The Equinoxe Humeral Reconstruction Prosthesis System is inteni or total shoulder arthroplasty where proximal humeral resection is deemed necessary in cases of: - Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis - · Rheumatoid arthritis - · Revision where other devices or treatments have failed - · Correction of functional deformity - Treatment of acute or chronic fracture with humeral head involvement, which are unmanageable using other treatment - · Traumatic arthritis - · Oncology applications including bone loss due to tumor resection. - · Significant humeral resection which are unmanageable using other treatment methods The Equinoxe Humeral Reconstruction Prosthesis System can be used in either primary or revision arthroplasty procedures. The Equinoxe Humeral Reconstruction Prosthesis System is indicated for proximal humeral replacement in conjunction with reverse shoulder arthroplasty in which significant resection of the proximal humerus is necessary, the rotator cuff is irreplaceable and grossly deficient, and a functional deltoid muscle is present. In the USA, the Equinoxe Humeral Reconstruction Prosthesis System is not indicated for use with the Equinoxe Reverse Shoulder System components in oncology applications. The Equinoxe Humeral Reconstruction Prosthesis Distal Stems are for cemented use only, while the HA coated Equinoxe Humeral Reconstruction Prosthesis Distal Stem Collars are only for uncemented, press-fit use.
Device Story
Humeral reconstruction prosthesis system; provides proximal humeral replacement for primary or revision shoulder arthroplasty. Device components include extra-small proximal bodies; used in conjunction with distal stems and collars. Implanted by orthopedic surgeons in clinical/hospital settings. Provides structural support for bone loss or resection; restores joint function. Modification of predicate device involves decreased tuberosity thickness. Device components are cemented (distal stems) or uncemented/press-fit (HA-coated collars).
Clinical Evidence
Bench testing only. Engineering analysis performed on mechanical properties including taper analysis, suture abrasion, axial pull-off, fatigue testing, and torsion testing. Pyrogen testing conducted per USP <161>, USP <85>, and ANSI/AAMI ST72.
Technological Characteristics
Shoulder joint prosthesis; metal/polymer construction. Includes extra-small proximal bodies, distal stems, and HA-coated collars. Distal stems are cemented; collars are uncemented/press-fit. Sterilization per FDA guidance. Mechanical testing performed to verify structural integrity following modification of tuberosity thickness.
Indications for Use
Indicated for patients requiring proximal humeral resection due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, revision arthroplasty, functional deformity, acute/chronic fracture with humeral head involvement, traumatic arthritis, or oncology (tumor resection). Also indicated for proximal humeral replacement with reverse shoulder arthroplasty when the rotator cuff is deficient and the deltoid is functional. Contraindicated for use with Equinoxe Reverse Shoulder System components in oncology applications in the USA.
Regulatory Classification
Identification
A shoulder joint metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-Aluminum 4-Vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-Chromium-Molybdenum Casting Alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-Chromium-Molybdenum Alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal Coatings,”
(vii) F 1378-97 “Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
K180089 — Medacta Shoulder System Short Humeral Diaphysis · Medacta International S.A. · Apr 5, 2018
K020045 — 3-PIECE PROXIMAL HUMERAL REPLACEMENT SYSTEM · Biomet Orthopedics, Inc. · Jan 30, 2002
K212710 — GLOBAL UNITE Platform Shoulder System · Depuy Ireland UC · Sep 15, 2021
K001706 — WMT MODULAR SHOULDER SYSTEM · Wrightmedicaltechnologyinc · Aug 16, 2000
K050761 — ZIMMER TRABECULAR METAL HUMERAL STEM, SIZES 6MM AND 8MM · Zimmer, Inc. · Apr 11, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 27, 2017
Exactech Inc Thomas Mcnamara Regulatory Affairs Specialist 2320 N.W. 66th Court Gainesville, Florida 32653
Re: K162903
Trade/Device Name: Exactech Equinoxe Humeral Reconstruction Prosthesis Extra-small Proximal Bodies Regulation Number: 21 CFR 888.3650 Regulation Name: Shoulder Joint Metal/Polymer Non-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: KWT, KWS, PHX, HSD Dated: January 26, 2017 Received: January 30, 2017
Dear Thomas McNamara:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
### 510(k) Number (if known)
K162903
#### Device Name
Exactech® Equinoxe® Humeral Reconstruction Prosthesis Extra-Small Proximal Bodies
Indications for Use (Describe)
The Equinoxe Humeral Reconstruction Prosthesis System is inteni or total shoulder arthroplasty where proximal humeral resection is deemed necessary in cases of:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis
- · Rheumatoid arthritis
- · Revision where other devices or treatments have failed
- · Correction of functional deformity
- Treatment of acute or chronic fracture with humeral head involvement, which are unmanageable using other treatment
- · Traumatic arthritis
- · Oncology applications including bone loss due to tumor resection.
- · Significant humeral resection which are unmanageable using other treatment methods
The Equinoxe Humeral Reconstruction Prosthesis System can be used in either primary or revision arthroplasty procedures.
The Equinoxe Humeral Reconstruction Prosthesis System is indicated for proximal humeral replacement in conjunction with reverse shoulder arthroplasty in which significant resection of the proximal humerus is necessary, the rotator cuff is irreplaceable and grossly deficient, and a functional deltoid muscle is present.
In the USA, the Equinoxe Humeral Reconstruction Prosthesis System is not indicated for use with the Equinoxe Reverse Shoulder System components in oncology applications.
The Equinoxe Humeral Reconstruction Prosthesis Distal Stems are for cemented use only, while the HA coated Equinoxe Humeral Reconstruction Prosthesis Distal Stem Collars are only for uncemented, press-fit use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Sponsor: | Exactech®, Inc<br>2320 NW 66th Court<br>Gainesville FL, 32653 |
|-------------------|-----------------------------------------------------------------------------------------------------------|
| | Phone: (352) 377-1140 Fax: (352) 378-2617 |
| | FDA Establishment Number 1038671 |
| Date: | January 26, 2017 |
| Contact Person: | Thomas McNamara, RAC<br>Regulatory Affairs Specialist<br>Telephone: (352) 327-1140<br>Fax: (352) 378-2617 |
| Proprietary Name: | Exactech® Equinoxe® Humeral Reconstruction Prosthesis Extra-Small<br>Proximal Bodies |
| Common Name: | Humeral Stem |
## Classification Name:
Shoulder Joint Metal/Polymer Non-Constrained Cemented Prosthesis, (21 CFR Section 888.3650. Class II, Product Code KWT). Shoulder Joint Metal/Polymer Semi-Constrained Cemented Prosthesis, (21 CFR Section 888.3660, Class II Product Code KWS), Shoulder Joint Metal/Polymer Semi-Constrained Cemented Prosthesis, (21 CFR Section 888.3660, Class II Product Code PHX) Shoulder Joint Humeral (Hemi-Shoulder) Metallic Uncemented Prosthesis, (21 CFR Section 888.3690, Class II, Product Code HSD)
## Legally Marketed Device to Which Substantial Equivalence Is Claimed:
| Name | Manufacturer | 510(k) Number |
|-----------------------------------------------|---------------|---------------|
| Equinoxe Humeral Reconstruction<br>Prosthesis | Exactech, Inc | K143659 |
## Indication for Use:
The Equinoxe Humeral Reconstruction Prosthesis System is intended for use in hemi or total shoulder arthroplasty where proximal humeral resection is deemed necessary in cases of:
- . Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis
- . Rheumatoid arthritis
- . Revision where other devices or treatments have failed
- . Correction of functional deformity
- . Treatment of acute or chronic fracture with humeral head involvement, which are unmanageable using other treatment
- Traumatic arthritis
{4}------------------------------------------------
# Exactech® Equinoxe® Humeral Reconstruction Prosthesis Extra-Small Proximal Bodies Traditional 510(k) - 510(k) Summary
- Oncology applications including bone loss due to tumor resection. .
- . Significant humeral resection which are unmanageable using other treatment methods
The Equinoxe Humeral Reconstruction Prosthesis System can be used in either primary or revision arthroplasty procedures.
The Equinoxe Humeral Reconstruction Prosthesis System is indicated for proximal humeral replacement in conjunction with reverse shoulder arthroplasty in which significant resection of the proximal humerus is necessary, the rotator cuff is irreplaceable and grossly deficient, and a functional deltoid muscle is present.
In the USA, the Equinoxe Humeral Reconstruction Prosthesis System is not indicated for use with the Equinoxe Reverse Shoulder System components in oncology applications.
The Equinoxe Humeral Reconstruction Prosthesis Distal Stems are for cemented use only, while the HA coated Equinoxe Humeral Reconstruction Prosthesis Distal Stem Collars are only for uncemented, press-fit use.
# Device Description
The proposed Exactech Equinoxe Humeral Reconstruction Prosthesis Extra-Small Proximal Bodies represent a modification of the predicate Exactech Equinoxe Humeral Reconstruction Prosthesis Proximal Bodies cleared in K143659. Both the predicate and proposed devices have the same intended use, general design features and basic fundamental scientific technology. The only difference between the predicate and the proposed device is a decreased tuberosity thickness.
# Testing:
An engineering analysis was conducted to evaluate the mechanical properties of the proposed Equinoxe Humeral Reconstruction Prosthesis Extra-Short Proximal Bodies, including:
- Taper Analysis ●
- Suture Abrasion
- Axial Pull-off ●
- Fatigue Testing ●
- Torsion Testing
Pyrogen testing was conducted in accordance with USP <161>, USP <85>, and ANSI/AAMI ST72 to meet recommended limits per FDA's Guidance Document Submission and Review of Sterility Information in Premarket (510(k)) Submission for Devices Labeled as Sterile.
# Substantial Equivalence Conclusion:
Results of engineering studies referenced in this 510(k) submission demonstrate the proposed Exactech Equinoxe Humeral Reconstruction Prosthesis Extra-Small Proximal Bodies are substantially equivalent to the cleared Exactech Humeral Reconstruction Proximal Bodies.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.