Fitmore® Hip Stem

K170072 · Zimmer GmbH · KWY · Feb 7, 2017 · Orthopedic

Device Facts

Record IDK170072
Device NameFitmore® Hip Stem
ApplicantZimmer GmbH
Product CodeKWY · Orthopedic
Decision DateFeb 7, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3390
Device ClassClass 2
AttributesTherapeutic

Indications for Use

This femoral stem is for total hip or hemi-hip arthroplasty and is indicated for the following conditions: Non-inflammatory degenerative joint disease (NID), e.g., avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., rheumatoid arthritis. Failed previous surgery where pain, deformity, or dysfunction persists. Revision of previously failed hip arthroplasty. Total hip replacements may be considered for younger patients if any unequivocal indication outweighs the risks associated with the age of the patient and modified demands regarding activity and hip joint loading are assured. This includes severely crippled patients with multiple joint involvement, for whom an immediate need of hip mobility leads to an expectation of significant improvement in the quality of their lives. This stem is for uncemented use only.

Device Story

Fitmore Hip Stem is a modular, curved, uncemented femoral stem for total or hemi-hip arthroplasty. Device features proximal Titanium Vacuum Plasma Sprayed (Ti-VPS) coating and rough-blasted distal surface; available in multiple sizes to accommodate patient morphology. Submission introduces line extension of stainless steel rasps used to prepare femoral canal for implant; rasps sized to match stem templates to ensure femoral stability. Used by orthopedic surgeons in clinical settings. Output is a prepared femoral canal facilitating stem implantation; benefits include restoration of hip mobility and pain relief. No changes made to stem implants; rasps are dimensionally larger variants of predicate rasps.

Clinical Evidence

No clinical data provided. Substantial equivalence supported by non-clinical performance testing, including usability testing and similar device analysis.

Technological Characteristics

Modular femoral stem; uncemented; proximal Ti-VPS coating; rough-blasted distal surface; 12/14 taper. Rasps constructed of stainless steel. Designed for femoral canal preparation and implant stability.

Indications for Use

Indicated for patients requiring total or hemi-hip arthroplasty due to non-inflammatory degenerative joint disease (e.g., avascular necrosis, osteoarthritis), inflammatory joint disease (e.g., rheumatoid arthritis), failed previous surgery, or revision of failed hip arthroplasty. Suitable for younger patients if clinical indications outweigh age-related risks and activity/loading demands are managed. For uncemented use only.

Regulatory Classification

Identification

A hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis is a two-part device intended to be implanted to replace the head and neck of the femur. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a snap-fit acetabular component made of an alloy, such as cobalt-chromium-molybdenum, and ultra-high molecular weight polyethylene. This generic type of device may be fixed to the bone with bone cement (§ 888.3027) or implanted by impaction.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ February 7, 2017 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Zimmer GmbH Thomas Lincoln Senior Regulatory Affairs Specialist Sulzerallee 8. P.O. Box 8404 Winterthur. Switzerland Re: K170072 Trade/Device Name: Fitmore Hip Stem Regulation Number: 21 CFR 888.3390 Regulation Name: Hip Joint Femoral (Hemi-Hip) Metal/Polymer Cemented Or Uncemented Prosthesis Regulatory Class: Class II Product Code: KWY, KWZ, JDI, LZO, LWJ, KWL Dated: January 3. 2017 Received: January 9, 2017 Dear Thomas Lincoln: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K170072 Not yet known (submitted as a Special 510(k) to K071723) Device Name Fitmore® Hip Stem Indications for Use (Describe) This femoral stem is for total hip or hemi-hip arthroplasty and is indicated for the following conditions: • Non-inflammatory degenerative joint disease (NID), e.g., avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., rheumatoid arthritis. · Failed previous surgery where pain, deformity, or dysfunction persists. - · Revision of previously failed hip arthroplasty. · Total hip replacements may be considered for younger patients if any unequivocal indication outweighs the risks associated with the age of the patient and modified demands regarding activity and hip joint loading are assured. This includes severely crippled patients with multiple joint involvement, for whom an immediate need of hip mobility leads to an expectation of significant improvement in the quality of their lives. This stem is for uncemented use only. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." EF PSC Publishing Services (301) 443-6740 {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Zimmer Biomet. The logo features a large, stylized blue letter "Z" enclosed in a blue circle. Below the circle, the word "zimmer" is written in a lowercase, sans-serif blue font. ## 510(k) Summary | Sponsor: | Zimmer GmbH<br>Sulzerallee 8, P.O. Box<br>8404 Winterthur, Switzerland | |--------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Thomas Lincoln<br>Senior Specialist, Regulatory Affairs<br>Telephone: +41 5885 48 257<br>Fax: + 41 5224 48 658 | | Date: | 03rd January, 2017 | | Subject Device: | Trade Name: Fitmore® Hip Stem<br>Common Name: Hip Prosthesis | | Classification Product Code /<br>Device Classification Name: | KWY - Prosthesis, hip, hemi-, femoral,<br>metal/polymer, cemented or uncemented<br>KWZ - Prosthesis, hip, constrained, cemented or<br>uncemented, metal/polymer<br>JDI - Prosthesis, hip, semiconstrained,<br>metal/polymer, cemented<br>LZO - Prosthesis, hip, semiconstrained,<br>metal/ceramic/polymer, cemented or nonporous,<br>uncemented<br>LWJ - Prosthesis, hip, semiconstrained, metal<br>polymer, uncemented<br>KWL - Prosthesis, hip, hemi-, femoral, metal | | Regulation Number / Description: | 21 CFR § 888.3390 Hip joint femoral (hemi-hip)<br>metal/polymer cemented or uncemented prosthesis<br>21 CFR § 888.3310 Hip joint metal/polymer<br>constrained cemented or uncemented prosthesis<br>21 CFR § 888.3350 Hip joint metal/polymer semi-<br>constrained cemented prosthesis<br>21 CFR § 888.3353 Hip joint metal/ceramic/polymer<br>semi-constrained cemented or nonporous uncemented<br>prosthesis.<br>21 CFR § 888.3360 Hip joint femoral (hemi-hip)<br>metallic cemented or uncemented prosthesis | | | 21 CFR § 888.3360 Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis | | Predicate Device: | Fitmore Hip Stem, manufactured by Zimmer GmbH, K071723, cleared March 7, 2008 (as Zimmer Porolock MIS Stem). | | Device Description: | The Fitmore Hip Stem is a modular femoral stem intended for total or hemi-hip arthroplasty. The curved, uncemented stem and is coated proximally with Titanium Vacuum Plasma Sprayed (Ti-VPS) and rough-blasted distally. The 12/14 femoral stem is available in multiple sizes in order to address different patient morphologies.<br><br>No changes are being made to the Fitmore Hip Stems themselves, a line extension to the rasps provided to facilitate implantation of the Fitmore Hip Stem is being proposed in this submission.<br><br>The rasps that are subject of this submission are intended to prepare the femoral canal for final implantation of the stem. They are made of Stainless Steel and are sized relative to the intended Fitmore Stem implant beginning with the starter rasp and becoming progressively larger in order that the femoral canal created by the rasps is sized appropriately to the pre-operative templated size of stem with the created canal also providing the required femoral stability | | Intended Use: | This femoral stem is for total hip or hemi-hip arthroplasty and is indicated for the following conditions:<br>Non-inflammatory degenerative joint disease (NIDJD), e.g., avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., rheumatoid arthritis. Failed previous surgery where pain, deformity, or dysfunction persists | {4}------------------------------------------------ - Revision of previously failed hip arthroplasty. . {5}------------------------------------------------ Total hip replacements may be considered for ● younger patients if any unequivocal indication outweighs the risks associated with the age of the patient and modified demands regarding activity and hip joint loading are assured. This includes severely crippled patients with multiple joint involvement, for whom an immediate need of hip mobility leads to an expectation of significant improvement in the quality of their lives. This stem is for uncemented use only. The Fitmore Hip Stem implants are not modified as Comparison to Predicate Device: compared to their predicates. The Fitmore Hip Stem implants are identical in intended use, materials, sterility, and performance characteristics to the predicate devices and remain unchanged. Instead, an additional rasp variant is introduced; these rasps are dimensionally larger than the currently available rasps and in all other aspects (intended use, materials, sterility, and performance characteristics) the proposed rasps are identical to the predicate rasps. Performance Data (Nonclinical Non-Clinical Performance and Conclusions: and/or Clinical): The results of non-clinical performance testing and analyses demonstrate that the devices are safe and effective and substantially equivalent to the predicate devices. Performance analyses included: 1. Usability Testing 2. Similar Device Analysis > Clinical Performance and Conclusions: Clinical data and conclusions were not needed for this device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...