitted to add MR safety information to the product labels and update the MR information within the package insert of the OR30 Dual Mobility System previously cleared in K220959 (08/18/2022), K191002 (10/31/2019), and K232667 (10/27/2023).

## 21 CFR 807 92(a)(5)

21 CFR 807.92(a)(6)

#### 21 CFR 807.92(a)(4)

{6}------------------------------------------------

#### K232375

Page 3 of 3

21 CFR 807.92(a)(5)

21 CFR 807.92(a)(6)

## Indications for Use Comparison

The subject OR30 Dual Mobility System has identical indications for use as the commercially available premarket notifications listed above.

## Technological Comparison

The subject Smith & Nephew OR3O Dual Mobility System is identical in design, material, technological characteristics, function of the devices, sterile packaging, and sterilization to the commercially available predicate devices cleared in K.20959 (08/18/2022), K191002 (10/31/2019), and K232667(10/27/2023). The subject 510(k) is being submitted to add MR safety information to the product labels and update the MR information within the package insert of the OR3O Dual Mobility System previously cleared in K220959 (08/18/2022), K191002 (10/31/2019), and K232667(10/27/2023).

#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

Magnetic Resonance Imaging (MRI) compatibility testing was performed and the results of the testing performed were evaluated and used to generate recommended MRI safety labeling for the subject devices. Testing and evaluation were performed in accordance with the standards listed below:

- 1. Magnetically induced displacement force (ASTM F2052)
2. Magnetically induced torque (ASTM F2213)

3. Radiofrequency (RF) induced heating (ASTM F2182)

4. MR image artifact (ASTM F2119)

Based on the MR assessment performed for the subject devices are MR Conditional and substantially equivalent to the predicate MRI 510(k) K211176.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K242375](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K242375)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
