← Product Code [KWS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWS) · K252416

# AETOS Shoulder System Meta Humeral Prosthesis Size 0 (K252416)

_Smith & Nephew, Inc. · KWS · Oct 29, 2025 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWS/K252416

## Device Facts

- **Applicant:** Smith & Nephew, Inc.
- **Product Code:** [KWS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWS.md)
- **Decision Date:** Oct 29, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3660
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The AETOS Shoulder System Meta Humeral Prosthesis Size 0 subject to this submission is intended for press-fit fixation as of an implant construct in anatomic and/or reverse total shoulder arthroplasty for the treatment of skeletally mature individuals with degenerative diseases of the glenohumeral joint.

## Device Story

AETOS Shoulder System Meta Humeral Prosthesis Size 0 is a humeral stem implant for anatomic or reverse total shoulder arthroplasty; provides press-fit fixation for joint replacement. Device used by orthopedic surgeons in clinical settings to treat degenerative shoulder conditions; restores joint mobility/stability and relieves pain. Implant serves as mechanical replacement for diseased humeral head/joint surfaces. System allows conversion from anatomic to reverse construct without stem removal if primary stem is well-fixed. Benefits include pain relief and functional restoration for patients with specific rotator cuff or deltoid status. Single-use device.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via non-clinical performance testing, specifically construct fatigue and corrosion resistance assessments.

## Technological Characteristics

Humeral prosthesis for press-fit (uncemented) fixation. Materials, sterilization, and packaging are identical to the predicate AETOS Shoulder System. Operates as a semi-constrained or non-constrained shoulder joint prosthesis depending on configuration (anatomic vs. reverse).

## Regulatory Identification

A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

## Predicate Devices

- AETOS Shoulder System ([K220847](/device/K220847.md))
- AETOS Shoulder System ([K230572](/device/K230572.md))
- AETOS Shoulder System Stemless Humeral Prosthesis ([K240716](/device/K240716.md))

## Reference Devices

- Titan Reverse Shoulder System ([K130050](/device/K130050.md))
- Titan Reverse Shoulder System ([K173717](/device/K173717.md))
- Titan Modular Total Shoulder System ([K100448](/device/K100448.md))
- Titan Modular Total Shoulder System ([K142413](/device/K142413.md))
- Tornier SAS Aequalis Ascend Flex Shoulder System ([K122698](/device/K122698.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

October 29, 2025

Smith &amp; Nephew, Inc.
Cassidy Whipple
Senior Regulatory Affairs Specialist
7135 Goodlett Farms Parkway
Cordova, Tennessee 38016

Re: K252416
Trade/Device Name: AETOS Shoulder System Meta Humeral Prosthesis Size 0
Regulation Number: 21 CFR 888.3660
Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis
Regulatory Class: Class II
Product Code: KWS, HSD, KWT, PHX, PKC
Dated: July 31, 2025
Received: August 1, 2025

Dear Cassidy Whipple:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}

K252416 - Cassidy Whipple
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K252416 - Cassidy Whipple
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Farzana Sharmin-S
Digitally signed by Farzana Sharmin-S
Date: 2025.10.29 23:03:19 -04'00'

Farzana Sharmin, PhD
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

AETOS Shoulder System Meta Humeral Prosthesis Sz 0
Page 1 of 2

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K252416 | ?  |
|  Please provide the device trade name(s). |  | ?  |
|  AETOS Shoulder System Meta Humeral Prosthesis Size 0  |   |   |
|  Please provide your Indications for Use below. |  | ?  |
|  In Anatomic:

The AETOS Shoulder System (when used with the AETOS Humeral Meta Stems) is to be used only in
patients with an intact or reconstructable rotator cuff, where it is intended to provide increased mobility and
stability and to relieve pain.

The AETOS humeral stems (AETOS Humeral Meta Stems) and head may be used by themselves, as a
hemiarthroplasty, if the natural glenoid provides a sufficient bearing surface, or in conjunction with the
glenoid, as a total replacement.

When used with the AETOS Humeral Meta Stems, The AETOS Shoulder System is indicated for use as a
replacement of shoulder joints disabled by:
•Rheumatoid arthritis
•Non-inflammatory degenerative joint disease
•Correction of functional deformity
•Fractures of the humeral head
•Traumatic arthritis
•Revision of other devices if sufficient bone stock remains

The AETOS Shoulder System (when used with AETOS Stemless Humeral Prosthesis) is to be used only in
patients with an intact or reconstructable rotator cuff.

When used with AETOS Stemless Humeral Prosthesis, the AETOS Shoulder System is indicated for
anatomic total shoulder replacement of shoulder joints disabled by:
•Non-inflammatory degenerative joint disease
•Traumatic arthritis
•Revision of other devices if sufficient bone stock remains

The coated humeral components are intended for uncemented use. The glenoid component is intended for
cemented use only.

In Reverse:

The AETOS Shoulder System, when used with AETOS Humeral Meta Stems, is indicated for use as a
replacement of shoulder joints for patients with a functional deltoid muscle and with massive and non-
repairable rotator cuff-tear with pain disabled by:
•Rheumatoid arthritis  |   |   |

{4}

- Non-inflammatory degenerative joint disease
- Correction of functional deformity
- Fractures of the humeral head
- Traumatic arthritis
- Revision of devices if sufficient bone stock remains

The humeral liner component is indicated for use in the AETOS Shoulder System as a primary reverse total shoulder replacement and for use when converting an anatomic AETOS Shoulder System into a reverse shoulder construct. This facilitates the conversion without the removal of the humeral stem during revision surgery for patients with a functional deltoid muscle. The component is permitted to be used in the conversion from anatomic to reverse if the humeral stem is well fixed, the patient has a functional deltoid muscle; the arthroplasty is associated with a massive and non-repairable rotator cuff tear.

The coated humeral stems are indicated for uncemented use. The coated glenoid base plate is intended for cementless application with the addition of screws for fixation.

Note: All implant components are single use.

Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

AETOS Shoulder System Meta Humeral Prosthesis Sz 0
Page 2 of 2

{5}

K252416
Page 1 of 3

510(k) Summary
AETOS Shoulder System Meta Humeral Prosthesis Size 0
Traditional 510(k)

|  Sponsor | Smith & Nephew, Inc.
Orthopedic Division
7135 Goodlett Farms Parkway
Cordova, Tennessee 38016  |
| --- | --- |
|  Establishment Number | 3008744062  |
|  Point of Contact | Cassidy Whipple
Senior Regulatory Specialist
512-466-1130  |
|  Date | October 29, 2025  |
|  Trade Name | AETOS Shoulder System Meta Humeral Prosthesis Size 0  |
|  Common Name | Shoulder Prosthesis  |
|  Product Code | KWS, HSD, KWT, PHX, PKC  |
|  Regulation | 1. Shoulder joint metal/polymer semi-constrained cemented prosthesis (21 CFR 888.3660)
2. Shoulder Joint humeral (hemi-shoulder) metallic uncemented (21 CFR 888.3690)
3. Shoulder joint metal/polymer non-constrained cemented prosthesis (21 CFR 888.3650)  |
|  Classification | Class II  |
|  Predicate Devices | AETOS Shoulder System: K220847 (primary), K230572
AETOS Shoulder System Stemless Humeral Prosthesis: K240716  |
|  Reference Devices | Titan Reverse Shoulder System: K130050, K173717
Titan Modular Total Shoulder System: K100448, K142413
Tornier SAS Aequalis Ascend Flex Shoulder System: K122698  |
|  Classification Panel | Orthopedic  |
|  Device Description / Intended Use | The AETOS Shoulder System Meta Humeral Prosthesis Size 0 subject to this submission is intended for press-fit fixation as of an implant construct in anatomic and/or reverse total shoulder arthroplasty for the treatment of skeletally mature individuals with degenerative diseases of the glenohumeral joint.  |

{6}

K252416
Page 2 of 3

|  Indications for Use (System) | In Anatomic: The AETOS Shoulder System (when used with the AETOS Humeral Meta Stems) is to be used only in patients with an intact or reconstructable rotator cuff, where it is intended to provide increased mobility and stability and to relieve pain. The AETOS humeral stems (AETOS Humeral Meta Stems) and head may be used by themselves, as a hemiarthroplasty, if the natural glenoid provides a sufficient bearing surface, or in conjunction with the glenoid, as a total replacement. When used with the AETOS Humeral Meta Stems, The AETOS Shoulder System is indicated for use as a replacement of shoulder joints disabled by: • Rheumatoid arthritis • Non-inflammatory degenerative joint disease • Correction of functional deformity • Fractures of the humeral head • Traumatic arthritis • Revision of other devices if sufficient bone stock remains The AETOS Shoulder System (when used with AETOS Stemless Humeral Prosthesis) is to be used only in patients with an intact or reconstructable rotator cuff. When used with AETOS Stemless Humeral Prosthesis, the AETOS Shoulder System is indicated for anatomic total shoulder replacement of shoulder joints disabled by: • Non-inflammatory degenerative joint disease • Traumatic arthritis • Revision of other devices if sufficient bone stock remains The coated humeral components are intended for uncemented use. The glenoid component is intended for cemented use only. In Reverse: The AETOS Shoulder System, when used with AETOS Humeral Meta Stems, is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and non-repairable rotator cuff-tear with pain disabled by: • Rheumatoid arthritis • Non-inflammatory degenerative joint disease • Correction of functional deformity • Fractures of the humeral head • Traumatic arthritis  |
| --- | --- |

{7}

K252416
Page 3 of 3

|   | • Revision of devices if sufficient bone stock remains
The humeral liner component is indicated for use in the AETOS Shoulder System as a primary reverse total shoulder replacement and for use when converting an anatomic AETOS Shoulder System into a reverse shoulder construct. This facilitates the conversion without the removal of the humeral stem during revision surgery for patients with a functional deltoid muscle. The component is permitted to be used in the conversion from anatomic to reverse if the humeral stem is well fixed, the patient has a functional deltoid muscle; the arthroplasty is associated with a massive and non-repairable rotator cuff tear.

The coated humeral stems are indicated for uncemented use. The coated glenoid base plate is intended for cementless application with the addition of screws for fixation.

Note: All implant components are single use.  |
| --- | --- |
|  Comparison to Predicate | The proposed device is an addition of a smaller size, Size 0, of the humeral stem to the AETOS Shoulder System portfolio.  |
|  Nonclinical Performance Data | Comparison to the predicate demonstrating the additional size has no effect on the following non-clinical performance tests.
• Construct fatigue and corrosion resistance assessment  |
|  Clinical Performance Data | Clinical performance data were not necessary to demonstrate substantial equivalence of the subject device.  |
|  Conclusion | Substantial equivalence of the AETOS Shoulder System Meta Humeral Prosthesis Size 0 to cited predicate can be demonstrated based on the following:
• The subject and predicate device have the same intended use and the same Indications for Use.
• The subject and predicate device share the same functional and technological characteristics via the same operational principles.
• The subject and predicate device are made from the same materials and packaged and sterilized using the same methods.

Performance evaluations support that differences between subject and predicate devices do not raise different questions of safety and effectiveness. As a result, Smith & Nephew concludes the subject AETOS Shoulder System Meta Humeral Prosthesis Size 0 is substantially equivalent to cited predicate device.  |

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWS/K252416](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWS/K252416)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
