← Product Code [KWY](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWY) · K200709

# Kyocera Bipolar Hip System (K200709)

_Kyocera Medical Technologies, Inc. · KWY · Apr 17, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWY/K200709

## Device Facts

- **Applicant:** Kyocera Medical Technologies, Inc.
- **Product Code:** [KWY](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWY.md)
- **Decision Date:** Apr 17, 2020
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3390
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis; 2. Rheumatoid arthritis; 3. Correction of functional deformity; 4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement. unmanageable using other techniques Qualifier: The Bipolar Head is for use in conjunction with KMTI femoral heads and femoral stems. In addition: -The KYOCERA Medical Corporation ("KYOCERA") Initia Total Hip System 28mm CoCr and 28mm BIOCERAM AZUL femoral head and Initia femoral stems can be used with the KMTI 28mm ID Bipolar Head. -The KYOCERA Initia Total Hip System femoral stem can be used with the KMTI 22mm CoCr femoral head and KMTI 22mm ID Bipolar Head. The Bipolar Head is for uncemented use only. Bipolar outer heads are not for use with acetabular shells and liners.

## Device Story

The Kyocera Bipolar Hip System is a modular orthopedic prosthesis used in hip arthroplasty. It consists of a bipolar head, femoral head, and femoral stem. The system allows for the integration of specific Kyocera Medical Corporation (KCJ) Initia Total Hip System components (femoral stems and heads) with existing KMTI bipolar head components. The device is intended for uncemented implantation in the proximal femur. It functions as a hemi-hip prosthesis to restore joint function and treat fractures or degenerative disease. It is used by orthopedic surgeons in a clinical/surgical setting. The device provides a mechanical bearing surface for the hip joint, replacing damaged bone structures to alleviate pain and restore mobility.

## Clinical Evidence

No clinical data. The submission relies on bench testing and design control evaluations to demonstrate substantial equivalence.

## Technological Characteristics

Modular hip prosthesis components including CoCr and BIOCERAM AZUL femoral heads and femoral stems. Designed for uncemented use. Complies with Class II Special Controls Guidance Document: Hip Joint Metal/Polymer Constrained Cemented or Uncemented Prosthesis and FDA guidance for Femoral Stem Prostheses.

## Regulatory Identification

A hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis is a two-part device intended to be implanted to replace the head and neck of the femur. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a snap-fit acetabular component made of an alloy, such as cobalt-chromium-molybdenum, and ultra-high molecular weight polyethylene. This generic type of device may be fixed to the bone with bone cement (§ 888.3027) or implanted by impaction.

## Predicate Devices

- Renovis Bipolar Hip System ([K131354](/device/K131354.md))

## Reference Devices

- Renovis Surgical Hip Replacement System ([K112897](/device/K112897.md))
- Kyocera Medical Corporation Japan Initia Total Hip System ([K160895](/device/K160895.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA acronym is in a blue square, and the full name is in blue text. The seal features an emblem with a staff and serpent, symbolizing healing and medicine.

Kyocera Medical Technologies, Inc. % Sharyn Orton Senior Consultant MEDIcept, Inc. 200 Homer Ave Ashland, Massachusetts 01721

April 17, 2020

Re: K200709

Trade/Device Name: Kyocera Bipolar Hip System Regulation Number: 21 CFR 888.3390 Regulation Name: Hip Joint Femoral (Hemi-Hip) Metal/Polymer Cemented Or Uncemented Prosthesis Regulatory Class: Class II Product Code: KWY Dated: March 16, 2020 Received: March 18, 2020

Dear Sharyn Orton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Vesa Vuniqi Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K200709

Device Name Kyocera Bipolar Hip System

Indications for Use (Describe)

- 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
2. Rheumatoid arthritis;

3. Correction of functional deformity;

4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement. unmanageable using other techniques

Qualifier:

The Bipolar Head is for use in conjunction with KMTI femoral heads and femoral stems. In addition:

-The KYOCERA Medical Corporation ("KYOCERA") Initia Total Hip System 28mm

CoCr and 28mm BIOCERAM AZUL femoral head and Initia femoral stems can be used with the KMTI 28mm ID Bipolar Head.

-The KYOCERA Initia Total Hip System femoral stem can be used with the KMTI

22mm CoCr femoral head and KMTI 22mm ID Bipolar Head.

The Bipolar Head is for uncemented use only.

Bipolar outer heads are not for use with acetabular shells and liners.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# KYOCERA

## Special 510(k) Premarket Notification Summary as required by 21 CFR 807.92(a) K200709

| A ) Submitted by: | Kyocera Medical Technologies, Inc.<br>1200 California St. Suite 210<br>Redlands, CA 92374<br>Phone: 909-557-2360<br>Fax: 909-839-6269 |
|-------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | Anthony DeBenedictis<br>Divisional Vice President of Quality Assurance                                                                |
| Consultant:       | Sharyn Orton, Ph.D.<br>MEDIcept, Inc.<br>200 Homer Ave<br>Ashland, MA 01721                                                           |

- B) Classification Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis
Proprietary Name: Kyocera Bipolar Hip System

Device Class: Class II

Regulation: 21 CFR 888.3390

Product Code: KWY

Classification panel: Orthopedic

- C) Predicates: Primary: K131354 Renovis Bipolar Hip System
- D) Date Prepared: March 25, 2020

E) Device Description:

The Renovis Surgical Hip Replacement System (K112897) includes multiple subsystem offerings, including:

- K131354: Renovis Bipolar Hip System ●
Renovis Surgical Technologies is now wholly owned by Kyocera International, Inc. (San

{4}------------------------------------------------

Diego, CA) as Kyocera Medical Technologies, Inc. ("KMTI"). The subject of this Special 510k Premarket Notification is additional KMTI offerings which include the use of components of the K160895 Kyocera Medical Corporation. Japan Initia Total Hip System (now Kyocera Corporation, Japan; "KCJ") that may be used with the KMTI K131354 Kyocera Bipolar Hip System.

F) Intended Use/Indications For Use:

- 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis:
- 2. Rheumatoid arthritis;
- 3. Correction of functional deformity;
- 4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques

### Qualifier:

The Bipolar Head is for use in conjunction with KMTI femoral heads and femoral stems. In addition:

- . The KYOCERA Medical Corporation ("KYOCERA") Initia Total Hip System 28mm CoCr and 28mm BIOCERAM AZUL femoral head and Initia femoral stems can be used with the KMTI 28mm ID Bipolar Head.
- . The KYOCERA Initia Total Hip System femoral stem can be used with the KMTI 22mm CoCr femoral head and KMTI 22mm ID Bipolar Head.

The Bipolar Head is for uncemented use only.

Bipolar outer heads are not for use with acetabular shells and liners.

- G) Substantial Equivalence Comparison and Discussion
There is no change in the Intended Use/Indications for Use; no change in implant materials, manufacturing, packaging, and/or sterilization. Equivalence of KCJ components to KMTI components have been demonstrated.

H) Compliance with Design Controls

All changes were assessed for risk and successfully evaluated under Design Controls.

I) Compliance with Standards or FDA Guidance

This application complies with:

- Class II Special Controls Guidance Document: Hip Joint Metal/Polymer Constrained ● Cemented or Uncemented Prosthesis; Guidance for Industry and FDA, April 29, 2002
- . Non-clinical Information for Femoral Stem Prostheses, September 17, 2007

{5}------------------------------------------------

## Conclusion

The KYOCERA Initia Total Hip System 28mm CoCr and 28mm BIOCERAM AZUL femoral head and Initia femoral stems used with the KMTI 28mm ID Bipolar Head, and the KYOCERA Initia Total Hip System femoral stem used with the KMTI 22mm CoCr femoral head and KMTI 22mm ID Bipolar Head, are substantially equivalent and expected to have equivalent performance when used in place of the corresponding KMTI implants used with the KMTI Bipolar Head.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWY/K200709](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWY/K200709)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).