RENOVIS BIPOLAR HIP SYSTEM

K131354 · Renovis Surgical Technologies, LLC · KWY · Sep 19, 2013 · Orthopedic

Device Facts

Record IDK131354
Device NameRENOVIS BIPOLAR HIP SYSTEM
ApplicantRenovis Surgical Technologies, LLC
Product CodeKWY · Orthopedic
Decision DateSep 19, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3390
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. Rheumatoid arthritis. Correction of functional deformity. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other technique

Device Story

Renovis Bipolar Hip System is a modular hemi-arthroplasty implant replacing the natural femoral head. System consists of a highly polished CoCr alloy outer shell and a UHMWPE insert (liner and locking ring). Device functions as a bipolar construct: outer CoCr shell articulates with patient's natural acetabular cartilage; inner UHMWPE liner articulates with a modular femoral head attached to a femoral stem. Implanted by orthopedic surgeons during hip replacement procedures. Provides a bearing surface to restore joint function and mobility in patients with degenerative or traumatic hip conditions. System includes full instrumentation for surgical implantation.

Clinical Evidence

Bench testing only. Performance testing conducted in compliance with ASTM F1820-13 (Standard Test Method for Determining the Forces for Disassembly of Modular Acetabular Devices) with specific deviations for bipolar head testing.

Technological Characteristics

Materials: CoCr alloy (ISO 5832-4, ASTM F1537) and UHMWPE (ISO 5834-2). Design: Modular bipolar head with inner/outer articulation surfaces. Sterilization: Gamma radiation. Standards: ASTM F983-86, ASTM F565-04, ASTM F2033-12, ISO 11137-2, ISO 17665-1.

Indications for Use

Indicated for patients with non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, or proximal femur fractures (non-union, femoral neck, trochanteric) requiring hemi-arthroplasty.

Regulatory Classification

Identification

A hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis is a two-part device intended to be implanted to replace the head and neck of the femur. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a snap-fit acetabular component made of an alloy, such as cobalt-chromium-molybdenum, and ultra-high molecular weight polyethylene. This generic type of device may be fixed to the bone with bone cement (§ 888.3027) or implanted by impaction.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### Traditional 510(k) Summary as required by 21 CFR 807.92(a) K131354 | A ) Submitted by: | Renovis Surgical Technologies<br>1901 W. Lugonia Ave, Ste 340<br>Redlands, CA 92374<br>Phone: 909-557-2360<br>Fax: 909-307-8571 | | | SEP 19 2013 | |---------------------------------|---------------------------------------------------------------------------------------------------------------------------------|--|--|-------------| | Official Contact: | Anthony DeBenedictis<br>Vice President of Quality Assurance | | | | | Consultant: | Sharyn Orton, Ph.D.<br>MEDIcept, Inc.<br>200 Homer Ave<br>Ashland, MA 01721 | | | | | Prepared: | August 13, 2013 | | | | | B) Classification Name: | Prosthesis. Hip, Hemi-, Femoral, Metal/Polymer, Cemented<br>or Uncemented | | | | | Common Name: | Bipolar Hip System | | | | | Proprietary Name: | Renovis Bipolar Hip System | | | | | Device Class: | Class II | | | | | Regulation<br>and Product Code: | 21 CFR 888.3390<br>KWY | | | | | Classification panel: | Orthopedic | | | | | C) Predicates: | Aesculap Bipolar Acetabular Cup, K060707, KWY<br>StelKast Bipolar Hip System, K 972961, KWY, JDI | | | | D) Device Description: The Renovis Bipolar Hip System includes a bipolar head and 22 mm femoral heads. The Renovis Bipolar heads are offered in a range of inner and outer dimensions from 39 to 65 mm to fit either 22 or 28 mm femoral heads. Renovis already has FDA clearance for a 28 mm femoral head (K112897). Renovis Bipolar Hip System {1}------------------------------------------------ The bioolar head is a sub-component of a hip replacement construct where it assumes the function of the natural femoral head and is combined with a femoral head and hip stem. It consists of a highly polished CoCr alloy outer shell and a polyethylene insert. The ultra-highmolecular weight polyethylene (UHMWPE) insert is comprised of an inner liner and a locking ring. The bipolar head has two articulation surfaces: the outer CoCr head articulates with the patient's natural acetabular cartilage and the inner UHMWPE liner articulates with the femoral head assembled to the femoral stem. The modular CoCr alloy femoral head, is held in place within the bipolar head by a retaining ring that is pre-assembled into the polyethylene insert. The bipolar heads and femoral heads are gamma sterilized. The system is offered with the full instrumentation required for the procedure. E) Intended Use/Indications For Use: - Non-inflammatory degenerative joint disease including osteoarthritis and avascular . necrosis - Rheumatoid arthritis . - Correction of functional deformity . - Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal . femur with head involvement, unmanageable using other technique - F) Substantial Equivalence Comparison and Discussion The Renovis Bipolar Hip System application includes a bipolar head (cup) and 22 femoral head. The Bipolar Hip System: - . Is the same technology (material and design) as the predicate devices - Has the same or similar Indications for Use as the predicate device . - Has the same bipolar head inner and outer dimensional sizes as the predicate devices ● - . Uses a slightly different size femoral head than one of the predicate devices, but the same as the other predicate device · #### Discussion The Renovis Bipolar Hip System is substantially equivalent to the predicate devices, and any slight differences do not raise new issues of safety or effectiveness. {2}------------------------------------------------ G) Performance Testing The Renovis Bipolar Hip System has successfully undergone ASTM performance testing in compliance with ASTM F1820-13 Standard Test Method for Determining the Forces for Disassembly of Modular Acetabular Devices, but with deviations specific for testing of a bipolar head. H) Compliance with Standards The Renovis Bipolar Hip System complies with the following Standards: - . ISO 5832-4:1996 Metallic materials -- Part 4: Cobalt-chromium-molybdenum casting alloy - . ISO 5834-2:2011 Implants for surgery -- Ultra-high-molecular-weight polyethylene --Part 2: Moulded forms - ASTM F1537-8 Standard Specification for Wrought Cobalt-28Chromium-6Molybdenum � Alloys for Surgical Implants (UNS R31537, UNS R31538, and UNS R31539) - ASTM F983-86 (Reapproved 2009) Standard Practice for Permanent Marking of . Orthopaedic Implant Components - ASTM F 565-04 (Reapproved 2009)el Standard Practice for Care and Handling of . Orthopedic Implants and Instruments - ASTM F2033-12 Standard Specification for Total Hip Joint Prosthesis and Hip . Endoprosthesis Bearing Surfaces Made of Metallic, Ceramic, and Polymeric Materials - o ASTM F1820-13 Standard Test Method for Determining the Forces for Disassembly of Modular Acetabular Devices with deviations - . ISO 11137-2:2006 Sterilization of health care products -- Radiation -- Part 2: Establishing the sterilization dose - . ISO 17665-1:2006 Sterilization of health care products - Moist heat -- Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices ## Conclusion The results of the Renovis Bipolar Hip System performance testing indicate that the device performs as expected. The Renovis Bipolar Hip System has the same Intended Use/Indications for use as the predicate devices, and any slight differences from the predicate devices do not raise new or different issues of safety or effectiveness. Therefore, the Renovis Bipolar Hip System is substantially equivalent to the predicate devices. {3}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 19, 2013 Renovis Surgical Technologies, LLC % Sharyn Orton, Ph.D. MEDIcept, Incoporated 200 Homer Avenue Ashland, Massachusetts 01721 Re: K131354 Trade/Device Name: Renovis Bipolar Hip System Regulation Number: 21 CFR 888.3390 Regulation Name: Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis Regulatory Class: Class II Product Code: KWY Dated: August 14, 2013 Received: August 19, 2013 Dear Dr. Orton: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 - Sharyn Orton. Ph.D. forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/McdicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Ronald P. Jean -S for - Mark N: Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### Indications for Use Form 510(k) Number (if known): K131354 Device Name: Renovis Bipolar Hip System Indications for Use: - Non-inflammatory degenerative joint disease including osteoarthritis and avascular . necrosis - Rheumatoid arthritis . - Correction of functional deformity . - Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal . femur with head involvement, unmanageable using other technique AND/OR Prescription Use_ X (21 CFR 801 Subpart D) Over-the-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NECESSARY) · Concurrence of CDRH, Office of Device Evaluation (ODE) # Elizabeth L. Frank -S Renovis Bipolar Hip System Traditional 510(k) Premarket Notification
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...