Total hip arthroplasty is intended to replace the damaged hip joint articulation in patients at a high risk of hip dislocation, where there is evidence of sufficient sound bone to seat and support the components.
Device Story
Modular total hip prosthesis component; replaces natural hip articular surface. Device consists of UHMWPE liner locking into porous-coated titanium alloy (Ti-Al6-V4) shell. Provides >180-degree femoral head capture via titanium constraining ring; mechanically constrains femoral head. Available in lateralized neutral or face-changing orientations; compatible with 28, 32, 36mm metal femoral heads and 40, 44mm unipolar/bipolar/modular femoral heads. Used in orthopedic surgery for total hip replacement. Surgeon implants device to stabilize hip joint in high-risk patients. Benefits include reduced risk of dislocation. Current submission updates labeling for MRI compatibility and standardizes symbols/labeling.
Clinical Evidence
Bench testing only. MRI safety evaluation performed per ASTM F2052-21 (displacement force), ASTM F2213-17 (torque), ASTM F2182-19 (RF heating), and ASTM F2119-07 (image artifacts). Data demonstrates conditional safety in MR environment.
Indicated for primary or revision total hip arthroplasty patients at high risk of dislocation due to history of prior dislocation, bone loss, joint/soft tissue laxity, neuromuscular disease, or intraoperative instability, where other constrained options have been considered. For use with PINNACLE Acetabular Cup in cementless application.
Regulatory Classification
Identification
A hip joint metal/polymer constrained cemented or uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device prevents dislocation in more than one anatomic plane and has components that are linked together. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular component made of ultra-high-molecular-weight polyethylene with or without a metal shell, made of alloys, such as cobalt-chromium-molybdenum and titanium alloys. This generic type of device is intended for use with or without bone cement (§ 888.3027).
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Hip Joint Metal/Polymer Constrained Cemented or Uncemented Prosthesis.”
DePuy Hip Portfolio MRI Bundled Traditional 510(k) (K231873)
Related Devices
K071117 — MODIFICATION TO PINNACLE CONSTRAINED ACETABULAR LINERS · DePuy Orthopaedics, Inc. · May 18, 2007
K052079 — PINNACLE CONSTRAINED ACETABULAR LINERS · DePuy Orthopaedics, Inc. · Oct 21, 2005
K043058 — PINNACLE CONSTRAINED ACETABULAR LINERS · Depuy, Inc. · Mar 14, 2005
K971718 — MICRO-SEAL ACETABULAR SYSTEM · Whiteside Biomechanics, Inc. · Jul 23, 1997
K071784 — DEPUY PINNACLE WITH GRIPTON ACETABULAR CUPS · DePuy Orthopaedics, Inc. · Jul 25, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
April 4, 2024
Depuy Ireland UC Jennifer Hill Regulatory Affairs Project Leader Loughbeg, Ringaskiddy County Cork, Ireland
Re: K240639
Trade/Device Name: PINNACLE 114 Constrained Acetabular Liners Regulation Number: 21 CFR 888.3310 Regulation Name: Hip Joint Metal/Polymer Constrained Or Uncemented Prosthesis Regulatory Class: Class II Product Code: KWZ Dated: March 6, 2024 Received: March 6, 2024
Dear Jennifer Hill:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
## Limin Sun - S
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K240639
Device Name
PINNACLETM Constrained Acetabular Liner
Indications for Use (Describe)
The PINNACLETM Constrained Acetabular Liner is indicated for use as a component of a total hip prosthesis in primary or revision patients at high risk of hip dislocation due to a history or prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability and for whom all other options of constrained acetabular components have been considered.
The PINNACLETM Constrained Acetabular Liner is indicated for use with the PINNACLETM Acetabular Cup in cementless application.
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| Submitter Information | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | DePuy Ireland UC |
| Address | Loughbeg<br>Ringaskiddy<br>Co. Cork<br>Ireland |
| Phone number | (+44) 7834974433 |
| Fax number | N/A |
| Establishment Registration Number | 3015516266 |
| Name of contact person | Jennifer Hill |
| Date prepared | 06th March 2024 |
| Name of device | |
| Trade or proprietary name | PINNACLETM Constrained Acetabular Liners |
| Common or usual name | Acetabular Cup Prosthesis |
| Classification name | 21 CFR 888.3310: Hip joint metal/polymer constrained cemented or uncemented prosthesis. |
| Class | II |
| Classification panel | 87 Orthopedics |
| Regulation | Class II – 21 CFR 888.3310 |
| Product Code(s) | KWZ: Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer |
| Legally marketed device(s) to which<br>equivalence is claimed | K043058/K052079/K071117 - PINNACLE™ Constrained Acetabular Liners |
| Reason for 510(k) submission | DePuy Ireland UC has compiled a Special 510(k) Premarket Notification to expand the labeling for<br>the DePuy PINNACLE™ Constrained Acetabular Liners, to provide updated information regarding<br>the MRI compatibility of the devices. A program of MRI Safety Evaluation testing has been carried<br>out with full test data present and reviewed in the FDA cleared submission 'K231873 - DePuy Hip<br>Portfolio MRI Bundled Traditional 510(k)'. Appropriate amends have been made to the product<br>labels and Instructions for Use (IFU), along with the provision of new MR Safety Cards for patients. |
| | At the same time as making these labeling updates in relation to MRI compatibility, DePuy will take<br>the opportunity to standardize language and symbols used across its legacy devices for consistency,<br>including the use of new internationally recognized symbols with corresponding updated symbols<br>glossaries. |
| | Additionally, in order to harmonize with International and EU labeling requirements, the labeling<br>of subject devices which are sold Internationally will be further updated to also include a new<br>Implant Card, Implant Card Guide, and Implant Card Label. These are not US requirements and can<br>be discarded by the end user. The labeling will also be updated to include a material code with<br>corresponding material descriptions. |
| | There is no change to the indications, intended use, safety, fit, form or technological characteristics<br>of the devices. |
| Device description | The PINNACLE™ Constrained Acetabular Liner is part of a modular system designed to replace the<br>natural articular surface of the hip joint in total hip replacement. The liner is manufactured from<br>ultra-high molecular weight polyethylene (UHMWPE), which locks into a porous coated,<br>hemispherical outer shell component manufactured from titanium alloy (Ti-Al6-V4). The liner<br>component articulates with a metal femoral head of an appropriate diameter. |
| | The PINNACLE™ Constrained Acetabular Liner mechanically constrains the femoral head within<br>the inner diameters of the liner by providing greater than 180 degrees femoral head capture<br>combined with a titanium constraining ring which fits over the opening diameter of the liner. The<br>UHMWPE liner is held in the metal shell by means of a titanium locking ring.<br>The PINNACLE™ Constrained Acetabular Liners are UHMWPE acetabular cup liners that are<br>available in a lateralized neutral or lateralized face-changing orientation. The liners have inner<br>diameters compatible with standard metal femoral heads sized 28, 32, and 36mm, as well as<br>larger diameter unipolar, Self-Centering (bipolar) and Modular M femoral heads sized 40 and<br>44mm. The outer diameters (OD) are geometrically the same as other Pinnacle Acetabular Liners,<br>in a 48mm-76mm size range offering. |
| Intended use of the device | Total hip arthroplasty is intended to replace the damaged hip joint articulation in patients at a<br>high risk of hip dislocation, where there is evidence of sufficient sound bone to seat and support<br>the components. |
| Indications for use | The PINNACLE™ Constrained Acetabular Liner is indicated for use as a component of a total hip<br>prosthesis in primary or revision patients at high risk of hip dislocation due to a history of prior<br>dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability<br>and for whom all other options to constrained acetabular components have been considered.<br>The PINNACLE™ Constrained Acetabular Liner is indicated for use with the PINNACLE™<br>Acetabular Cup in cementless application. |
| Substantial Equivalence | DePuy Ireland UC has compiled a Special 510(k) Premarket Notification to expand the labeling for<br>the DePuy PINNACLE™ Constrained Acetabular Liners, to provide updated information regarding<br>the MRI compatibility of the devices. A program of MRI Safety Evaluation testing has been carried<br>out with full test data present and reviewed in the FDA cleared submission 'K231873 - DePuy Hip<br>Portfolio MRI Bundled Traditional 510(k)'. Appropriate amends have been made to the product<br>labels and Instructions for Use (IFU), along with the provision of new MR Safety Cards for patients. |
| At the same time as making these labeling updates in relation to MRI compatibility, DePuy will take the opportunity to standardize language and symbols used across its legacy devices for consistency, including the use of new internationally recognized symbols with corresponding updated symbols glossaries. | |
| Additionally, in order to harmonize with International and EU labeling requirements, the labeling of subject devices which are sold Internationally will be further updated to also include a new Implant Card, Implant Card Guide, and Implant Card Label. These are not US requirements and can be discarded by the end user. The labeling will also be updated to include a material code with corresponding material descriptions. | |
| There is no change to the indications, intended use, safety, fit, form or technological characteristics of the devices. | |
| Non-clinical testing is provided to support the conditional safety of the PINNACLETM Constrained Acetabular Liners in the MR environment, including assessment of:<br>Magnetically Induced Displacement Force (ASTM F2052-21) Magnetically Induced Displacement Torque (ASTM F2213-17) Radio Frequency (RF) Heating (ASTM F2182-19) Image Artifacts (ASTM F2119-07) | |
| The non-clinical performance data demonstrate that the Subject Devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy. | |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.